JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hagberg, C.
Right arrow Articles by Khurana, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hagberg, C.
Right arrow Articles by Khurana, S.

Anesth Analg 2006;102:621-625
© 2006 International Anesthesia Research Society
doi: 10.1213/01.ane.0000189101.26403.06


GENERAL ARTICLES

An Evaluation of the Insertion and Function of a New Supraglottic Airway Device, the King LT TM, During Spontaneous Ventilation

Carin Hagberg, MD, Yefim Bogomolny, MD, Clarence Gilmore, MD, Valentine Gibson, MD, Mark Kaitner, MD, and Somya Khurana, MD

Department of Anesthesiology, The University of Texas Medical School at Houston, Houston, Texas

Address correspondence and reprint requests to Carin A. Hagberg, MD, Department of Anesthesiology, The University of Texas Medical School at Houston, 6431 Fannin, MSB 5.020, Houston, Texas 77030-1503. Address e-mail to Carin.A.Hagberg{at}uth.tmc.edu.


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 
Trials of the King LT TM (LT, King Systems, Noblesville, IN) in controlled ventilation of the lungs have shown that it is an effective supraglottic airway device. We designed this study to evaluate the King LT TM regarding ease of insertion, position within the airway, and anatomic sealing properties during spontaneous ventilation in 50 ASA physical status I–III, Mallampati I–III, adult patients undergoing routine general anesthesia. Anesthesia was induced with up to 2 µg/kg fentanyl and 1.5–2 mg/kg propofol and maintained with 70% N2O/30%O2 and isoflurane. Insertion time, oropharyngeal leak pressures, fiberoptic position, and spirometry and hemodynamic data were recorded. Any complications were noted. Insertion was determined to be easy and a patent airway was achieved in all patients. First, second, and third attempt insertion rates were 86%, 12%, and 2%, respectively. Time to place the King LT TM was <5 s in 90% of cases. Baseline leak pressures were 31 ± 8.8 cm H2O (17–50 cm H2O). Complications included laryngospasm (1) and coughing (3) on extubation. The incidence of sore throat at 1 h and 24 h postoperatively was 22% and 15%, respectively. The King LT TM is a simple and reliable supraglottic airway device for airway management during spontaneous ventilation.


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 
The King Laryngeal Tube (King LT TM; King Systems, Noblesville, IN) is a relatively new supraglottic airway (SGA) for airway management patterned after the Combitube. Similar to the Combitube, it can be inserted blindly or with the aid of a laryngoscope, yet it requires a smaller mouth opening (23 mm) (1) and its S-shape design decreases the likelihood of tracheal intubation (Fig. 1). Since its introduction into clinical practice, several models of the King LT TM have been developed. The model used in this study is an S-shaped transparent silicon tube consisting of a proximal pharyngeal cuff and a distal esophageal cuff which are both high-volume, low-pressure cuffs. Between these cuffs are two oval-shaped ventilation apertures and two secondary lateral ventilatory openings that allow ventilation to pass from the oropharynx into the trachea (Fig. 2). The proximal cuff blocks the oropharynx and nasopharynx, while the distal esophageal cuff blocks the opening of the esophagus. The King LT TM is available for neonates to adults in sizes 0–5.


Figure 156
View larger version (31K):
[in this window]
[in a new window]
 
Figure 1. The King LT TM.

 

Figure 256
View larger version (167K):
[in this window]
[in a new window]
 
Figure 2. View of ventilatory holes of the King LT TM (main oval-shaped apertures and 2 lateral ventilatory openings [arrows]).

 

A number of studies have proven the King LT TM to be effective in mechanical ventilation (1–4). However, few studies have been conducted to examine the effectiveness of the King LT TM during spontaneous ventilation. This study evaluates the nondisposable King LT TM regarding ease of insertion, position within the airway, and anatomic sealing properties during spontaneous ventilation in anesthetized patients scheduled for elective surgery.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 
After approval by the IRB and written informed patient consent, 50 ASA I–III, Mallampati I–III, patients scheduled for elective surgery in which tracheal intubation was not deemed necessary were enrolled in this study. Patients were excluded if they presented as Mallampati III–IV, ASA physical status IV–V, or emergency status. Furthermore, patients were excluded if they met one of the contraindication criteria in using an laryngeal mask airway (LMA), including: 1) obesity, 2) pregnancy, 3) history of gastric regurgitation, heart burn, ileus or "full stomach," 4) history of low pulmonary compliance or high pulmonary resistance, 5) known history of difficult intubation or signs suggesting the possibility of difficult intubation, 6) pharyngeal pathology, or 7) upper airway obstruction resulting from laryngeal pathology.

An IV was placed in each patient before transfer to the operating room. Standard routine ASA monitors were placed, and baseline measurements of arterial blood pressure, heart rate, oxygen saturation, and CO2 were recorded. Vital signs were then recorded every minute from the time of induction of anesthesia until 5 min after induction and then at 5-min intervals thereafter for 15 min. After administration of oxygen (4 L O2 for 3 min), anesthesia was induced with up to 2 µg/kg fentanyl and 1.5–2 mg/kg propofol and maintained with 70% N2O/30% O2. An appropriately sized King LT TM (a size 4 was used for patients 5–6 ft and a size 5 was used for patients >6 ft in height) was inserted by resident anesthesiologists with no prior King LT TM experience 1 min after completion of anesthetic induction and loss of eyelash reflex. With the patient’s head in the neutral position, each device was placed into the oropharynx and passed into the hypopharynx until distinct resistance was felt or the tube could advance no further.

The number of attempts required to insert the King LT TM was recorded. Insertion time (time from picking up the King LT TM until placement in the oropharynx was completed) and ease of insertion were also recorded. Both King LT TM cuffs were simultaneously inflated and cuff pressures were measured by using the manometer connected to the pilot tube of the pharyngeal cuff and were measured at 10-min intervals (Cuff Pressure Gauge; King Systems, Noblesville, IN), beginning after the introduction of the King LT TM, for the first 30 min of anesthesia and then every 30 min thereafter. Pressures were maintained at 60 cm H2O until just before device removal. Each device was taped securely in similar fashion to that of a LMA, with indirect position maintained. The depth of insertion was recorded and auscultation was made over the epigastrium to determine if there was any evidence of gastric insufflation.

An LF II Olympus fiberoptic bronchoscope (Olympus, Orangeburg, NJ) was then passed through a bronchoscopic adapter (Bodai Suction Safe TM Swively, Phase II Medical Manufacturing, Inc., Rollinsford, NH) attached to the King LT TM to visualize all 4 ventilatory orifices and their relation to the laryngeal inlet. The airway view was recorded as visualized from the distal orifice as 1 = full vocal cords, 2 = posterior arytenoids, 3 = epiglottis, and 4 = no glottis/epiglottis.

Airway leak pressures were then determined by listening for the first sign of a leak over the mouth while closing the pop-off valve and ventilating the patient’s lungs against an outlet valve pressure of 15 mm Hg (5) while adjusting the fresh gas to 6 L/min. Assisted ventilation was performed until patients began spontaneous ventilation. If ventilation occurred easily with auscultation of bilateral breath sounds, the position of the device was considered good and its use satisfactory. If placement of the device was considered unsatisfactory (poor ventilation, inadequate chest excursion, leak pressure <15 cm H2O), the King LT TM was removed and the patient’s lungs were ventilated by facemask. The King LT TM could be reinserted up to 3 times maximum. Measurements of ventilation, oxygen saturation, end-expiratory carbon dioxide, expiratory tidal volume, and peak airway pressure were recorded at both 2 and 10 min.

At the conclusion of surgery, anesthetics were discontinued and the King LT TM was removed when the patient was awake and protective airway reflexes returned. After its removal, the King LT TM was examined carefully for any evidence of blood or signs of regurgitation. All patients were interviewed at 2 h and 24 h postoperatively for the presence or absence of sore throat, dysphonia, and dysphagia.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 
Patient demographics and types of surgery are displayed in Table 1 and Table 2, respectively. A size 4 or 5 King LT TM was used in 66% and 34% of patients, respectively. Both spontaneous ventilation and a patent airway were successfully achieved in all patients. Insertion was determined to be easy, with a physical placement time of <5 s in 98% of the cases and a placement time of 5–15 s in the remaining 2% of cases. Repositioning of the King LT TM was required in 9 (18%) patients. Placement attempts were as follows: 1 = 86%, 2 = 12%, and 3 = 2%.


View this table:
[in this window]
[in a new window]
 
Table 1. Demographic Data

 


View this table:
[in this window]
[in a new window]
 
Table 2. Types of Surgery

 
The mean volume of air placed in the cuffs to attain an intracuff pressure of 60 cm H2O was 73.5 ± 5.23 mL for the size 4 and 81.5 ± 4.24 mL for the size 5 King LT TM. The minimum mean airway pressure at which gas leaked around the cuff of the device was 31 ± 8.8 cm H2O, (9-50 cm H2O). Mean expiratory tidal volumes were 253 ± 85.2 mL (100–480 mL). Adequacy of ventilation through the King LT TM was either excellent or good in most (94%) cases (Table 3). Gastric insufflation was not detected by epigastric auscultation in any patient. The median fiberoptic view was 2 (posterior arytenoids) in the distal ventilation aperture and 3 (epiglottis) in the proximal ventilation aperture. Hemodynamic stability was maintained in all subjects.


View this table:
[in this window]
[in a new window]
 
Table 3. Adequacy of Ventilation Through the King LT TM

 

The average duration of surgery was 82.5 ± 52.11 min (10–260 min). Intraoperatively, there were only single episodes (2%) of bucking, obstruction, and laryngospasm. Extubation events included visualization of blood on the device in 1 patient (2%), 1 episode of laryngospasm (2%), and 3 subjects (6%) experienced coughing. No signs of regurgitation were detected. The incidence of sore throat at 1 h and 24 h postoperatively was 22% and 15%, respectively. Furthermore, one patient experienced mild dysphagia and another mild dysphonia at the aforementioned postoperative periods.


    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 
The results of this study demonstrate that the King LT TM is a reliable device for airway management during elective surgery with spontaneous ventilation. As previously mentioned, numerous studies have been performed that demonstrate its adequacy for controlled ventilation. There is sparse data regarding the adequacy of the King LT TM for spontaneous ventilation but some studies, most notably Miller et al. (6), determined that the King LT TM was unsatisfactory for this type of ventilation. Miller et al.’s findings were based on a frequent failure rate (7 of 17 patients) secondary to loss of airway control during surgery. It was inferred that "loss of airway control" was synonymous with obstruction.

Miller et al. speculated that there were three possible reasons for the frequent failure rate in their study. One reason was operator error; the other two reasons focused on tube design and its possible role in the causation of obstructive events. They reasoned that increases in the distal cuffs’ pressures may cause the King LT TM to be pulled inward, thus causing obstruction of the laryngeal opening. They noted that deflating the cuff seemed to clear the airway, whereas inflating it thereafter increased or reproduced the obstruction. Miller et al. also theorized that it was possible that the epiglottis was responsible for physically obstructing the sole ventilation aperture, thereby contributing to the observed obstructive incidents. This form of obstruction could be more easily overcome by the ventilator during positive pressure ventilation than by spontaneous ventilation.

Miller et al.’s findings were based on a first generation model of the LT manufactured by VBM (Sulz, Germany), which did not feature a second large ventilatory aperture and two lateral ventilatory openings that are now present in the first generation King LT TM. Also, this newer model features a newly designed pharyngeal cuff that may be mechanically advantageous. The mean depth of insertion was higher than expected for each size of the King LT TM; thus the company now manufacturers this device 1 cm longer and the depth markings have been moved more proximally. This extra length allows the device to be positioned more deeply so that the proximal aperture is in alignment with the glottis, thus facilitating either spontaneous or positive pressure ventilation. The disposable version (LT-D) is currently manufactured 2 cm longer than the original King LT TM.

The infrequent incidence of obstruction/loss of airway control (1/50) demonstrated in our study is in stark contrast to Miller et al.’s study. It is inferred that the current features of this version of the King LT TM allow for fewer airway complications during its use. Consistent with our findings is a study of 100 patients mechanically ventilated with the new version of the King LT TM, in which Asai et al. (7) noted the absence of obstruction on the return of spontaneous ventilation.

With the concerns of obstruction resolved, the clinical usage of the King LT TM for elective surgical procedures is intriguing for several reasons: 1) easy insertion, 2) frequent first time placement rate, 3) increased protection of the airway from the aspiration of gastric contents, 4) blockage of the esophageal opening, thereby decreasing the risk of gastric insufflation, if positive pressure is desired, and 5) fewer complications related to sore throat, dysphonia, or dysphagia.

Regarding its ease of insertion, our study found the first-time placement rate to be 86% (98% after 2 attempts). This is consistent with the first-time placement success rates of 85%–95% found in previous studies (2,7,8). Yet there are reports of 100% first attempt success rate even with inexperienced users (14). Interestingly, 8%–33% of LMA placements require more than one attempt for proper placement (9). Moreover, other SGAs may require hyperextension of the neck for placement of the device. However, the King LT TM’s design allows for placement without such manipulation, further adding to its attractiveness.

The mean airway pressure at which gas leaks around the LMA has been reported to range from 18–20 cm H2O (± 5 cm H2O) (10). The King LT TM provided a good airtight seal in most patients (Table 2), and often there was no gas leak around the cuff at an airway pressure of 25 cm H2O. Our results are comparable to those reported by Asai et al. and others (11–13).

A known disadvantage of the LMA is its inability to protect the airway from the aspiration of gastric contents. Assuming an adequate seal, the distal esophageal cuff provides an invaluable means of airway protection. By blocking the opening of the esophagus, the King LT TM should minimize the risk of aspiration. Our study supports this concept, albeit in a small number of patients, as there were no incidences of regurgitation or aspiration.

Also, while this study focused on use of the King LT TM for spontaneous ventilation, it is foreseeable that clinical application may require transition to assisted ventilation. In these instances, the distal esophageal cuff prevents exposure of the esophagus to positive pressure ventilation, thereby decreasing gastric insufflation and minimizing aspiration risks. Only the LMA-ProSeal can offer such protection. Also, several studies have demonstrated equivalent or higher attainable ventilatory seal pressures during positive pressure ventilation with the King LT TM compared with the LMA (2,6,14), along with fewer incidences of gastric insufflation (15). Therefore, although our study demonstrates that the King LT TM is effective for spontaneous ventilation, positive pressure ventilation may also be used, if desired or necessary.

Finally, we determined the incidence of adverse perioperative events to be comparative, if not favorable, to rates involving the usage of other SGAs. In previous studies, concern of obstructive events and loss of airway control precluded the use of King LT TMs in spontaneous ventilation (6). However, as addressed earlier, these results were not duplicated in our study, possibly because of the new design. Although there was 1 case in which airway trauma (2%) occurred, the incidence of sore throat at 2 hours and 24 hours (22% and 15%, respectively) were more frequent than that reported in other LT studies, which were as infrequent as 6% (16). On the other hand, the incidence of sore throat for the LMA and Combitube are 5.8%–18% (17,18) and 25–45% (19,20), respectively.

There are several limitations of this study. First, this is not a comparative trial to other SGAs, such as the LMA-Unique. This study was simply designed to determine the efficacy of the King LT TM for spontaneous ventilation. Second, resident anesthesiologists with limited experience with either the LMA or the King LT TM were involved in the insertion of the devices. Thus, the results reflect the use of the King LT TM in novice, rather than experienced, hands.

In conclusion, the King LT TM is an effective SGA device for the spontaneous ventilation of patients undergoing elective surgery. Its unique design allows for ease of placement and advancement, minimizes the risk of aspiration, and has acceptable rates of both intraoperative and postoperative complications. Further study is warranted in a large number of patients at higher risk for regurgitation and aspiration.


    Footnotes
 
Financial and equipment support was provided by King Systems, Noblesville, Indiana.

Accepted for publication August 24, 2005.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 References
 

  1. Gaitini L, Vaida SJ, Somri M, et al. An evaluation of the laryngeal tube during general anesthesia using mechanical ventilation. Anesth Analg 2003;96:1750–5.[Abstract/Free Full Text]
  2. Brimacombe J, Keller C, Brimacombe L. A comparison of the laryngeal mask airway ProSeal and the laryngeal tube airway in paralyzed anesthetized adult patients undergoing pressure-controlled ventilation. Anesth Analg 2003;96:1535.[Free Full Text]
  3. Dorges V, Ocker H, Wenzel V, Schmucker P. The laryngeal tube: a new simple airway device. Anesth Analg 2000;90:1220–2.[Free Full Text]
  4. Cook TM, McCormick B, Asai T. Randomized comparison of laryngeal tube with classical laryngeal mask airway for anaesthesia with controlled ventilation. Br J Anaest 2003;91:373–8.[Abstract/Free Full Text]
  5. Joshi S, Sciacca RR, Solanki DR, et al. A prospective evaluation of clinical tests for placement of laryngeal mask airways. Anesthesiology 1998;89:1141–6.[Web of Science][Medline]
  6. Miller DM, Youkhana I, Pearce AC. The laryngeal mask and VBM laryngeal tube compared during spontaneous ventilation: a pilot study. Eur J Anaesthesiol 2001;18:593–8.[Web of Science][Medline]
  7. Asai T, Murao K, Shingu K. Efficacy of the laryngeal tube during intermittent positive-pressure ventilation. Anaesthesia 2000;55:1099–102.[Web of Science][Medline]
  8. Genzwuerker HV, Hilker T, Hohner E, Kuhnert-Frey B. The laryngeal tube: a new adjunct for airway management. Prehosp Emerg Care 2000;4:168–72.[Medline]
  9. Smith I, White PF. Use of laryngeal mask airway as an alternative to a face mask during outpatient arthroscopy. Anesthesiology 1992;77:850–5.[Web of Science][Medline]
  10. Asai T, Morris S. The laryngeal mask airway: its features, effects and role. Can J Anaesth 1994;41:930–60.[Web of Science][Medline]
  11. Asai T, Kawashima A, Hidaka I, Kawadri S. The laryngeal tube compared with the laryngeal mask: insertion, gas leak pressure and gastric insufflations. Br J Anaesth 2002;89:729–32.[Abstract/Free Full Text]
  12. Ocker H, Nenzel V, Schmucher P, et al. A comparison of the laryngeal tube with the laryngeal mask airway during routine surgical procedures. Anesth Analg 2002;95:1094–7.[Abstract/Free Full Text]
  13. Asai T, Kawashima A, Hidaka I, Kawachi S. Laryngeal tube: its use for controlled ventilation. Masui (Jap J Anesthesiol) 2001;50:1340–1.
  14. Asai T, Hidaka I, Kawachi S. Efficacy of LT by inexperienced personnel. Resuscitation 2002;55:171–5.[Web of Science][Medline]
  15. Miller D, Light D. Storage capacities of the laryngeal mask and laryngeal tube compared and their relevance to aspiration risk during positive pressure ventilation. Anesth Analg 2003;96:1821–2.[Abstract/Free Full Text]
  16. Asai T, Shingu K, Cook T. Use of the laryngeal tube in 100 patients. Acta Anaesthesiol Scand 2003;47:828–32.[Web of Science][Medline]
  17. Keller C, Sparr HJ, Brimacombe JR. Laryngeal mask lubrication: a comparative study of saline versus 2% lignocaine gel with cuff pressure control. Anaesthesia 1997;52:592–7.[Web of Science][Medline]
  18. Dingley J, Whitehead MJ, Wareham K. A comparative study of the incidence of sore throat with the laryngeal mask airway. Anaesthesia 1994;49:251–4.[Web of Science][Medline]
  19. Oczenski W, Krenn H, Dahaba AA, et al. Complications following the use of the Combitube, tracheal tube and laryngeal mask airway. Anaesthesia 1999;54:1161–5.[Web of Science][Medline]
  20. Gaitini LA, Vaida SJ, Somri M, et al. Fiberoptic-guided airway exchange of the esophageal-tracheal Combitube in spontaneously breathing versus mechanically ventilated patients. Anesth Analg 1999;88:193–6.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hagberg, C.
Right arrow Articles by Khurana, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hagberg, C.
Right arrow Articles by Khurana, S.


Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2006 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press