During liver transplantation two arterial catheters are often placed. The Vasotrac® is a noninvasive monitor that provides radial arterial blood pressures by a tonometric method. We investigated whether the Vasotrac® would be an accurate substitute for an arterial catheter by comparing Vasotrac® blood pressures with simultaneous direct radial blood pressures recorded from the contralateral arm in 14 patients undergoing liver transplantation. Correlation between the two methods was calculated and a Bland-Altman analysis performed to assess agreement. A total of 6468 simultaneous measurements were made over a duration of 1.57.5 h per case. For mean arterial blood pressure 57% of Vasotrac® measurements were within 10% of direct arterial measurement. Correlation (r) was 0.82. Vasotrac® bias was +5.4 mm Hg and limits of agreement were ±18.6 mm Hg. For systolic arterial blood pressure 65% of Vasotrac® measurements were within 10% of direct arterial measurement. Correlation was 0.78. Vasotrac® bias was +7.6 mm Hg and limits of agreement ±25 mm Hg. For diastolic arterial blood pressure 57% of Vasotrac® measurements were within 10% of direct arterial measurement. Correlation was 0.82. Vasotrac® bias was +3.3 mm Hg and limits of agreement ±15 mm Hg. We conclude that the Vasotrac® is not adequately accurate to substitute for direct arterial blood pressure monitoring in liver transplantation.
|