The Relationship Between Bispectral Index and Electroencephalographic Parameters During Isoflurane Anesthesia
Yasuhiro Morimoto, MD*,
Satoshi Hagihira, MD ,
Yumika Koizumi, MD*,
Kazuyoshi Ishida, MD*,
Mishiya Matsumoto, MD*, and
Takefumi Sakabe, MD*
*Department of Anesthesiology-Resuscitology, Yamaguchi University School of Medicine, Yamaguchi, Japan; and
Department of Anesthesiology, Osaka Prefectural Habikino Hospital, Osaka, Japan
Address correspondence and reprint requests to Yasuhiro Morimoto, MD, Department of Anesthesiology-Resuscitology, Yamaguchi University School of Medicine, 1-1-1 Minami-Kogushi Ube, Yamaguchi, 755-8505, Japan. Address e-mail to naa01346{at}nifty.ne.jp
Bispectral index (BIS) integrates various electroencephalographic (EEG) parameters into a single variable. However, the exact algorithm used to synthesize the parameters to BIS values is not known. The relationship between BIS and EEG parameters was evaluated during nitrous oxide/isoflurane anesthesia. Twenty patients scheduled for elective ophthalmic surgery were enrolled in the study. After EEG recording with a BIS monitor (A-1050) was begun, general anesthesia was induced and maintained with 0.5%2% isoflurane and 66% nitrous oxide. Using software we developed, we continuously recorded BIS, spectral edge frequency 95% (SEF95), and EEG parameters such as relative beta ratio (BetaRatio), relative synchrony of fast and slow wave (SynchFastSlow), and burst suppression ratio. BetaRatio was linearly correlated with BIS (r = 0.90; P < 0.01; n = 253) at BIS more than 60. At a BIS range of 30 to 80, SynchFastSlow (r = 0.60; P < 0.01; n = 3314) and SEF95 (r = 0.75; P < 0.01; n = 3339) were linearly correlated with BIS. The correlation between BIS and SEF95 was significantly better than the correlation between BIS and SynchFastSlow (P < 0.01). At BIS less than 30, the burst suppression ratio was inversely linearly correlated with BIS (r = 0.76; P < 0.01; n = 65). At BIS less than 80, burst-compensated SEF95 was linearly correlated with BIS (r = 0.78; P < 0.01; n = 3404). In the range of BIS from 60 to 100, BIS can be calculated from BetaRatio. At surgical levels of anesthesia, BIS and SynchFastSlow (a parameter derived from bispectral analysis) or burst-compensated SEF95 (derived from power spectral analysis) are well correlated. However, our results show that SynchFastSlow has no advantage over SEF95 in calculation of BIS.
IMPLICATIONS: The relationship between bispectral index (BIS) and electroencephalographic parameters was evaluated during nitrous oxide/isoflurane anesthesia. At surgical levels of anesthesia, BIS and the relative synchrony of fast and slow wave (a parameter derived from bispectral analysis) or burst-compensated spectral edge frequency 95% (a parameter derived from power spectral analysis) are well correlated.
This article has been cited by other articles:

|
 |

|
 |
 
M. S. Avidan, L. Zhang, B. A. Burnside, K. J. Finkel, A. C. Searleman, J. A. Selvidge, L. Saager, M. S. Turner, S. Rao, M. Bottros, et al.
Anesthesia Awareness and the Bispectral Index
N. Engl. J. Med.,
March 13, 2008;
358(11):
1097 - 1108.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y. Morimoto, S. Hagihira, S. Yamashita, Y. Iida, M. Matsumoto, S. Tsuruta, and T. Sakabe
Changes in electroencephalographic bicoherence during sevoflurane anesthesia combined with intravenous fentanyl.
Anesth. Analg.,
September 1, 2006;
103(3):
641 - 645.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
V. Fodale, C. Pratico, M. Tescione, S. Tanania, T. Lucanto, and L. B. Santamaria
Tramadol does not modify the Bispectral Index during anaesthesia with sevoflurane and remifentanil
Br. J. Anaesth.,
August 1, 2005;
95(2):
212 - 215.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Heller, R. Hatami, P. Mullin, R. R. Sciacca, A. G. Khandji, M. Hamberger, R. Emerson, and E. J. Heyer
Bilateral Bispectral Index Monitoring During Suppression of Unilateral Hemispheric Function
Anesth. Analg.,
July 1, 2005;
101(1):
235 - 341.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y. Morimoto, A. Matsumoto, Y. Koizumi, T. Gohara, T. Sakabe, and S. Hagihira
Changes in the Bispectral Index During Intraabdominal Irrigation in Patients Anesthetized with Nitrous Oxide and Sevoflurane
Anesth. Analg.,
May 1, 2005;
100(5):
1370 - 1374.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y. Morimoto, Y. Monden, K. Ohtake, T. Sakabe, and S. Hagihira
The Detection of Cerebral Hypoperfusion with Bispectral Index Monitoring During General Anesthesia
Anesth. Analg.,
January 1, 2005;
100(1):
158 - 161.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|