JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


Anesth Analg 2008; 107:2079-2084
© 2008 International Anesthesia Research Society
doi: 10.1213/ane.0b013e318187c9bb
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Remerand, F.
Right arrow Articles by Fusciardi, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Remerand, F.
Right arrow Articles by Fusciardi, J.
Related Collections
Right arrow Monitoring (Cardiac)
Right arrow Regional Anesthesia
Right arrow Technology
Right arrow Pharmacology


ANALGESIA

Elastomeric Pump Reliability in Postoperative Regional Anesthesia: A Survey of 430 Consecutive Devices

Francis Remerand, MD, Anne Sophie Vuitton, MD, Michel Palud, MD, Sylvie Buchet, MD, Xavier Pourrat, PharmD, Annick Baud, MD, Marc Laffon, MD, PhD, and Jacques Fusciardi, MD

From the Groupement d’Anesthésie Réanimation, Hôpital Trousseau, CHU Tours, Université François Rabelais, 37041 Tours cedex 1, France.

Address correspondence to Dr Francis Remérand, Groupement d’Anesthésie Réanimation, Hôpital Trousseau, Chambray lès Tours, CHU Tours, 37041 Tours cedex 1, France. Address e-mail to sabetrem{at}yahoo.com.

Abstract

BACKGROUND: Postoperative analgesia via continuous perineural infusion of local anesthetics compares favorably with systemic analgesia. Elastomeric pumps increase patient satisfaction compared with electronic models. In in vitro investigations, infusions remained within 15% of their designated set rates. We assessed in vivo the infusion rate of elastomeric pumps in regional analgesia after orthopedic surgery.

METHODS: All consecutive elastomeric pumps were retrospectively studied during a 10-mo period. Perineural catheters were inserted preoperatively and connected postoperatively to elastomeric pumps filled with ropivacaine 0.2%. Before infusion, elastomeric pumps and ropivacaine were stored at room temperature. Two models of pumps were randomly used: InfusorTM LV5 (Baxter, France) or EasypumpTM (Braun, Germany), both set at 5 mL/h. Nurses weighed the devices at the bedside using a portable electronic scale several times a day until catheter removal. Weights over time allowed accurate deflation profile assessment and flow rate calculation. An unchanged weight over time indicated either an obstructed catheter or an ineffective device.

RESULTS: After connection to the catheter, 88 devices did not deflate (80 Easypump of 300 and 8 Infusor of 130, P < 0.0001). One Easypump was impossible to deflate, even after disconnection from its catheter. In two cases, catheters were obstructed. In 21 cases, catheters were removed 11 to 72 h later without being tested for patency. In 24 cases, pumps correctly deflated after catheters were injected without difficulty with a local anesthetic bolus. The remaining 40 devices spontaneously started to deflate 6 to 43 h after their connection. These 88 elastomeric pumps were associated with higher maximal visual analog scale scores during the first postoperative night than devices showing immediate deflation after connection (34 ± 21 mm vs 26 ± 19 mm, P = 0.006). Flow rates were calculated over a mean period of 54 ± 18 h (Easypump) and 49 ± 19 h (Infusor). The flow rates differed from those set by manufacturers (5 mL/h ± 15%) in 47% of Easypump and in 34% of Infusor devices (P = 0.01).

CONCLUSIONS: In vivo reliability of elastomeric pumps is different than in vitro. In the event of early insufficient postoperative perineural analgesia, an absence of deflation of the elastomeric pump must be considered. We recommend weighing these devices every 3 h during the first 24 h of infusion.







Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2008 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press