Received from the Neuroanesthesia Research Laboratory, Department of Anesthesiology, University of California at San Diego, La Jolla, California and the Department of Anesthesiology, Veterans' Administration Medical Center, San Diego, California.
Abstract
The authors sought to define the relative sensitivities of endtidal carbon dioxide analysis (ETco2), end-tidal nitrogen analysis (ETN2), and pulmonary artery pressure (PAP) monitoring in the detection of venous air embolism (VAE). Serial injections of air (0.25, 0.5, 0.75, 1.0, arid 1.5 ml/kg) were performed in six mongrel dogs. The frequency with which positive responses (PAP increase > 2 mm Hg; ETCO2 decrease > 0.2%; ETN2 increase > 0.04%) were observed following VAE was not different for the three methods. The response time (time to maximum change following VAE) was significantly more rapid for PAP and ETN2 than for ETCO2; although the range for the three methods was narrow, e.g., for 1.5 ml/kg—PAP, 0.92 ± 0.7 (SD) min; ETN2, 1.20 ± 0.5 min; ETCO2, 1.85 ± 0.7 min. The time from injection of air to return to baseline levels was significantly more rapid for ETN2 than for ETCO2 which was in turn significantly significantly faster than PAP, e.g., for 1.5 ml/kg—ETN2, 8.0 ± 4.3 min; ETCO2, 19.4 ± 6.0 min; PAP, 23.8 ± 6.1 min. The results indicate that, where the capacity to identify increases in expired nitrogen on the order of 0.04% can be achieved, ETN2 monitoring mill identify VAE events with a sensitivity similar to that of PAP and ETCO2. However, the difficulties inherent in achieving this level of nitrogen detection sensitivity probably represent a current major limitation in the application of this method. Furthermore, the data indicate that, after VAE, ETN2 will return to preinjection levels although PAP and ETCO2 remain abnormal. This observation suggests that ETN2 may not be a reliable indicator of recovery from the physiologic impact of VAE, and may therefore not be the optimum method to base decisions regarding resumption of the head-up posture and continuation of surgery during procedures in which VAE bas occurred.
Key Words: EMBOLISM—air MONITORING—endtidal gases
This article has been cited by other articles:
![]() |
S. T. Schafer, J. Lindemann, P. Brendt, G. Kaiser, and J. Peters Intracardiac Transvenous Echocardiography Is Superior to Both Precordial Doppler and Transesophageal Echocardiography Techniques for Detecting Venous Air Embolism and Catheter-Guided Air Aspiration Anesth. Analg., January 1, 2008; 106(1): 45 - 54. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Zar and W. W. Wu The Inability to Detect Expired Carbon Dioxide After Endotracheal Intubation as a Result of One-Way Valve Obstruction of the Endotracheal Tube Anesth. Analg., October 1, 2001; 93(4): 971 - 972. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. T. Hegde and R. N. Avatgere Air embolism during anaesthesia for shoulder arthroscopy Br. J. Anaesth., December 1, 2000; 85(6): 926 - 927. [Abstract] [Full Text] [PDF] |
||||
|