The complete details of the BIS algorithms core technology have not been published. All we know is that the BIS monitor is a "black box" headset and that the BIS value is merely a reflection of a "head-related" biosignal that correlates with changes in certain hypnotic drug effects. Thus, the BIS cannot be considered a true reflection of the depth of anesthesia nor an independent measure of electroencephalographic (EEG) cerebral function.
The purpose of this review is to assess the utility of BIS monitoring by examining the various published reports of all BIS values not coinciding with a clinically judged sedative-hypnotic state, whether arising from an underlying pathophysiology of EEG cerebral function or because of shortcomings in the performance and design of the BIS monitor.
Several models of the BIS are still being used, namely, the A-1000, A-1050, A-2000, and, lately, the BIS-XP. Among these models there are numerous software algorithm revisions and iterations. We must emphasize that the reported performance of a certain BIS model might not necessarily apply to other models. Furthermore, we emphasize that, obviously, the electrophysiological changes in extreme situations, such as hypoglycemic coma or hypovolemic cardiac arrest, cannot be rigorously tested in controlled studies as being a definitive cause of accompanying BIS changes. Thus, the explanations we present for these particular situations are speculations based on electrophysiological changes in similar studies.
| Basic Structure |
|---|
|
|
|---|
(7.512.5 Hz) wave predominance. Light anesthesia is accompanied by a decrease in the
and an increase in ß (12.530 Hz) power. With deepening of anesthesia, slow wave activitynamely
(1.53.5 Hz) and
(3.57.5 Hz) wavesincrease and become more prominent, with a simultaneous decrease in the
and ß activities in all regions. This represents a decrease in the cortical generator of
and ß activities, with a shift toward control by the thalamo-hippocampal-septal generators of
and
activities. All these changes are reversed in the same order with the return of consciousness (1).
|
The BIS is a dimensionless number ranging from 100 (fully awake) to 0 (isoelectric EEG). The BIS is the weighted sum of three descriptors (Fig. 2): Relative BetaRatio, a frequency-domain feature, is the EEG spectral power log (P3047 Hz/P1120 Hz), SynchFastSlow, a bispectral-domain feature, is the bispectral power wave band log (B0.547 Hz/B4047 Hz), and Burst Suppression is a time-domain feature that quantifies the extent of isoelectrical silence. None of these disparate descriptors is particular per se because each has a specific range of effect at which they perform best. BetaRatio is the most influential during light hypnotic states; SynchFastSlow predominates during surgical levels of anesthesia; and Burst Suppression detects very deep anesthesia. The BIS analysis uses a proprietary algorithm that allows the different descriptors to sequentially dominate as the EEG changes character with increasing depth of anesthesia (2).
|
| Paradoxical BIS Changes with Anesthetics |
|---|
|
|
|---|
and
activities (3). This could be attributed to N2Os peculiar withdrawal-suppression EEG phenomenon, in which 710 min from N2Os sudden discontinuation, an "overswing" of paroxysmal bursts of low-frequency
and
waves was shown to diffusely occur throughout the EEG recording, a pattern very similar to that of deep anesthesia (5).
|
However, N2O has a weak cortical action, because it acts mainly through activating the descending inhibitory noradrenergic pathway in the brainstem and spinal cord (6). This effect is completely undetectable by the algorithm that computes the BIS. That is why N2O, both as a sole anesthetic and as an adjunct to IV anesthesia, was shown, in several studies, to result in loss of consciousness without a noticeable change in the BIS A-1000 values (79).
Ketamine
Ketamine does not follow the basic EEG pattern of general anesthesia, as it was shown to cause an opposite effect of an increase in the ß range activity accompanied by a reduction in the
power (10). This odd EEG pattern would be reflected by BIS monitoring, as the BIS A-1050 (7) and the BIS A-2000 version 3.4 (11) were reported to paradoxically increase after ketamine administration.
Inhaled Anesthetics
An increase in isoflurane concentrations from 0.79% to 1.26% was shown to cause a "paradoxical arousal reaction" as it results in an increase in the
and the ß waves in a pattern similar to that of light anesthesia (12). A similar paradoxical arousal in the mean BIS (A-1000 version 3.12) value from 35 to 46 was reported with the increase of isoflurane concentrations from 0.8% to 1.6%. BIS returned to baseline values with the return to 0.8% (13).
Different inhaled anesthetics have different EEG "signatures." For instance, clinical concentrations of halothane produce slow waves and fast "spindle" EEG rhythms, whereas sevoflurane produces sharp, slow EEG waves with relatively few fast rhythms (14). Consequently, BIS values are not the same for equipotent concentrations of different inhaled anesthetics. With the BIS-XP version 4.0 (15), and BIS A-2000 version 3.21 (16), the mean BIS value of 57 with halothane was significantly higher than a mean BIS value of 3233 for equipotent sevoflurane (15) or isoflurane concentrations (16). This indicates that the BIS algorithm, which was not written or validated for halothane, does not accurately reflect the hypnotic effect of halothane anesthesia and may lead to inadvertent halothane overdose.
Propofol
In 2 volunteers, the BIS (A-1000 version 3.22) was reported to remain unchanged at 3540 with escalating propofol TCI, despite the simultaneously recorded raw EEG indicating burst suppression (17). The authors of the report speculated that BIS 3540 could be the "range of uncertainty" or a "blind spot" between the BetaRatio and the Burst Suppression descriptors, where the BIS algorithm could be less sensitive to propofol effect.
The Opioid Effect
Opioid doses of almost 5 times the analgesic concentrations would be required for the appearance of a noticeable EEG depression (18). Unlike IV or inhaled anesthetics, opioids in analgesic concentrations produce minimal or no electrophysiological alterations on the cerebral cortex. This is because "noncortical" structures that are undetectable by the EEG, such as the locus coeruleus-noradrenergic system, are involved in the mechanism of opioid drug effect (19). At constant propofol TCI, remifentanil in escalating concentrations did not change BIS (A-1000 version 3.12) values (20), nor did remifentanil step-by-step reductions change the BIS (A-1000 version 2.0) values (21). This clearly indicates that the BIS monitor is not capable of detecting a direct opioid cortical EEG effect.
However, opioids could still influence BIS monitoring by another peculiar effect, as the addition of fentanyl, sufentanil, remifentanil or alfentanil to TCI propofol resulted in loss of consciousness at smaller propofol effect site concentrations and consequently at higher BIS A-1000 version 3.12 (22), A-2000 version 3.4 (23), and A-1050 (2225) values. This clearly shows that the hypnotic effect of propofol is enhanced by µ agonist opioids. However, BIS does not show this increased hypnotic effect and would only reflect the lower propofol requirements and hence higher BIS values, which may lead to an inadvertent anesthetic overdose.
Opioids could still directly influence BIS monitoring in a totally different manner, namely, through the attenuation of responses to noxious stimuli. Under constant propofol TCI, µ agonist opioids attenuated increases in BIS (A-1000) in response to tracheal intubation (20), to the pinning of the Mayfield head holder fixator (26), and to endotracheal suction (27). This means that BIS values under a constant level of an anesthetic regimen not only indicate a certain level of hypnosis but also reflect the degree of opioid-induced inhibition of response to noxious stimuli. If BIS suddenly increases in response to a noxious stimulus, this could be a cortical arousal reaction reflecting a deficit in the analgesic component of anesthesia that would require increasing the analgesic doses.
| Electric Device Interference with BIS Monitoring |
|---|
|
|
|---|
However, the BIS signal quality bar might not always be helpful, as in numerous occasions when the subtle signal pollution was not detected by the signal quality bar nor displayed as artifacts. For instance, BIS A-1000 version 3.12 (29) and BIS A-2000 (30) were reported to falsely increase from <56 to 7090 when a forced-air-warming blanket, placed directly on a patients forehead, was switched on. BIS returned to 3560 when it was switched off (29,30). Similarly, BIS (A-2000) suddenly increased from 40 to 62 with the start of endoscopic shoulder shaver oscillations and decreased after the shaver device was switched off (31). Furthermore, an otorhinolaryngology positioning system, which created a 20-cm electromagnetic field around the patients head, was reported to increase the BIS (A-2000) from 4045 to 6090 when switched on. BIS immediately decreased to 2832 with switching the system off (32).
In all the above-mentioned reports, the BIS signal quality indicator displayed "optimal signal quality" in all patients (2932), whereas, in some cases, the simultaneously displayed raw EEG showed fast moving waves of high amplitudes resembling those of the awake state (30). This clearly indicates that the air-warming-blanket vibrations (30), the shoulder shaver oscillations (31), and the electromagnetic field (32) could have all created minimal vibrations or frequencies in the BIS electrodes simulating the EEG waves of light anesthesia or an awake state. None of these subtle signal pollutions was detected by the BIS monitor and hence were not displayed as artifacts, thus opening the potential for inadvertent anesthetic overdose.
| Effect of Different Clinical Conditions on BIS Monitoring |
|---|
|
|
|---|
and
waves (Table 1). This proceeds to a widespread increase in the
and
waves at 54 mg/dL blood glucose level (33). With a further decrease to 32 mg/dL, the increase in the
and
waves is associated with a significant decrease in
waves, a pattern very much similar to that of general anesthesia (34). It comes as no surprise that BIS could reflect these EEG changes, as in 2 separate reports with BIS A-2000 (35) and BIS-XP (36), 2 hypoglycemic coma patients with 35 mg/dL (35) and 21 mg/dL blood glucose levels (36) were reported to manifest BIS values as low as 45. BIS rapidly increased to 80 along with the increase of blood glucose levels and the return of consciousness (35,36). Thus, blood glucose levels should be considered as a contributing factor during interpretation of the BIS values.
Hypovolemic Cardiac Arrest
Real time EEG monitoring is not a part of routine anesthesia monitoring. Thus, there are very few documented cases of EEG changes during the pericardiac arrest period. In one rare, recorded case, real time EEG showed generalized isoelectricity 10 s after the onset of asystole. Low-voltage, high-frequency EEG activity began to return 1520 s after manual chest compression restored cerebral perfusion and oxygenation with the essentially "standing by" well-oxygenated arterial blood. Return of cardiac rhythm was associated with the return of normal EEG signal (37). Unlike real time EEG, with the increasing use of BIS monitoring in routine anesthesia, we are encountering more cases of documented BIS "witnessed cardiac arrests." In two reported cases, hypovolemic cardiac arrest evoked a parallel decrease in the BIS values to zero with an isoelectric EEG. As arterial blood pressure was restored by volume replenishment, BIS score increased to premorbid levels (38,39). A major difference between BIS monitoring in the first case and real time EEG was that the decrease and increase in BIS lagged 2 min behind the arterial blood pressure, because of the 60-s delay of the "smoothing algorithm" used to calculate the BIS value (38). However, this was not observed in the second case as the gradual decrease of arterial blood pressure over 20 min minimized the lag between the decrease in arterial blood pressure and the decrease in BIS score (39).
BIS changes could even precede hemodynamic changes. In an interesting report, BIS was shown to decrease from a steady value of 35 to 20, 10 min before the decrease of arterial blood pressure from 120/70 to 65/30 in a patient with fatal major bleeding in an aortic graft (40). This BIS "prodromal" change was attributed to the BIS early detection of propofol pharmacokinetics alteration before the actual decrease of arterial blood pressure (40).
Cerebral Ischemia
BIS was shown to successfully reflect the global cerebral ischemia of asystole (38,39) as well as other forms of localized cerebral ischemia and brain injuries. In 2 case reports, BIS decreased from 4060 to <10 with carotid artery clamping (41,42). The fact that the BIS did not subsequently increase revealed an intracerebral hemorrhage in the first case (41) and severe cerebral ischemia in the second case (42). The BIS (A-2000 version 2.1) later returned to normal values with the restoration of normal cerebral circulation (42).
Hypothermic Cardiopulmonary Bypass (CPB)
In a large group of patients (n = 100), during hypothermic CPB under constant isoflurane anesthesia, BIS (A-1050 version 3.3) was estimated to decrease by 1.12 BIS units for each degree Celsius decrease in body temperature (43). Hypothermia produces a linear decrease in inhaled anesthetic requirements. In fact, hypothermia itself serves as a complete anesthetic at 20°C (44). This is attributed to the increase in the solubility of inhaled anesthetics in the lipid membrane with lower temperature, resulting in larger concentrations of anesthetics being available at a cellular level (44), which would result in a BIS decrease.
Similarly, under constant propofol-alfentanil anesthesia, a mean BIS (A-1000) value of 41 was significantly lower during hypothermic CPB than a mean BIS value of 49 during normothermic CPB (45). Hypothermia (25°27°C) was shown to significantly increase serum propofol concentrations through a decrease in hepatic microsomal enzyme activity and a decrease in propofol biotransformation (46), which would consequently result in a BIS decline.
However, this effect was not consistent among different reports, as during hypothermic CPB (n = 12), BIS (A-1000 version 3.12) slightly decreased with decreasing temperature. A possible reason for not finding a consistent association between BIS and temperature could be that the sample size was too small to demonstrate such an effect (47).
| BIS Values Modified by Abnormal EEG Patterns |
|---|
|
|
|---|
waves, a pattern very much resembling deep anesthesia (48) (Table 1). The BIS monitor reflected this "post-ictal suppression state" as patients awakened from propofol anesthesia at drastically low mean BIS (A-1000 version 3.31) values of 4557 (49,50). Some patients could even open their eyes at a BIS value as low as 7 (50). This post-ictal suppression does not depend on the hypnotic used, as in patients regaining consciousness after electroconvulsive therapy under methohexital anesthesia, the BIS-XP (A-2000) values were still <60 in 75% of the patients at eye opening (51).
Neurological Disorders
The BIS algorithm was developed from volunteers with normal EEG, thus neurological disorders that manifest abnormal EEG patterns would probably affect the BIS monitoring. In patients with Alzheimers type dementia, who show reduced power in the ß band (52), the mean awake BIS (A-2000 version 4.0) was 89 compared with a mean BIS value of 95 in control elderly patients (53). Whereas the mean BIS (A-2000 version 3.21) values both at sevoflurane 1% and after emergence from anesthesia were significantly lower in cerebral palsy mentally retarded children compared with normal children (54). Furthermore, 24 h after all sedative drugs were withdrawn, the mean BIS (A-2000 version 3.4) value was 43 in comatose state intensive care unit (ICU) patients who did not later recover from severe brain injury, whereas the mean BIS value was 63 in those comatose state patients who did later recover from severe brain injury (55).
A BIS (A-2000) value of 0 was shown to accurately indicate brain death (56). However, unlike brain death, cortical cognition of painful stimuli could persist in some permanent vegetative state subjects (57). In a permanent vegetative state patient undergoing dental surgery, BIS (A-1000 version 3.2) values before anesthetic induction were 7485, indicating neurological damage. BIS decreased to 4042 with sevoflurane administration and later increased to 98100 after termination of anesthesia, resembling that of a normal subject (58). This clearly indicates that the BIS is not robust enough to distinguish between integrated and nonintegrated cortical neuron activity. In other words, a high BIS score indicates that the overall level of cortical activity is high, which in a normal subject implies wakefulness, but in a subject with chronic brain disease does not necessarily indicate integrated or meaningful cortical activity.
Low-Voltage EEG
Genetically determined low-voltage EEG variant is defined as EEG amplitude of <20 mV over all head regions. This is a normal variant that occurs in 5%10% of the population and is not associated with any brain dysfunction (59). The gene responsible for this variant is localized on the distal part of chromosome 20q (60). In an interesting case of a fully conscious volunteer exhibiting an unreasonably low awake BIS (A-1000 version 3.11) value of 40 on 2 separate sessions 3 days apart, a 16-lead EEG showed a genetically determined low-voltage EEG (61). Because the BIS algorithm was developed from volunteers with normal EEG, it is therefore expected that the BIS be misled by this abnormal EEG pattern. Thus, it is essential to confirm a normal BIS value in all patients before induction of anesthesia.
The low-voltage EEG is not restricted to the genetically determined variant, as it could also be drug-induced. The mean BIS (A-1000 version 3.31) value plummeted paradoxically during the remifentanil washout phase in 6 patients (62). The same effect was reported with 2 different inhaled anesthetics as BIS (A-1050 version 3.4) values decreased from 4050 to 2728 during the washout phase of 1.5%1.6% sevoflurane and the washout phase of 0.9% isoflurane (63). In all of the above-mentioned patients, the simultaneously recorded raw EEG showed a very low-voltage EEG (
15 µV) that was misinterpreted by the BIS monitor as burst suppression (62,63). Unfortunately, a normal BIS value before induction of anesthesia cannot guarantee a normal BIS performance with this form of drug-induced low-voltage EEG.
Artifact Signals During Isoelectricity
BIS electrodes are intended to detect cortical EEG signals. However, in the absence of EEG signals, as in brain dead subjects, or extreme EEG suppression of deep hypothermia, the BIS algorithm is very vulnerable to a wide range of artifact signals. This was shown in 2 confirmed brain dead subjects (56,64), in whom the BIS score increased from 05 to 38. The BIS being exactly synchronized with the subjects electrocardiography (ECG) signals indicated that the BIS algorithm misinterpreted their ECG signal as EEG activity (56,64).
A similar effect was demonstrated in a 2-mo-old patient undergoing cardiac surgery. During complete isoelectric EEG suppression of deep hypothermic (18°C) circulatory arrest, BIS (A-2000) abruptly increased from 0 to 98100, indicating radiofrequency pollution. After 10 min of rewarming, BIS returned to 40, indicating that rewarming resulted in resumption of EEG activity that enabled the BIS to discount the radiofrequency noise in the algorithm and appropriately reinterpret the EEG signal (65).
| Awareness and Explicit Recall with Different BIS Revisions |
|---|
|
|
|---|
According to the manufacturer, the latest algorithm revision (BIS-XP version 4.0) is a major revision intended to make the BIS more resistant to electrocautery and electromyography (EMG) artifacts. However, 3 volunteers, under propofol-midazolam anesthesia, were recently reported to remain responsive to verbal command at a BIS-XP value of as low as 40 (71). This indicates that the latest algorithm iteration could have resulted in the new BIS-XP providing lower values than the previous BIS models for the same hypnotic level. This may well be the case, as our research group (72) demonstrated, with 2 simultaneously placed BIS monitors, that the mean BIS-XP (version 4.0) value of 33 was significantly lower than the mean BIS (A-2000 version 3.4) value of 40. This clearly indicates that, in addition to all the above-mentioned factors, the choice of the BIS model is a major factor that would greatly influence the interpretation of the BIS value.
| Effect of EMG Activity and Muscle Relaxants |
|---|
|
|
|---|
EMG activities are artifact signals that occur within the frequency "range of interest" of the bispectrum, because the EMG frequencies overlap the BIS algorithms BetaRatio in the 30- to 47-Hz range. The EMG3047 Hz portion was estimated to be one-tenth the magnitude of the EEG3047 Hz signal (76). Contamination of BIS by EMG activity is therefore inherent in the calculation of the BIS because EMG frequencies could simulate the 30- to 47-Hz component of the BetaRatio typically associated with awake or light levels of anesthesia. This would be misinterpreted by the BIS algorithm as EEG activity and assigned a spuriously increased BIS value, making deeply anesthetized patients appear more awake than they really are. The administration of NMBD, in that case, would decrease the BIS value by alleviating the artifact and unveiling the true calculated BIS. However, in the absence of EMG activity, the BIS A-1000 version 3.3 (77) and BIS-XP version 4.0 (72) mean value of 40 was unaltered by NMBD administration under propofol TCI anesthesia.
NMBD could still influence BIS monitoring in another manner. Messner et al. (78) volunteered to receive succinylcholine while being fully awake, demonstrating the BIS (A-1000 version 3.31) artifactually decreasing to 33 and to 9 on repeated administration.
In conclusion, high EMG activity and electric device interference could create subtle artifact signal pollution without being necessarily displayed as artifacts. This would be misinterpreted by the BIS algorithm as EEG activity and assigned a spuriously increased BIS value. Numerous clinical conditions that have a direct effect on EEG cerebral function could also directly influence the BIS value.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. K. Panni and M. K. Panni Increased Anesthetic Depth, Not Metoprolol, May Have Caused the Change in Bispectral Index Anesth. Analg., February 1, 2008; 106(2): 670 - 670. [Full Text] [PDF] |
||||
![]() |
P. A. Sloan Interference of bispectral index monitoring with intraoperative use of the electromyograph endotracheal tube Can J Anesth, December 1, 2007; 54(12): 1028 - 1029. [Full Text] [PDF] |
||||
![]() |
V. Bonhomme and P. Hans Muscle relaxation and depth of anaesthesia: where is the missing link? Br. J. Anaesth., October 1, 2007; 99(4): 456 - 460. [Full Text] [PDF] |
||||
![]() |
A. Ekman, E. Stalberg, E. Sundman, L. I. Eriksson, L. Brudin, and R. Sandin The Effect of Neuromuscular Block and Noxious Stimulation on Hypnosis Monitoring During Sevoflurane Anesthesia Anesth. Analg., September 1, 2007; 105(3): 688 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Malviya, T. Voepel-Lewis, A. R. Tait, M. F. Watcha, S. Sadhasivam, and R. H. Friesen Effect of Age and Sedative Agent on the Accuracy of Bispectral Index in Detecting Depth of Sedation in Children Pediatrics, September 1, 2007; 120(3): e461 - e470. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Lauwick, M. English, and T. M. Hemmerling An unusual case of cerebral hypoperfusion detected by bispectral index monitoring Can J Anesth, August 1, 2007; 54(8): 680 - 681. [Full Text] [PDF] |
||||
![]() |
P. Tewari and H. Skinner Cardiopulmonary Bypass Machine Can Interfere with Accuracy of BIS Monitor Anesth. Analg., August 1, 2007; 105(2): 534 - 534. [Full Text] [PDF] |
||||
![]() |
P. J. Manberg and S. D. Kelley Cardiopulmonary Bypass Machine Can Interfere with Accuracy of BIS Monitor Anesth. Analg., August 1, 2007; 105(2): 534 - 535. [Full Text] [PDF] |
||||
![]() |
J. Dinsmore Anaesthesia for elective neurosurgery Br. J. Anaesth., July 1, 2007; 99(1): 68 - 74. [Abstract] [Full Text] [PDF] |
||||