Anesth Analg 2006;102:405-410
© 2006 International Anesthesia Research Society
doi: 10.1213/01.ane.0000194300.56739.1a
PEDIATRIC ANESTHESIA
The Size 1 ProSealTM Laryngeal Mask Airway in Infants: A Randomized, Crossover Investigation with the ClassicTM Laryngeal Mask Airway
Kai Goldmann, MD, DEAA,
Christine Roettger, FW, and
Hinnerk Wulf, MD
Department of Anaesthesia and Intensive Care Therapy, Philipps University Marburg, Germany
Address correspondence and reprint requests to Kai Goldmann, MD, DEAA, Attending Anesthesiologist, Department of Anaesthesia and Intensive Care Therapy, Philipps University Marburg, 35033 Marburg, Germany. Address e-mail to Kaigoldmann1{at}aol.com.
Many problems with the ClassicTM laryngeal mask airway (CLMA) in infants are believed to be related to its inadequate cuff design. One of the main limitations of the CLMA is that the resulting low-pressure seal can be inadequate for positive pressure ventilation (PPV). The ProSealTM LMA (PLMA), a new laryngeal mask airway with a modified cuff, has been shown to form a more effective seal than the CLMA in children. The first infant size PLMA, size 1 , became available recently. We studied 30 anesthetized, nonparalyzed infants aged 15 mo (230 mo) and weighing 9 kg (512 kg). The CLMA and PLMA were inserted in random order into each patient. Airway leak pressure and maximum tidal volume were measured. Ease of insertion, quality of initial airway, and fiberoptic position were also determined. Gastric tube placement was assessed for the PLMA. The mean airway leak pressure in neutral head position (26.7 versus 18.9 cm H2O), maximum flexion (35.6 versus 28.2 cm H2O), and the mean maximum tidal volume (312 versus 260 mL) were significantly higher for the PLMA (P < 0.01). Air entered the stomach in eight patients with the CLMA but did not with the PLMA. Gastric tube placement was possible in all but one patient. In three patients, the use of the PLMA led to some degree of clinically relevant compression of the larynx. The size 1 PLMA seems to be a more suitable device for airway maintenance in infants than the same size CLMA. The ability to insert a gastric tube at the same time, and a significantly higher airway leak pressure than with the CLMA, may have important implications for its use for PPV in infants.
This article has been cited by other articles:

|
 |

|
 |
 
M. Micaglio, R. Bonato, M. De Nardin, M. Parotto, D. Trevisanuto, V. Zanardo, N. Doglioni, and C. Ori
Prospective, randomized comparison of ProSealTM and ClassicTM laryngeal mask airways in anaesthetized neonates and infants
Br. J. Anaesth.,
August 1, 2009;
103(2):
263 - 267.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Patel and R. Bingham
Laryngeal mask airway and other supraglottic airway devices in paediatric practice
CEACCP,
February 1, 2009;
9(1):
6 - 9.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Micaglio, M. Parotto, D. Trevisanuto, and V. Zanardo
Size 1 ProSealTM Laryngeal Mask Airway in Neonates
Anesth. Analg.,
October 1, 2006;
103(4):
1044 - 1045.
[Full Text]
[PDF]
|
 |
|
|