We designed the present study to investigate the electrical resistance of commercially available epidural catheters and to search for products and procedures suitable for nerve stimulation-guided insertion. Four types of epidural catheters were evaluated: 2 nonwire-reinforced catheters (19-gauge and 20-gauge nylon) and 2 wire-reinforced catheters (19-gauge without stylet and 20-gauge with stylet). The resistance of a catheter was calculated from the voltage level proportional to the fixed resistance in series circuit. In case of physiologic saline, the resistance of nonreinforced catheters was more than 700 k , whereas the wire-reinforced catheter was 14.4 ± 0.20 k without stylet and 10.1 ± 0.42 k with stylet. When the stylet was passed through a 20-gauge nylon catheter, the resistance decreased to 49.2 ± 1.96 k . When catheters were primed with 10% hypertonic saline, the resistance of both nonreinforced catheters decreased by one third compared with physiologic saline. The electrical resistance of the saline-filled epidural catheters significantly differed among products tested. We conclude that epidural catheterization that is guided by electrical stimulation should be performed only with catheters equipped with spiral stainless steel wire reinforcement or with a stainless steel stylet.
Epidural analgesic methods are widely performed for postoperative pain control and pain therapy. There have been some reports in which the position of the epidural catheter tip was confirmed by monitoring the contraction of skeletal muscle using nerve stimulators (i.e., Tsui test) (13). This method is especially useful in pediatric epidural catheterization via the caudal approach and is expected to reduce the risk of neurological damage during catheter insertion under general anesthesia (2,3). However, all previous reports concerning the Tsui test were performed using stainless steel wire-reinforced catheters (13). According to the Tsui test, 10 mA direct current should be sufficient to cause muscle contraction. An ordinary electrical nerve stimulator yields a maximal 400 V output in clinical use. Therefore, by definition (i.e., V = IR), the resistance of an epidural catheter must be <40 k to yield a 10-mA current. Using a simplified electrical model, we have examined the resistances of four types of epidural catheters to determine whether these catheters were suitable for the Tsui test (1).
Four types of epidural catheters were evaluated: 2 without stainless steel coiling (19-gauge Epineed®; Terumo, Inc. and 20-gauge Perifix®; B. Braun, Inc.) and 2 with stainless steel coiling (19-gauge FlexTip Plus® and 20-gauge FlexTip Plus® with a stylet; Arrow International, Inc.). The 20-gauge Flex Tip Plus with stylet was developed for pediatric caudal placement. The materials and characteristics of each catheter are shown in Table 1.
The experimental apparatus is shown in Figure 1. The epidural catheter being tested was filled with either physiologic saline or hypertonic saline and the proximal end was connected to the anode of an electrical stimulator (SEM-4201, Nihon Kohden, Japan) with an electrocardiograph (ECG) adapter (Johans ECG adapter, Arrow International, Inc.). The distal end of the catheter was inserted into a U-shaped tube (8.0 mm inner diameter) filled with physiologic saline and the tip was positioned 1 cm from the metal electrode. The metal electrode was connected to a fixed resistor (100 k
Initially, the resistance of 4 types of catheters primed with physiologic saline was measured. Then the effect of catheter tip configuration was investigated; 20 mm of the 20-gauge nylon catheter (originally multihole, closed-tip) was removed to make it a single-hole, open-tip catheter, and its resistance was measured. The effect of the stainless steel stylet was then investigated. The stainless steel stylet, from a 20-gauge wire-reinforced catheter, was inserted into a 20-gauge nylon catheter and the resistance was measured. For this purpose, the 20-gauge nylon catheter was cut to approximately 20 cm so that the distance from the tip of the stylet to the tip of the catheter was 25 mm, as in the original styletted catheter. Furthermore, the influence of the distance between catheter tip and stylet tip on resulting resistance was examined. Finally, the resistance of catheters primed with 10% hypertonic saline was investigated for two types of nonwire-reinforced catheters (19-gauge and 20-gauge nylon). In each experiment, 5 catheters were examined and results were expressed as mean ± sd. Data between the two groups were analyzed by Students t-test and P < 0.05 was considered statistically significant.
Calculated resistance are shown in Table 1. Two types of nonwire-reinforced catheters showed very high resistance: the 19-gauge nylon catheter had a resistance of 754.9 ± 14.5 k and the 20-gauge nylon catheter had a resistance of 974.7 ± 58.3 k . Conversely, catheters with stainless steel coiling showed considerably less electrical resistance: the 19-gauge wire-reinforced catheter had a resistance of 14.4 ± 0.20 k . Furthermore, the wire-reinforced catheter with stainless stylet yielded a smaller resistance (10.1 ± 0.42 k ). However, when the stylet was removed from this catheter, catheter resistance increased to almost the same level as for the 19-gauge wire-reinforced catheter.
When catheter tip configuration was changed from multihole and closed end to single open end, the catheter resistance decreased significantly from 974.7 ± 58.3 k
A stainless steel stylet was examined in the 20-gauge nonwire-reinforced catheter. When the 20-gauge nylon catheter was adjusted to fit the stylet length, the resistance measured 782.8 ± 19.5 k
When physiologic saline was exchanged for hypertonic saline during the priming step, the resistance of the two types of nonwire-reinforced catheters, i.e., 19-gauge and 20-gauge nylon, were significantly decreased to 226.6 ± 14.0 k
In general, it is difficult to measure pulse waved direct current immediately in milliamperage, and thus catheter resistance cannot be calculated using the direct current value applied by nerve stimulators. Although our experimental method shown in Figures 1 and 2 is simple, catheter resistance can be calculated using stable electric current, and voltage levels can be easily and accurately measured with oscilloscopes. Once primed with physiologic saline, catheter resistance actually includes a 1-cm saline portion, as shown in Figure 1. When electric stimulation of a nerve is applied in a clinical setting the tip of the catheter must complete an electrical circuit with the nerve. In the Tsui test, saline is injected through the epidural catheter before electrical stimulation. Our method has a 1-cm saline portion, shown in Figure 1, to account for the resistance between the catheter and the nerve in situations where the two do not touch.
The two wire-reinforced epidural catheters, 19-gauge without stylet and 20-gauge with stylet, showed very low resistances, 14.4 ± 0.20 k
When electric stimulation is performed in a patient, additional resistance must be considered, that is, the resistance of the human body, estimated between 16 k
Introducing a stainless steel stylet through a 20-gauge nylon catheter causes the resistance to decrease to 49.9 ± 1.96 k
When we used 10% hypertonic saline for the priming step in the 20-gauge nylon catheter, resistance was reduced from 974 ± 58.3 k In summary, we have examined several commercially available epidural catheters with respect to their suitability in nerve stimulation-guided insertion and have concluded that this procedure requires either a catheter with stainless steel coiling or a stainless steel stylet.