Anesth Analg 2002;95:1331-1336
© 2002 International Anesthesia Research Society
TECHNOLOGY, COMPUTING, AND SIMULATION
The Delivery Rate Accuracy of Portable Infusion Pumps Used for Continuous Regional Analgesia
Brian M. Ilfeld, MD*,
Timothy E. Morey, MD*, and
F. Kayser Enneking, MD*
Departments of *Anesthesiology and Orthopedics and Rehabilitation, University of Florida College of Medicine, Gainesville, Florida
Address correspondence and reprint requests to F. Kayser Enneking, MD, Department of Anesthesiology, PO Box 100254, Gainesville, FL 32610-0254. Address e-mail to enneking{at}anest2 anest.ufl.edu.
Portable pumps used for local anesthetic infusion during continuous regional analgesia are gaining acceptance. These pumps are often used for ambulatory patients who are medically unsupervised throughout most of the infusion. However, the performance of these pumps, which infuse potentially toxic medication, has not been independently investigated. We investigated the flow rate accuracy, consistency, and profiles of various portable pumps often used for local anesthetic infusion during continuous regional analgesia. By using a computer/scale combination within a laboratory to record infusion rates, 6 pumps were tested with their flow regulators at expected (30°32°C) and increased (34°36°C) temperatures. Infusion rate accuracy differed significantly among the pumps, exhibiting flow rates within ±15% of their expected rate for 18%100% of their infusion duration. An increase in temperature also affected pumps to differing degrees, with infusion rates increasing from 0% to 25% for each model tested. These results suggest that factors such as flow rate accuracy and consistency, infusion profile, and temperature sensitivity should be considered when choosing and using a portable infusion pump for local anesthetic administration.
IMPLICATIONS: Portable pumps often used for local anesthetic infusion during continuous regional analgesia exhibit varying degrees of delivery rate accuracy and consistency. Furthermore, increases in temperature result in an increased infusion rate for various pumps investigated. These factors should be taken into consideration when choosing and using a portable infusion pump.
This article has been cited by other articles:

|
 |

|
 |
 
F. Remerand, A. S. Vuitton, M. Palud, S. Buchet, X. Pourrat, A. Baud, M. Laffon, and J. Fusciardi
Elastomeric Pump Reliability in Postoperative Regional Anesthesia: A Survey of 430 Consecutive Devices
Anesth. Analg.,
December 1, 2008;
107(6):
2079 - 2084.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. R. Cichocki JR., T. A. Barbolt, S. J. Savidge, G. Chen, and J. Zhou
A Bioabsorbable Microcatheter for the Continuous and Direct Treatment of Surgical Sites
J Biomater Appl,
September 1, 2007;
22(2):
145 - 162.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. A. Skryabina and T. S. Dunn
Disposable infusion pumps.
Am. J. Health Syst. Pharm.,
July 1, 2006;
63(13):
1260 - 1268.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. M. Ilfeld and F. K. Enneking
Continuous Peripheral Nerve Blocks at Home: A Review
Anesth. Analg.,
June 1, 2005;
100(6):
1822 - 1833.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. M. Ilfeld, T. E. Morey, X. Capdevila, C. Dadure, N. Bernard, S. Lopez, P. Macaire, and P. Aknin
Use of Term "Patient-Controlled" May Be Confusing in Study of Elastometric Pump * Response
Anesth. Analg.,
September 1, 2003;
97(3):
916 - 917.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. M. Ilfeld, T. E. Morey, T. W. Wright, L. K. Chidgey, and F. K. Enneking
Continuous Interscalene Brachial Plexus Block for Postoperative Pain Control at Home: A Randomized, Double-Blinded, Placebo-Controlled Study
Anesth. Analg.,
April 1, 2003;
96(4):
1089 - 1095.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|