BACKGROUND: Cardiac output (CO) and invasive hemodynamic measurements are useful during liver transplantation. The pulmonary artery catheter (PAC) is commonly used for these patients, despite the potential complications. Recently, a less invasive device (Vigileo®/FloTracTM) became available, which estimates CO using arterial pressure waveform analysis without external calibration. In this study, we compared CO obtained with a PAC using automatic thermodilution, instantaneous CO stat-mode (ICOSM), and CO obtained with the new device, arterial pressure waveform analysis (APCO) in patients undergoing liver transplantation.
METHODS: Twenty sets of simultaneous measurements of APCO and ICOSM were determined in sedated and mechanically ventilated patients undergoing liver transplantation. Time points were as follows: after PAC insertion (T1–3), after portal clamping (T4–6), during anhepathy (T7–9), after graft reperfusion (T10–15), and in the postoperative period in the intensive care unit (T15–20).
RESULTS: We enrolled 20 patients and 400 measurements were obtained. No data were rejected. Bias between ICOSM and APCO was 0.8 L/min, 95% limits of agreement were –1.8 to 3.5 L/min. The percentage error was 43%. Bias between ICOSM and APCO was correlated with systemic vascular resistance [r2 = 0.55, P < 0.0001, y = 15.8–2.2 ln(x)] and subgroup analysis revealed an increase in the bias and in the percentage error in patients with low systemic vascular resistance (Child-Pugh grade B and C patients). There was no difference between the different surgical periods.
CONCLUSIONS: Our results suggest that Vigileo/FloTrac CO monitoring data do not agree well with those of automatic thermodilution in patients undergoing liver transplantation, especially in Child-Pugh grade B and C patients with low systemic vascular resistance.
This article has been cited by other articles:
![]() |
M. R. Concha, V. F. Mertz, L. I. Cortinez, K. A. Gonzalez, and J. M. Butte Pulse Contour Analysis and Transesophageal Echocardiography: A Comparison of Measurements of Cardiac Output During Laparoscopic Colon Surgery Anesth. Analg., July 1, 2009; 109(1): 114 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gouvea, R. Diaz, L. Auler, R. Toledo, and J. M. Martinho Evaluation of the pulse pressure variation index as a predictor of fluid responsiveness during orthotopic liver transplantation Br. J. Anaesth., May 19, 2009; (2009) aep123v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Chatti, S. de Rudniki, S. Marque, A. S. Dumenil, A. Descorps-Declere, A. Cariou, J. Duranteau, M. Aout, E. Vicaut, and B. P. Cholley Comparison of two versions of the Vigileo-FloTracTM system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements Br. J. Anaesth., April 1, 2009; 102(4): 463 - 469. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Compton and J.-H. Schafer Noninvasive Cardiac Output Determination: Broadening the Applicability of Hemodynamic Monitoring Seminars in Cardiothoracic and Vascular Anesthesia, March 1, 2009; 13(1): 44 - 55. [Abstract] [PDF] |
||||
![]() |
M. Cannesson, H. Musard, O. Desebbe, C. Boucau, R. Simon, R. Henaine, and J.-J. Lehot The Ability of Stroke Volume Variations Obtained with Vigileo/FloTrac System to Monitor Fluid Responsiveness in Mechanically Ventilated Patients Anesth. Analg., February 1, 2009; 108(2): 513 - 517. [Abstract] [Full Text] [PDF] |
||||
|