JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


Anesth Analg 2009; 108:1193-1197
© 2009 International Anesthesia Research Society
doi: 10.1213/ane.0b013e3181949afd
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Dosch, M. P.
Right arrow Articles by Lechner, S.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Dosch, M. P.
Right arrow Articles by Lechner, S.
Related Collections
Right arrow Monitoring (Non-cardiac)
Right arrow Technology
Right arrow Pharmacology


TECHNOLOGY, COMPUTING, AND SIMULATION

Time to a 90% Change in Gas Concentration: A Comparison of Three Semi-Closed Anesthesia Breathing Systems

Michael P. Dosch, MS, CRNA*, Robert G. Loeb, MD{dagger}, Tiffany L. Brainerd, MD{dagger}, John F. Stallwood, MS, CRNA*, and Steven Lechner, MS, CRNA*

From the *Nurse Anesthesia, University of Detroit Mercy, Detroit, Michigan; and {dagger}Department of Anesthesiology, University of Arizona, Tucson, Arizona.

Address correspondence to Michael P. Dosch, MS, CRNA, Nurse Anesthesia, University of Detroit Mercy, 4001 W McNichols Rd., Detroit, MI 48221-3038. Address e-mail to doschmi{at}udmercy.edu.

BACKGROUND: The speed with which gas concentration can be changed in the anesthesia breathing system affects the rate of denitrogenation, anesthesia induction, and emergence. Breathing system design also affects the speed at which gas concentration can be changed during maintenance. In this study, we sought to determine the speed of changes in gas concentration in modern semi-closed breathing systems. We hypothesized that equilibrium would be reached most quickly in breathing systems with smaller volume, and at high fresh gas flows.

METHODS: Three anesthesia workstations were studied in vitro: the ADU (Datex-Ohmeda, now a division of GE Medical, Madison, WI), the Fabius GS with a COSY-1 breathing system (Draeger Medical, Telford, PA), and the Aestiva (Datex-Ohmeda, now a division of GE Medical, Madison, WI). The breathing systems were connected to a test lung and ventilated with air. The fresh gas flow was then changed to oxygen at rates of 1, 2, 4, 6, or 8 L/min, and times to 50%, 63%, 66%, 75%, and 90% change in oxygen concentration within the test lung were recorded. Ten trials were performed for each breathing system, at each fresh gas flow. The results were analyzed with a split-plot analysis of variance followed by post hoc tests with a Bonferroni correction.

RESULTS: At flows of 6 or 8 L/min, times to equilibration did not differ among the three breathing systems. At flows of 1 to 2 L/min, the gas concentration changed faster with the ADU than with the Aestiva or Fabius (P < 0.001). At 4 L/min, the ADU was faster than Aestiva (P < 0.001), but not Fabius.

CONCLUSIONS: We concluded that, other than fresh gas flow rate, breathing system volume has the biggest effect on time to equilibrium when the composition of the fresh gas inflow is changed. The position of components (e.g., valves, carbon dioxide absorber, fresh gas inlet, ventilator bellows or piston) within the breathing system has a less pronounced effect.







Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2009 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press