JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Visaria, R. K.
Right arrow Articles by Westenskow, D. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Visaria, R. K.
Right arrow Articles by Westenskow, D. R.
Related Collections
Right arrow Monitoring (Non-cardiac)
Right arrow Technology

Anesth Analg 2006;103:888-893
© 2006 International Anesthesia Research Society
doi: 10.1213/01.ane.0000237271.78867.1f


TECHNOLOGY, COMPUTING, AND SIMULATION

Model-Based Detection of Endobronchial Intubation

Rachana K. Visaria, PhD{dagger}, and Dwayne R. Westenskow, PhD*{dagger}

From the Departments of *Anesthesiology and {dagger}Bioengineering, University of Utah, Salt Lake City, Utah.

Address correspondence and reprint requests to Dwayne R. Westenskow, PhD, Department of Anesthesiology, University of Utah, 3C 444 SOM, Salt Lake City, UT 84132. Address e-mail to Dwayne.Westenskow{at}hsc.utah.edu.

To detect endobronchial intubation (EBI) noninvasively in real time, we developed a novel, automated, lumped model-based approach. The model uses routinely monitored airway pressure and flow as inputs. The specificity of the method in detecting EBI was determined by testing events of stiff chest wall (SCW) in the absence of EBI. EBI was induced in 10 anesthetized, paralyzed, and mechanically ventilated mongrel dogs (19–45 kg) by advancing the endotracheal tube into the right mainstem bronchus. The event of SCW was created by wrapping a pressure cuff around the chest. Airway pressure and flow were continuously recorded at the mouth, and respiratory impedance was estimated from these signals. Model parameters were iteratively identified until the root mean square error between the respiratory and model-predicted impedance was minimum. The change in model parameters during EBI from baseline was analyzed. In nine of 10 cases, it was determined that during EBI, the model’s compliance element (C1) decreased ≥50% and model’s resistance element (R2) changed ≤10-fold from baseline. Testing this rule on 40 cases of SCW, four false positives were obtained. During SCW, R1 and R2 increased, whereas C2 decreased significantly from baseline. This preliminary study is a promising step toward noninvasive, real-time detection of EBI to aid clinicians in decision making.




This article has been cited by other articles:


Home page
J Intensive Care MedHome page
P. Rudraraju and L. A. Eisen
Analytic Review: Confirmation of Endotracheal Tube Position: A Narrative Review
J Intensive Care Med, September 1, 2009; 24(5): 283 - 292.
[Abstract] [PDF]


Home page
Anesth. Analg.Home page
S. Tejman-Yarden, A. Zlotnik, L. Weizman, J. Tabrikian, A. Cohen, N. Weksler, and G. M. Gurman
Acoustic Monitoring of Lung Sounds for the Detection of One-Lung Intubation
Anesth. Analg., August 1, 2007; 105(2): 397 - 404.
[Abstract] [Full Text] [PDF]




Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2006 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press