JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Horowitz, P. E.
Right arrow Articles by Smith, R. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Horowitz, P. E.
Right arrow Articles by Smith, R. A.
Related Collections
Right arrow Critical Care
Right arrow Blood
Right arrow Equipment
Right arrow Technology

Anesth Analg 2004;99:788-792
© 2004 International Anesthesia Research Society
doi: 10.1213/01.ANE.0000129995.42008.A4


TECHNOLOGY, COMPUTING, AND SIMULATION

Flow Rates and Warming Efficacy with Hotline and Ranger Blood/Fluid Warmers

Peter E. Horowitz, MD, Miguel A. Delagarza, MD, Jaime J. Pulaski, MD, and Robert A. Smith, PhD

Department of Anesthesiology, College of Medicine, University of South Florida, Tampa, Florida

Address correspondence and reprint requests to Peter E. Horowitz, MD, 12901 Bruce B. Downs Blvd., MDC Box 59, Tampa, FL 33612-4799. Address e-mail to phorowit{at}hsc.usf.edu

The heating capabilities of a water bath blood/fluid warmer, Hotline, have proven superior to those of other devices. The dry heat warmer Ranger has not previously been compared with the Hotline. We evaluated these devices in terms of flow rates and efficacy of warming. We delivered room temperature (21°C) saline and 10°C packed red blood cells (RBCs) by using 90 mm Hg (gravity equivalent) and 300 mm Hg bag pressure and various sizes of IV catheters. The outflow from each device was connected to an inline thermistor, and simultaneous measurements of outflow temperature and flow volume per minute were recorded. Additional data points were obtained with a roller pump that delivered flows of 1–6 L/h through each device. We calculated the effect of these flow rates and outflow temperatures on the mean body temperature (MBT) of a 70-kg patient. The Hotline and Ranger had similar flow rates at 90 and 300 mm Hg pressure infusion when studied with various sizes of IV catheters. Hotline was able to deliver warmer RBCs and saline at slower flow rates (1–4 L/h), but because changes in MBT were almost identical, there was no clinically important advantage, and almost no heat was transferred at these slower flow rates. At more rapid flow rates (>4 L/h), the Ranger warmed RBCs and saline better and produced smaller decreases in MBT than the Hotline. The use of the Hotline for rapid infusions, especially of cold RBCs, is not recommended because of low outflow temperatures and decreases in MBT that were three times larger than those seen with the Ranger.

IMPLICATIONS: We compared the ability of a new dry heat blood/fluid warmer (Ranger) with the water bath warmer Hotline in terms of flow rates and outflow temperatures by using room temperature saline and 10°C diluted packed red blood cells (RBCs). Flow rates were similar, but the Ranger provided higher outflow temperatures at flows >4 L/h, especially during rapid pressurized infusions of both RBCs and saline.







Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2004 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press