From the Division of Anesthesiology, The Cleveland Clinic Foundation, Cleveland, Ohio.
Address correspondence and reprint requests to Bala Gopakumaran, PhD, E30, Division of Anesthesiology, The Cleveland Clinic Foundation, 9500 Euclid Ave., Cleveland, OH 44195. Address e-mail to nairb{at}ccf.org.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Over the last decade, there have been considerable advances in signal processing techniques, particularly for biomedical signals. Many of these techniques, particularly those using nonlinear filters and artificial intelligence, have the potential to significantly improve artifact detection and elimination. In fact, the adoption of these techniques could revolutionize the way patient monitors process wave form signals, leading to more accurate displays of clinical information. In this article we review the clinical problem of artifacts and the artifact filtering methods adopted by currently marketed patient monitors. We also review artifact correction methods that have been recently developed by the research community. Possible approaches that the new generations of patient monitors could adopt to detect and eliminate artifacts are also proposed. Only artifacts that are related to technical and environmental factors are considered in this review. Artifacts that have a physiological origin are beyond the scope of the review, and are not considered.
| ORIGIN OF COMMON ARTIFACTS |
|---|
|
|
|---|
|
| METHODS USED IN PATIENT MONITORS |
|---|
|
|
|---|
Only a few manufacturers have developed some specialized schemes to remove specific types of artifacts. They are sometimes implemented as enhanced features of the monitor that the user can select. However, in many cases, such features are never enabled because of either user ignorance or because the user is not confident of the abilities of such algorithms.
The steps taken by industry to eliminate specific artifacts in the main signals of the patient monitor are described below.
ECG Signal
Most monitors use some method to minimize the effect of electrosurgical unit (ESU) interference. These methods comply with American Association of Medical Instrumentation (AAMI) standards and generally use band-pass filters to minimize the effects of ESU noise. Most monitors keep the pass-band of such filters constant, but there has also been an attempt to adaptively vary the pass-band (Ivy Biomedical Systems) if ESU interference is detected. ESU interference is detected by monitoring the "slew rate" (the first differential of the signal). With this scheme, rapid fluctuation in the ECG signal caused by ESU interference causes the measured slew rate to be above a threshold. When ESU interference is detected, the pass-band of the filters is reduced, so that ESU interference is minimized without excessively distorting the primary ECG characteristics. Such a step will ensure that the heart rate parameter is computed with more reliability.
Interference from power sources could also affect ECG signals. To eliminate artifacts from power-line frequency, the monitors use notch filters to remove the frequency component specific to power-line interference.
None of the monitors use any artifact correction scheme specifically intended to eliminate baseline wander or motion artifacts. The effect of baseline wander is minimized by using a filter to remove low frequency components (generally below 0.5 Hz).
Invasive Arterial Blood Pressure Signals
Most models of patient monitors do not use any schemes to correct specific artifacts in arterial blood pressure signals. In general, simple band-pass filters are used to limit the effect of artifacts that could affect the computation of systolic, diastolic, and mean blood pressures. One manufacturer (GE Medical Systems) has used a "Smart BP" feature to automatically detect artifacts caused by zeroing and flushing of the pressure line or by drawing blood samples. Detection of these artifacts is performed through identifying patterns in the blood pressure signals that are created by each of the three sources of artifacts. The patterns represent the general trends in systolic, diastolic, and mean blood pressure values when a zeroing, flushing, or blood draw is performed. These patterns are detected by simple signal processing steps and heuristic rules. The "Smart BP" feature tested on a limited data set (18) seems to detect aforementioned artifacts accurately and consistently. However, further testing of this feature is necessary to fully assess its true accuracy. Moreover, the basic assumption of the algorithm that the patterns in blood pressure wave forms due to the aforementioned artifacts remain constant might be flawed. This could lead to artifacts being detected with low sensitivity and specificity.
Damping (overdamping or underdamping) of pressure signal is a common source of artifacts that could distort the pressure wave form. This could lead to over- or underestimation of pressure parameters. The current patient monitors do not use any technique to detect or compensate for this problem.
To the best of our knowledge, the currently marketed patient monitors do not use any specific algorithms to detect and eliminate specific artifacts that affect PA and CVP.
Noninvasive Arterial Blood Pressure
Most patient monitors use the oscillometry method to measure noninvasive arterial blood pressure (NIBP) (1,19,20). Though this method is immune to ambient noise, motion artifact is a major potential source of error (1,14,2022). Irregularities in the oscillometric pulses are detected by most monitors, and if a pulse search fails, arterial blood pressure measurement is reattempted. To minimize the effect of motion artifacts, most monitors use techniques to sense oscillometric pulses in synchrony with the patients pulse (from Spo2 signal) or ECG. An example is the Dynamap technology (14,23) used by GE monitors that uses stepped deflation to identify pulses matched in frequency and amplitude. This algorithm has been improved (Dynamap SuperStat) to use a modified Gaussian curve to fit the oscillometric data, thus minimizing measurement time and the effects of artifacts (24). Similarly, the "Smart Cuff" technique (Protocol Systems, Beaverton, OR) directly synchronizes ECG and NIBP data to eliminate noncardiac pulses related to artifacts. An evaluation of the Smart Cuff algorithm has shown promising results (21,22). To improve the accuracy of NIBP measurements, additional steps have been adopted. An example is the use of a two-tube system (Dynamap) that uses separate tubes to inflate and sense pulsations, thus minimizing pressure damping effects, particularly when measuring low arterial blood pressures. Another manufacturer (Nihon-Kohden, Tokyo, Japan) uses a parameter indicative of changes in arterial blood pressure to trigger NIBP measurements, thus automatically initiating more frequent NIBP measurements when the arterial blood pressure is changing rapidly. Though all manufacturers comply with American Association of Medical Instrumentation (AAMI)/American National Standards Institute (ANSI) or European standards for NIBP measurements, there is still wide variability in measured values when using different monitors, or even when using different algorithms used by the same manufacturer (25,26).
Radial tonometry is an alternative to the oscillometry method and can measure arterial blood pressure more frequently. Additionally, this noninvasive method can provide an averaged tracing of the arterial wave form. However, this technique is sensitive to motion artifact and position of pressure sensor (27). To the best of our knowledge, currently marketed patient monitors do not use this technique, though stand-alone pressure measurement devices using this technique have been on the market for some time (27).
Pulse Oximetry
As described previously, motion artifact is the main type of artifact that affects pulse oximetry. Almost all pulse oximetry manufacturers have developed their own methods to minimize the effect of motion artifacts. The attempts include using adaptive noise cancellation (Masimo, Irvine, CA), frequency-based fourier artifact suppression (Phillips Medical Systems), combinations of time and frequency domain analysis and fuzzy logic (Nellcor, Pleasonton, CA). These methods have been shown to have varying success on the basis of the measurement type and situations (28). The patient monitors in general do not directly perform the pulse oximetry measurements but, rather, interface with measurement modules supplied by the pulse-oximeter manufacturer. Hence the patient monitors act as passive display devices as far as the Spo2 signal is concerned. They simply display the O2 saturation wave form and the Spo2 parameter value that are detected and computed by the pulse oximetry hardware module. For this reason, specific methods to correct artifacts in pulse oximetry data are generally not implemented in patient monitors.
Capnography
Most capnography modules that interface to patient monitors use an infrared spectrography technique of gas analysis (15,16,29,30). These modules generally use a side-stream method to analyze gas samples aspirated from the breathing circuit (15,31,32). Moisture and secretions entering and clogging the breathing circuit and/or sampling line are common problems, although most monitors can detect this scenario and generate alarms (15,31). However, interference due to the presence of inhaled anesthetics, nitrous oxide, and high concentrations of O2 could potentially produce erroneous EtCO2 values (15,17). Monitors, in general, do not detect or correct for such interference. However, with recent monitors that use Raman scattering technology, this is less of a problem. Though Raman spectrography can measure CO2 faster and with greater accuracy than infrared technology, it is comparatively more expensive (15,16).
Temperature
Most monitors do not detect or alarm when a temperature probe is dislodged from its intended measurement position. Similarly, artifacts in the PA catheter temperature measurements are not directly detected. However, if this leads to failure in cardiac output measurement, monitors generally generate error messages.
| METHODS PROPOSED BY THE RESEARCH COMMUNITY |
|---|
|
|
|---|
The ECG
A survey of the research literature revealed active research in developing methods that eliminate artifacts in the ECG signals. However, most of the efforts are geared towards improving ECG quality in nonoperating room (OR) environments, such as with ambulatory and Holter ECG recordings. Methods based on signal extraction (3,33,34) as well as artifact elimination (4,3537) have been tested.
An example of the first method was the use of P, QRS, and T-wave frequency components (3) to differentiate the various features of the ECG from noise. A sample template of the ECG wave form was constructed using averaging techniques applied over several beats. Using the sample template, the frequency components of the ECG were derived and used to specify the pass frequency band of filters. However, the earlier application of such a technique did not adapt to the variation in the ECG signal among different subjects, and sometimes even in the same subject (5). Also, the technique failed to consider that the frequency spectrum of the ECG signal often overlaps with the spectrum of noise. As an improvement, in later studies (38), linear band-pass filters were modified to compensate for the nonstationary nature of the ECG signals. The newer filters had the capability to adapt to the changes in the ECG signal so that their operational parameters were based upon some feedback information. The feedback information is, in general, the difference between an estimate of the signal and the measured signal. The filter parameters were adjusted such that the difference between the estimated and the measured signals was forced to a minimum. This approach is used in techniques such as least mean square, recursive least square, and matching pursuit (39) to improve the accuracy of the adaptive filters. However, even these improved methods did not change the filters linear characteristics, and hence were still unreliable in dealing with the nonlinearity of the ECG noise characteristics. To accommodate nonlinearity, more recent ECG processing techniques have adopted methods of artificial intelligence (3). Artificial neural network is one such artificial intelligence method that is inspired by the way biological nervous systems, such as the brain, process information, and comprises a set of interconnected processing nodes that simulate neurons. Artificial neural network has the ability to successfully learn the embedded characteristics of the ECG signal and its inherent nonlinearity (3). However, the significant computational cost and processing time required by the algorithm has prevented its real-time application. Considerable research has also been done to extract ECG features by using wavelet filters (4044). Wavelet transform decomposes the signal into multiple scales while preserving the temporal features of the signal. The scale functions are further analyzed to extract ECG features successfully (41,42) from noisy and nonstationary signals. Recently, the accuracy of ECG feature extraction has been improved by using a wavenet technique that combines wavelets with artificial neural networks (45,46).
Methods based on artifact elimination have used different types of adaptive filters to extract artifact signals (4,39). As an example, Aase et al. (4) used a multichannel Wiener adaptive filter to extract the cardiopulmonary resuscitation (CPR) artifact. A Wiener filter operates in the frequency domain to extract the noise signal by statistical means. However, this method suffered from computational complexity and was impractical for real-time application. In later studies (39), a matching pursuit algorithm was used to provide an efficient mechanism for detecting the CPR artifact. Using a moving average filter is another example of an artifact subtraction method to detect motion artifacts (47). In this approach, a least mean square method was used to adapt the noise filter parameters to accommodate the varying artifact situations. The artifact extraction and subtraction method generally requires several filters to extract the different types of artifact. Because ECG artifacts have various origins, it is not realistic to extract every type of artifact and then subtract it from the main ECG signal.
Invasive Blood Pressure and NIBP
Most of the current research work (35,48) focuses on artifacts, particularly motion artifacts that affect NIBP measurement. Initial attempts (48) to deal with motion artifacts were based on averaging techniques to minimize the fluctuations due to artifacts. Artifact elimination was improved by applying adaptive techniques, such as Kalman filtering (35), that was superior in dealing with time varying oscillation amplitudes and stochastic noise. The Kalman filter operates in time domain and recursively estimates a signal based on the previous estimate and the current measurement. The feasibility of artificial intelligence methods such as fuzzy logic has been studied successfully (6) to remove artificial interference and reconstruct the oscillation amplitude pattern. Though active research is conducted to minimize NIBP artifacts, our literature search revealed that very few attempts have been made to eliminate artifacts affecting the ART signal. A particular attempt by Nagai and Nagata (49) is worth mentioning. They applied different digital filters to detect systolic and diastolic arterial blood pressure values. Artifact values were eliminated by applying Smirnovs rejection method, which is based on finding of outlier blood pressure values when the standard deviation of blood pressure values was more than 5%. The presence of outlier values was used as a trigger to eliminate segments of blood pressure values. Though this technique was successful in eliminating artifacts with reasonable accuracy, it adopted a very conservative approach and eliminated several nonartifact data points along with actual artifactual points. An interesting approach by Zong et al. (50) used fuzzy logic to analyze the relationship between ECG and arterial blood pressure wave forms to assess blood pressure signal quality and artifacts. Initial evaluation of this method in intensive care units (ICU) demonstrated that arterial blood pressure artifacts can be detected with high accuracy and specificity.
Pulse Oximetry
The pulse oximetry device evaluates blood photoplethysmographic (PPG) characteristics and detects hypoxia when the blood is not well perfused. The PPG can be corrupted by motion artifacts, ambient light interference, and low blood perfusion. Several algorithms are described in the literature that aims to eliminate artifacts in the Spo2 signal (7). Generally, these algorithms identify the artifacts by analytical means, through either spectral analysis of the signal or statistical correlation with other transducer signals. As an example, Hall (51) correlated the occurrence of a true blood pulse to the onset of cardiovascular activity, which was determined by observing the R wave in the QRS complex of an ECG signal. A microprocessor calculated the oxygen saturation using a preset formula and read the result from an experimentally determined reference table. This approach predominantly characterized the incoming PPG signal and rejected any part of the signal that did not fit the prior assumed signal characteristics, and the preset formula implicitly discriminated artifact components. Other approaches (7,52) used an artifact reduction method, where an additional measurement channel is used for deterministic removal of the modeled form of the motion artifact. The techniques described in the literature show promise in minimizing artifacts, but extensive studies evaluating their accuracy in a clinical environment are yet to be performed.
Quite recently, an entirely different way of improving the quality of the data extracted from physiological signals has been suggested. The method uses multisensor data fusion. For instance, Ebrahim et al. (53) and Feldman et al. (54) investigated the notion of combining sensors with similar information to improve the quality and reliability of the extracted data. Thus, they sought to develop a method for combining heart rate measurements from multiple sensors to obtain: 1) an estimate of heart rate that is free of artifact; 2) a confidence value associated with every heart rate estimate which indicates the likelihood that an estimate is correct; and 3) a more accurate estimate of heart rate than is available from any individual sensor. The raw signals used to estimate heart rate were the ECG, the arterial blood pressure signal, and the pulse oximetry signal. Their method sought to discriminate between high-quality and low-quality sensor signals and combine only the high-quality signals to derive an improved heart rate estimate. Past estimates of heart rate were used to derive a predicted value for the current heart rate that was mathematically "fused" with the sensor measurements. The level of consensus between the sensor measurements, the predicted value, and the physiologic credibility of the readings was used to distinguish between high-quality and low-quality readings. The authors stated that the method performed well using clinical data, noting that the fused heart rate estimate "was consistently as good or better than the estimate available from any individual sensor" and that the fused estimates "consistently reduced the incidence of false alarms compared with individual sensors without an unacceptable incidence of missed alarms."
Capnography
Few attempts are being made to improve detection and elimination of artifacts by better processing of capnography signal. However, improvements in capnography measuring technique and design of airway adapters have minimized the occurrences of some common artifacts. Molecular correlation spectroscopy, an improvement over conventional infrared spectroscopy, uses an infrared emission to precisely match the absorption spectrum of CO2 [Oridion, Needham, MA (16,55)]. This method not only improved the measurement accuracy, but also minimized the interference because of the presence of inhalations agents and nitrous oxide. Filterline airway adapters (Oridion) minimize the chance of sample line blockage due to fluids by using three channels that faces different directions. With this adaptor, aspiration of sample can take place even if only one channel is open. The use of Naflon to make the sampling tube also minimizes condensation of water vapor in the breathing circuit (16). Naflon selectively allows water vapor to pass from the interior of the tube to the exterior.
| DISCUSSION |
|---|
|
|
|---|
To minimize the problem of artifacts, several measures can be adopted. These measures can be considered from clinical and engineering perspectives. The clinical measure should be toward adopting monitoring practices that minimize the chance of artifacts affecting measurement of biomedical signals. The engineering measures would be two-fold: 1) improving the design of biomedical sensors so that they are less tolerant toward artifacts; and 2) improving signal processing methods so that artifacts can be accurately detected and filtered. From the patient monitor design perspective, it is the second engineering measure that is the most relevant. Artifact detection and correction is a two-step process. The initial step is artifact detection, in which it is determined whether the acquired biomedical signal is corrupted by artifact. This is followed by the step of signal extraction, in which the signal of interest is extracted from the acquired signal with the artifact portion removed. The success of the second step will be determined by the quality of the signal extraction technique as well as the quality of the acquired signal. If the signal quality or signal-to-noise ratio is so low that signal extraction is neither possible nor reliable, the monitor should take the necessary steps not to process the signal, and thus not to provide erroneous parametric data. Additionally, the patient monitor should take the necessary steps to alert the clinician to the presence of artifacts so that necessary clinical steps can be taken to eliminate or minimize them. Monitors now on the market do perform the above tasks, although in a simplistic manner. Detection of artifacts is generally performed by measuring rapid and large fluctuations in the signal. Signal extraction or artifact removal is typically performed by applying linear filters. Though the techniques adopted by most patient monitors can detect obvious artifacts, they have limited success in detecting a wide variety of artifacts encountered in the intraoperative environment. Additionally, the simpler filtering techniques used in current monitors have been in existence for decades and are not ideally suited to filter time-varying and complex biomedical signals, and therefore their ability to extract the actual signal from one contaminated by artifacts is severely limited.
There is much room for improvement in the artifact detection methods adopted by current patient monitors. This is especially true considering that in recent years, there has been considerable progress in biomedical signal processing techniques and the computational power of microprocessors. These technological advancements should be used to provide more reliable and accurate artifact detection and signal extraction schemes. The new generation of patient monitors could possibly detect artifacts using a combination of artificial intelligence and signal processing techniques (63). The techniques of artificial intelligence, such as fuzzy logic, artificial neural networks, and genetic algorithms are designed to duplicate the human thought and decision-making processes. Such techniques could duplicate the decision-making process of a clinician when deciding whether a parameter value contains artifact. In the OR setting, the clinician decides whether to discriminate a parameter value as an artifact on the basis of several factors. The primary factor is the quality of the signal (or wave form) from which the parameters are derived. Inferior quality wave forms with irregular fluctuations or wave forms that show abnormal or nonphysiological patterns are obvious indicators that the parameters derived from such wave forms are artifacts. In such cases, the clinicians can select the parameters from an alternate source, if one is available. As an example, when the ECG signal is distorted by artifacts, the clinicians could choose to use the heart rate value derived from the blood pressure or pulse oximetry signals. Clinicians sometimes use additional pieces of information that are provided by the signal acquisition module, to make decisions about artifactual data. For example, some pulse oximetry manufacturers provide a signal quality index along with the oxygenation signal. During periods when the signal quality index is low, the clinician may ignore the Spo2 parameter value as artifactual. Occasionally, parameters containing artifacts are not obvious by simply qualitatively analyzing the source wave form. In such cases, clinicians might choose to refer to alternate wave forms or sources of information to judge whether a parameter is an artifact. For example, double counting of ECG complexes can be deduced by referencing the blood pressure and pulse oximetry wave forms, and the heart rate derived from these wave forms. Another clinical example occasionally encountered is the erroneous display of Spo2 values when the peripheral perfusion is poor. This situation can be deduced by gauging the pressure wave form and the cardiac output values. Artificial intelligence techniques can mimic the aforementioned human thought process. These techniques can take advantage of the fact that the patient monitor serves as a central source, where multiple signal information derived from a variety of biomedical signals is available. Additionally, such techniques can also use the underlying factor that the ECG, arterial blood pressure, and pulse oximetry signals are all generated by the beating of the heart. Hence there is a fair amount of information overlap among these signals, with the obvious common parameter being the heart rate. However, these three signals are acquired by different transducers: ECG-electrical, blood pressure-mechanical, and pulse oximetry-optical. Different signal transduction methodologies mean that there is little chance of common sources of artifact that affect these signals simultaneously.
In addition to using artificial intelligence methods, better signal processing methods can also be adopted by patient monitors to deal with artifacts. The use of better signal processing methods can help in two ways. First, it can provide more accurate information to the artificial intelligence module so that it can make better decisions on classifying signals as artifacts. Second, it will be able to efficiently and accurately extract the signal of interest from an artifactual signal. Most signal processing filters used in the existing patient monitors are linear and based on frequency domain designs. However, in reality, not only are the biomedical signals contaminated by nonlinear sources of noise, but also their frequency characteristics change over time. Several nonlinear and adaptive filtering techniques, such as the Kalman and H
filters, have been developed in recent decades and have been shown to have superior and optimal performance when removing nonlinear noise. Wavelet filtering is another option that patient monitors can use. Such filters, based on both time and frequency domain designs, are especially suited to filter signals whose frequency characteristics vary over time.
While dealing with artifacts, the primary task of the patient monitor is the detection and possible elimination of artifacts. However, many times, the signal is severely corrupted by artifacts such that extraction of reliable physiological information becomes impossible. Under these conditions, the patient monitor should be able to switch to the secondary task of alerting the clinician about the presence of the artifact. A message describing the possible source of artifact should be displayed. Additionally, the parametric data derived from the source signals should be displayed in a different format (different background color) to notify the clinician that the displayed values are old and that computation of current values cannot be performed because of poor signal quality. Most current patient monitors perform the above tasks to some extent; however, their ability to alert clinicians to the presence of artifacts is rather simplistic, probably because of the limited types of artifacts that are detected by current patient monitors.
In summary, our review found that artifacts remain a significant problem for processing and displaying correct clinical information by OR patient monitors. Most of the current patient monitors still use simple linear filters which are often ineffective to detect and eliminate artifacts. The new generation of patient monitors should adopt improved methods of artifact detection, elimination, and alerts. These methods could be based on artificial intelligence and advanced signal processing methods that take advantage of the fact that the patient monitors serve as a central source for multiple signals that have a fair amount of information overlap.
| Footnotes |
|---|
| REFERENCES |
|---|
|
|
|---|
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|