| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
BACKGROUND: The AnaConDa® filter permits administration of volatile anesthetic without the use of an anesthesia machine. It is intended for use in the intensive care unit. METHODS: We studied the AnaConDa® reflection filter on the bench and in anesthetized patients. The bench analysis used a test lung, a gas analyzer, an intensive care ventilator, the AnaConDa® filter, and a syringe pump. We studied a range of tidal volume, respiratory rate, and positive end-expiratory pressure values. We simulated errors during syringe refilling and patient transportation. In 15 anesthetized patients, we used the AnaConDa® with constant ventilation variables, a constant sevoflurane infusion rate (45 mL/h), and two consecutive fresh gas flow levels. RESULTS: In the bench study, the expired volatile anesthetic fraction decreased linearly with respiratory frequency at constant minute ventilation, and decreased markedly in a hyperbolical manner when tidal volume increased at a constant respiratory rate. Changing the positive end-expiratory pressure level and inspiration/expiration ratio did not modify the AnaConDa®s performance. Several safety failures were observed: refilling caused a transient change in AnaConDa® output because of a pumping effect, and a standard Luer lock made it possible to connect the halogenate syringe on an IV infusion line. In anesthetized patients, reducing fresh gas flow from 8 to 1 L/min led to a median 40% increase in the expired volatile anesthetic fraction. CONCLUSIONS: This study shows that the device is generally reliable, but that there are several conditions under which it might deliver more anesthetic than intended.
Inhaled volatile anesthetics are useful drugs in the intensive care unit (ICU) for the treatment of status asthmaticus (1,2) and for providing fast reversible sedation (3). However, their administration traditionally requires a cumbersome closed-circuit anesthesia machine. A novel approach to administration of inhaled volatile anesthetics is the use of a reflection filter in the breathing circuit. The reflection filter is a device connected between the patient and ventilator to allow reinhalation of anesthetics. These filters conserve volatile anesthetics in a manner similar to conservation of heat and moisture by a heat and moisture exchanger. Reflection filters containing zeolite as the absorber were introduced in the 1980s (4). However, studies suggesting that zeolite fibers might induce fibrosis and cancer (57) precluded widespread adoption. A new generation of reflection filter (AnaConDa®, Hudson RCI, Uppsland Väsby, Sweden) with charcoal as the absorber and an imbedded heat and moisture exchanger was recently introduced for use in the ICU (7,8). The device allows infusion of liquid volatile anesthetics by a syringe pump via a porous rod on the patient side of the filter. Unfortunately, the manufacturer provides only limited safety and efficacy data on the AnaConDa® filter. No studies have evaluated the influence of ventilatory settings on the safety and efficacy of the device. The main purpose of our study was to conduct a bench evaluation of AnaConDa® to obtain safety data and to investigate the influence of ventilator settings on performance. In addition, we hypothesized that AnaConDa® used with a low-flow closed-circuit anesthesia machine would reduce volatile anesthetic consumption during anesthesia for surgery. We tested this hypothesis in patients undergoing anesthesia for surgical procedures.
Bench Study The bench apparatus consisted of a Servo 900D ICU ventilator (Siemens Elema®, Solna, Sweden). An 8-mm endotracheal tube was used to connect the ventilatory circuit to a lung model made of a plastic jar partly filled with water to produce a compliance of 50 mL/cm H2O. The AnaConDa® filter (Figs. 1 and 2) was placed between the Y piece and the endotracheal tube. Air pollution by the volatile anesthetic was prevented by a charcoal-containing cartridge connected to the gas outlet of the ventilator. An infrared gas analyzer (Brüel and Kjaer, Bron, France) was used to sample gas via the sampling port of the AnaConDa® filter to measure the expired fraction of volatile anesthetic (FE). Sampled gas was reinjected from the analyzer outlet into the breathing circuit. FE was recorded every 15 s on a personal computer. The analyzer was calibrated before the study, as recommended by the manufacturer.
AnaConDa® Filter Output AnaConDa® Filter Output at Step Increasing Infusion Rate, Under a Standard Ventilatory Setting. In this technique, isoflurane was infused into the filter via a syringe pump (Fresenius Vial, Pilot C®, France) at step fixed rates (1, 2, 4, 6, 8 mL/h), while ventilation was maintained constant (tidal volume (VT): 500 mL, respiratory rate (RR): 12 min1, inspiratory time/total respiratory cycle time (Ti/Ttot): 33%, constant inspiratory flow, and no positive end-expiratory pressure (PEEP).
Assessment of the AnaConDa® Efficacy and the Influence of Respiratory Frequency Under Three Different Minute Ventilation Regimen. To assess the FE without the reflecting filter, we also tested an AnaConDa® device from which we had removed the reflecting filter under the conditions described earlier.
Assessment of the Influence of VT. The measurement error did not exceed ±2%, as assessed by performing runs in triplicate.
Test of Two Scenarios that Might Lead to Anesthetic Overdose
Risk of Halogenate Accumulation.
Patient Study
Bench Study As shown in Figure 2, the relationship between infusion rate and isoflurane FE was linear when ventilator settings were maintained constant. When the infusion rate was kept constant (5 mL/h), the output of volatile anesthetic from the AnaConDa® filter varied with the ventilator settings. Increasing RR with a constant minute ventilation caused an increase in FE (Fig. 3a). The efficacy of the filter is illustrated by the fact that FE is about eight times lower when charcoal has been removed. Increasing Vt with a constant RR caused a marked nonlinear decrease in FE (Fig. 3b). In this figure, one can see that Vt seemingly plays a larger role than breathing frequency in the AnaConDa®s performance.
Isoflurane output did not vary with PEEP or when Ti/Ttot was changed. During syringe refill, gravity-induced flow of anesthetic liquid caused an increase in FE when the syringe was disconnected 30 cm above the filter, and reverse flow caused a decrease in FE when disconnection was done 30 cm below the filter (Fig. 4). The 1 h disconnection while leaving the syringe pump led to a large, transient increase in FE because of the accumulation of volatile anesthetic in the filter (Fig. 5).
Patient Study
Our main findings are that the vapor output of the AnaConDa® varies with the anesthetic infusion rate and also with the ventilator settings, and that, in patients, the device reduced volatile anesthetic consumption in a closed-circuit system. Our study also identified a number of safety concerns. We observed on bench that ventilation settings influenced the output of the AnaConDa®. Vt especially was the most influencing variable on FE. The increasing effect of dead space may explain the hyperbolic increase of FE when Vt decreased. Conversely, at high Vt, one may observe a full washout of the filter with FE converging to an asymptotic value. We hypothesize that the small dependence of FE on RR at constant Vt may result from slightly different isoflurane mixing conditions in the filter. The first risk we identified, although not specifically studied, as it is obvious on inspection of the equipment, is that of an inadvertent connection of the volatile anesthetic syringe to an IV infusion line. The Luer lock-fitted volatile anesthetic infusion line has the same appearance as a standard infusion line even if the syringe is labeled in color with a printed warning "not for IV use". At present, no safeguard is available to avoid such an error. Our bench study, although its results cannot be directly applied to clinical use, suggests that clinicians should be mindful of the possibility that changing ventilator settings may alter inhaled anesthetic delivery. Our clinical study should be interpreted in the light of the data obtained by Enlund et al. (8). These authors compared isoflurane consumption in patients receiving isoflurane to maintain anesthesia at a FE around 0.46%, either via a conventional vaporizer or using the AnaConDa® filter. Ventilation was provided by a Mapleson D system configuration with FGF at about 2.2 L/min. The AnaConDa® reduced isoflurane consumption by 39% via a 55% reduction in the isoflurane escape rate to the atmosphere (8). In our study, the AnaConDa® filter with FGF at 8 L/min required the same amount of sevoflurane as observed by Tempia et al. (around 5 mL/h) (9) in the absence of the filter with the FGF at 1 L/min. The FE achieved in both studies was similar 0.8% [range: 0.51.0] in ours and 1.1% ± 0% in that of Tempia et al. Thus, the AnaConDa® filter performs in an open circle configuration (FGF at 8 L/min) equivalently to a conventional closed circuit configuration with a FGF set at 1 L/min. However, because the median increase of the filter output was 40% [range: 2067] when FGF was set at 1 L/min, using the AnaConDa® on a closed-circuit machine with a low FGF produced greater volatile anesthetic conservation than reported in previous studies (8,9). Connecting the AnaConDa® to a closed circuit with a FGF of 1 L/min should improve the volatile anesthetic-conserving properties of a closed circuit anesthesia. The AnaConDa® filter may be useful in ICUs if the safety issues are resolved. Once the risks are addressed, the device may be useful for sedation and bronchodilation in mechanically ventilated patients. A randomized crossover (10) study in ICU patients showed that propofol- or isoflurane-based long-term sedation aiming at a predefined sedation level led, on average, to similar times to awakening after sedative discontinuation. Thus, recovery time is not a major advantage of a volatile anesthetic for sedation. In contrast, bronchodilation may be a more specific goal AnaConDa® may achieve in intensive care. However, if the device is used for long term administration of volatile anesthetics, the risk of inorganic fluoride accumulation must be addressed, as this does not seem to cause detectable renal toxicity in humans (1113). A scavenging system will be needed to prevent pollution of ambient air. A charcoal cartridge effectively prevents pollution (14). We believe this is likely useful, although others have not found high levels of pollution when using the AnaConDa® with no scavenging system (15). In conclusion, the performance of the AnaConDa® varies with ventilator settings in a predictable manner, at least during bench testing. The AnaConDa® adds to the anesthetic-sparing effect of conventional low-FGF closed-circuit anesthesia machines, and may be of use during administration of volatile anesthetics in the operating room. However, there are safety issues that should be addressed with the current device.
Accepted for publication September 21, 2006. Supported by Pôle dAnesthésie Réanimation. Author correspondence and reprint requests to Laurent Beydon, MD, Pôle dAnesthésie Réanimation, CHU dAngers, 49933 Angers Cedex 09, France. Address e-mail to lbeydon.angers{at}invivo.edu.
This article has been cited by other articles:
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|