JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (15)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mahajan, A.
Right arrow Articles by Marijic, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mahajan, A.
Right arrow Articles by Marijic, J.
Related Collections
Right arrow Cardiovascular
Right arrow Monitoring (Cardiac)
Right arrow Pediatrics

Anesth Analg 2003;97:1283-1288
© 2003 International Anesthesia Research Society


PEDIATRIC ANESTHESIA

Pulse Contour Analysis for Cardiac Output Monitoring in Cardiac Surgery for Congenital Heart Disease

Aman Mahajan, MD, Afshin Shabanie, MD, Judi Turner, MD PhD, Michael J. Sopher, MD, and Jure Marijic, MD

Department of Anesthesiology, David Geffen School of Medicine, University of California Los Angeles

Address correspondence and reprint requests to Aman Mahajan, MD, Department of Anesthesiology, Box 951778, David Geffen School of Med at UCLA, Los Angeles, CA 90095. Address e-mail to amahajan{at}mednet.ucla.edu

Conventional methods of cardiac output monitoring using pulmonary artery catheters may not be feasible in patients with congenital heart disease because of patients’ small size or aberrant anatomy. We studied the accuracy of a new device, which uses pulse contour analysis to measure continuous cardiac output, in children and adults undergoing congenital heart surgery. Sixteen patients, median ages 7 yr old, were included in this prospective study. One-hundred-ninety-one data points were obtained in the pre- and postcardiopulmonary bypass periods and in the first 12 h after intensive care unit admission. We evaluated the relationship between cardiac index (CI) derived from transpulmonary thermodilution (TDCI) and CI derived from pulse contour analysis (PCCI). Bias and limits of agreement between TDCI and PCCI over all time periods were 0.1 ± 1.94, indicating a wide dispersion of the data. Coefficient of correlation (r) between the TDCI and PCCI was 0.7. Although in previous studies, PCCI has been suggested to be accurate in adult cardiac surgery, we found it to be less reliable in our study patients, even after shunt correction. The relationships of the volume and pressure based measures of preload, intrathoracic blood volume index (ITBI), and central venous pressure with CI were also investigated. After repair, correlation (r) between PCCI or TDCI and ITBI (0.56 and 0.71, respectively) was better than that between PCCI or TDCI and CVP (0.16 and 0.11, respectively), indicating greater validity of ITBI as a measure of preload.

IMPLICATIONS: Our results suggest that the pulse contour analysis cardiac output (CO) monitoring in patients undergoing congenital heart surgery may not provide as accurate or reliable measures of CO as previously suggested. The volume-based variable of preload intrathoracic blood volume index (ITBI) has better correlation with cardiac index (CI) than the central venous pressure, suggesting that ITBI may be a better indicator of preload.




This article has been cited by other articles:


Home page
Anesth. Analg.Home page
O. J. Liakopoulos, J. K. Ho, A. Yezbick, E. Sanchez, C. Naddell, G. D. Buckberg, R. Crowley, and A. Mahajan
An Experimental and Clinical Evaluation of a Novel Central Venous Catheter with Integrated Oximetry for Pediatric Patients Undergoing Cardiac Surgery
Anesth. Analg., December 1, 2007; 105(6): 1598 - 1604.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
U. Fakler, Ch. Pauli, G. Balling, H.P. Lorenz, A. Eicken, M. Hennig, and J. Hess
Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery
J. Thorac. Cardiovasc. Surg., January 1, 2007; 133(1): 224 - 228.
[Abstract] [Full Text] [PDF]




Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins with the assistance of Stanford University Libraries' HighWire Press®. Copyright 2006 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press
Copyright © 2003 by the International Anesthesia Research Society.