BACKGROUND: "Optimal" positive end-expiratory pressure (PEEP) can be defined as the PEEP that prevents recollapse after a recruitment maneuver, avoids over-distension, and, consequently, leads to optimal lung mechanics at minimal dead space ventilation. In this study, we analyzed the effects of PEEP and recruitment on functional residual capacity (FRC), compliance, arterial oxygen partial pressure (Pao2) and dead space fraction, and we determined the most suitable variables indicating optimal PEEP. METHODS: We studied 20 anesthetized patients with healthy lungs undergoing faciomaxillary surgery. After a stepwise increase of PEEP/inspiratory pressures (0/10, 5/15, 10/20, 15/25 cm H2O, each level lasting for 20 min) using a pressure-controlled ventilation mode, a recruitment maneuver (at 20/45 cm H2O for a maximum of 20 min) was performed, followed by a stepwise pressure reduction (15/25, 10/20, 5/15, 0/10 cm H2O, with 20 min at each level). At each pressure level, FRC, compliance, Pao2, and dead space fraction were measured. RESULTS: When comparing the values before and after recruitment at identical PEEP levels, all variables showed significant changes at 10/20 cm H2O; compliance was also significantly higher at the pressure step 15/25 cm H2O. In addition, FRC values showed significant differences at 5/15 cm H2O and 15/25 cm H2O. CONCLUSIONS: All variables showed the positive effects of PEEP in conjunction with a recruitment maneuver. Optimal PEEP was 10 cm H2O because at this pressure level the highest compliance value in conjunction with the lowest dead space fraction indicated a maximum amount of effectively expanded alveoli. FRC and Pao2 were insensitive to alveolar over-distension.
General anesthesia affects ventilation and gas exchange by causing atelectasis.1–4 Various studies in patients with healthy lungs undergoing general anesthesia have demonstrated that recruitment maneuvers remove atelectasis, improve lung mechanics and gas exchange, and reduce dead space fraction.5–10 If positive end-expiratory pressure (PEEP) was applied after recruitment, it could prevent recollapse of alveoli.9,11,12 However, PEEP levels should not be too high, to avoid over-distending the lung. The PEEP, which prevents re-collapse, but avoids over-distension while optimizing lung mechanics at minimal dead space ventilation, can be defined as optimal PEEP. However, the question of which variables reliably determine this optimal PEEP remains unanswered. We hypothesized that functional residual capacity (FRC), compliance, arterial oxygenation, and dead space fraction should show the positive effects of PEEP after a recruitment maneuver, and compliance and dead space fraction are the two most suitable variables for determining optimal PEEP. Therefore, the aims of this study were 1) to evaluate the effects of PEEP on FRC, compliance, arterial oxygenation, and dead space fraction before and after a recruitment maneuver in healthy patients undergoing general anesthesia, 2) to decide which of the above variables indicates optimal PEEP, and 3) to determine optimal PEEP for this group of patients.
The study was conducted at the University Hospital of Hamburg, Germany, after approval was obtained from the local ethics committee and written informed consent from each patient. Twenty patients were studied: 12 men and 8 women with an age of 41 ± 16 yr (mean ± sd) and a body mass index of 24 ± 4 kg/m2. Nine of the patients were smokers. The patients had no relevant cardiopulmonary disease and were classified as ASA I (5) or II (15). They were studied during general anesthesia for elective faciomaxillary surgery. Surgery in the supine position had to have an expected duration of general anesthesia of at least 3.5 h. During these operations, the surgeon did not influence the thorax or position of the diaphragm.
Anesthesia, Ventilation, and Monitoring Lungs were ventilated with an EVITA 4 (Drägerwerk AG, Lübeck, Germany) in a pressure-controlled ventilation mode (bilevel positive airway pressure) and with an inspired oxygen fraction (Fio2) of 0.3. Ventilation started with a PEEP of 5 cm H2O and an inspiratory pressure (IP) of 15 cm H2O. Respiratory rate was set between 10 and 14 breaths/min, inspiratory time was set at 2.0 s, and expiratory time was maintained between 2.3 and 4.0 s. The respiratory rate was regulated according to arterial carbon dioxide partial pressure (Paco2), which was kept between 36 and 42 mm Hg. For an initial period of 30 min after start of surgery, lungs were ventilated using the ventilator settings described above. Before starting PEEP titration, arterial blood gases were measured. Each pressure level was maintained for approximately 20 min, except for the recruitment phase, during which the highest pressures (PEEP/IP = 20/45 cm H2O) were applied for 20 min or shorter if any of the hemodynamic variables deviated from baseline by more than 15%. Alveolar recruitment and PEEP titration were applied as follows: Pressures were increased stepwise from a PEEP of 0 cm H2O and an IP of 10 cm H2O (further reported as PEEP/IP = 0/10 cm H2O) to 5/15 cm H2O, 10/20 cm H2O, and 15/25 cm H2O. For recruiting collapsed alveoli, a PEEP of 20 cm H2O combined with an IP of 45 cm H2O was used. Afterwards, pressures were reduced to a PEEP of 0 cm H2O and an IP of 10 cm H2O in a manner that mirrored the previous pressure increase. Figure 1 shows the phases of incremental PEEP, recruitment maneuver, and decremental PEEP. Every patient received 500 mL of Ringers solution and 500 mL of 6% hydroxyethyl starch 130/0.5 (Voluven®; Fresenius Kabi, Bad Homburg, Germany) during induction of anesthesia. Another 500 mL of Ringers solution was infused into every patient before the start of the recruitment maneuver.
Heart rate, invasive systolic, diastolic, and mean arterial blood pressure, and peripheral oxygen saturation by pulse oximetry were measured by the Datex Ohmeda S/5 monitor (Datex-Ohmeda Division, Instrumentarium Corp., Helsinki, Finland) during the entire study period. From the readings of the EVITA 4, the variables end-tidal carbon dioxide concentration (Petco2), PEEP, IP, tidal volume, and minute ventilation were recorded. All patients had a chest radiograph within 6 h after surgery to exclude a pneumothorax or other potential complications.
Measurements On every PEEP level, 2 end-inspiratory holds ("inspiration holds") and 2 end-expiratory holds ("expiration holds") were performed (each lasting for at least 5 s). Intrinsic PEEP, i.e., a remaining positive alveolar pressure as a result of dynamic hyperinflation with a FRC exceeding the relaxation volume of the respiratory system,15 was determined automatically by the EVITA 4. Static compliance of the respiratory system was calculated by dividing the expiratory tidal volume by the pressure difference between end-inspiratory plateau pressure and total end-expiratory pressure.16 Dynamic compliance was determined automatically by the EVITA 4. Arterial blood samples were drawn at the end of each protocol step and analyzed for arterial oxygen partial pressure (Pao2), Paco2, arterial oxygen saturation (Sao2), and the pHa. These values were determined using the blood-gas analyzer ABL 700 (Radiometer, Copenhagen, Denmark). Samples were processed within 5 min. For analyzing dead space fraction, the average alveolar expired Pco2 (PAECO2) was determined off-line by using the mean value of slope III of the capnogram recorded by the EVITA 4. Using a time-based capnogram for determining PAECO2 is a modification of the well-known method, which obtains PAECO2 from a volume-based capnogram.17–20 Dead space fraction is defined as the ratio of physiological dead space to tidal volume (VD/VT), and was calculated by Enghoffs modification of Bohrs equation.21
Statistical Analysis
Twenty patients were included in and completed the study. We could not evaluate PAECO2 in two patients because of defective recordings of expired Pco2. Results of FRC, static compliance of the respiratory system, Pao2 dead space fraction, and Paco2 are listed in Figures 2–4.
Comparing the same pressure level before and after recruitment, FRC values and compliance values were significantly higher on all pressure steps, except for the pressure level PEEP/IP 0/10 cm H2O. Significant changes of Pao2 values and dead space fraction after the recruitment maneuver were found only at the corresponding pressure levels of 10/20 cm H2O. Static compliance of the respiratory system and the dynamic compliance determined by the EVITA 4 were highly correlated (with r = 0.98). Intrinsic PEEP remained below 2 cm H2O in all patients. The optimal PEEP was found to be 10 cm H2O after the recruitment maneuver. At this pressure level, all four variables showed significant changes when compared with the corresponding pressure steps before the intervention. SaO2, arterial blood pressure, and heart rate did not differ between test periods. End-tidal CO2 concentrations were not significantly different, except for a diminution of Petco2 during the recruitment maneuver. With respect to the recruitment maneuver, there were no clinically relevant hemodynamic alterations, ventilatory side effects, or any other complications. However, arterial blood pressure decreased by more than 15% from baseline in two patients, leading to early termination of the recruitment procedure after 15 and 11 min, respectively. The maximum tidal volume per kilogram body weight during the recruitment maneuver showed a mean value of 11.4 ± 3.3 mL/kg for all patients and a maximum value of 16.3 mL/kg in one patient. As a result of this hyperventilation, Petco2 during the recruitment period temporarily decreased to values between 20 and 32 mm Hg (mean ± sd was 24.3 ± 3.6 mm Hg).
The first main finding of our study was that a recruitment maneuver in conjunction with PEEP caused a significant increase in FRC, compliance, and Pao2 in concert with a significant decrease in dead space fraction when compared with the same levels of PEEP before the maneuver. These findings imply that patients with healthy lungs undergoing general anesthesia during operations without the surgical procedure influencing the lung, thorax, or position of the diaphragm might benefit from a recruitment maneuver in conjunction with a PEEP of 10 cm H2O. The second main finding was that compliance and dead space fraction can identify an optimal PEEP level because the highest compliance value in conjunction with the lowest dead space fraction indicates a maximum amount of effectively expanded alveoli. This gain in the efficiency of ventilation implies an increase in the number of functional lung units and not a mere expansion of previously open alveoli, thereby eliminating any significant over-distension. In our study, optimal PEEP after the recruitment maneuver was 10 cm H2O. Although both Pao2 and FRC showed a recruitment effect, they could not differentiate between an increase in the number of functional alveolar units and an increase in lung volume due to distension. Although both compliance and dead space fraction suggested an identical value for optimal PEEP, this result may have been coincidental. Therefore, we prefer using both variables together to determine optimal PEEP until further studies can clarify whether compliance or dead space fraction is superior in determining optimal PEEP. Compliance could be easier to measure in the clinical setting, but may have some limitations. A study by Rothen et al.22 showed that the changes over time of static compliance and in the amount of atelectasis estimated by computed tomography (CT) were not in parallel. On the other hand, data from a recent animal study support the use of compliance as a marker of optimal PEEP. The point of maximum compliance indicated the PEEP level at which lung collapse started to occur, as confirmed by the appearance of atelectasis on the CT scan.23 Although dead space fraction might be more difficult to analyze, it provides information about the matching of ventilation and perfusion and is, therefore, particularly suited for establishing an optimal PEEP that is linked to an efficient gas exchange.24,25 The positive effects of recruitment on static compliance of the respiratory system are obvious (Figs. 2 and 3). The distinct differences seen in compliance at identical pressure levels before and after recruitment support the idea of a gain in the number of open alveoli by the recruitment maneuver itself. The marked decrease in compliance at high pressures can be explained by a distension of open alveoli. Because static compliance of the respiratory system and the dynamic compliance automatically determined by the EVITA 4 were very comparable and showed similar behavior over the course of the entire protocol, the latter can easily be used in clinical routine requiring no intervention by the user.23 Dead space fraction expressed as the relation of physiological Vd/vt represents over-distension of alveoli. Therefore, a gain in ventilated alveoli due to a recruitment maneuver should decrease dead space fraction, and its minimum value should correspond to optimal PEEP.25 According to Breen et al.,21 it is important to calculate dead spaces using PAECO2 instead of Petco2 to obtain correct dead space fraction values. A well-known method for calculating the dead space is VD/VT = (Paco2 – PAECO2)/Paco2, in which the values for PAECO2 were obtained from a volumetric capnogram.16–19 We used a slight modification of the original method, in which PAECO2 was obtained by plotting exhaled Pco2 against time. Comparison of the time-based and the volume-based capnograms shows no major difference in the PAECO2 values, since both reflect the mean value of the alveolar part of the CO2 curve. Even if there were small differences between these methods, they would not have influenced our results, since we compared different ventilatory conditions within each patient while using identical apparatuses and methods. Determining the effects of PEEP after recruitment by using dead space fraction values is somewhat cumbersome because arterial blood samples and PAECO2 values from the recorded capnogram are required. In an attempt to simplify dead space determination, other noninvasive indicators of dead space should be evaluated for their usefulness in determining optimal PEEP.25 For detecting FRC, we used a new device which was validated recently in vitro and in vivo.14,26 The FRC results of this study revealed a plausible physiologic picture of the effects of PEEP and PEEP in conjunction with a recruitment maneuver. At identical pressure levels, a larger FRC can be found after the recruitment when compared with the corresponding value before it (Figs. 2 and 3). However, FRC seems to be insensitive for detecting lung over-distension. A gain in absolute lung volume (FRC) after the recruitment, when compared with the same PEEP before it, cannot result from a pure distension of aerated alveoli or airways but must be the result of an increase in the number of functional alveolar units. Thus, this is proof for the intended "opening effect" of an alveolar recruitment. Determination of the effects of PEEP after recruitment is often done by analyzing Pao2, assuming that the extent of lung collapse can be derived from arterial oxygenation. According to different authors, a Pao2 of more than 400 or 450 mm Hg at pure oxygen breathing is characteristic for a nonatelectatic lung.11,27,28 Arterial oxygenation, however, is nonspecific for judging recruitment effect, since Pao2 depends on the hemodynamic and metabolic state as well. Moreover, Pao2 determination is insensitive to the over-distension of alveoli. Even in grossly over-distended lungs, verified by reduced compliance and increased dead space fraction, Pao2 increases until capillary and vascular compression occurs. This fact was also confirmed in our study. As can be seen in Figure 2, the highest Pao2 was reached at recruitment pressures of 20/45 cm H2O. It can be argued that even at these relatively high airway pressures, either no relevant compression of lung vessels or a parallel decline in both ventilation and perfusion in over-distended lung regions was present. Another disadvantage of using Pao2 to determine the effects of PEEP after recruitment is the trouble and expense of drawing and analyzing a series of arterial blood samples in a timely manner. The deterioration of gas exchange as a result of lung collapse is caused mainly by the following pathomechanisms: first, intrapulmonary shunting, and second, a reduced lung volume with a concomitant diminution of compliance.2–4,29–31 Studies have shown that atelectatic lung areas can be opened actively by increased end-IPs and then kept open by sufficient PEEP.5,9,32 One method to overcome atelectasis during general anesthesia is the "Alveolar Recruitment Strategy " introduced by Tusman et al.,10,20,33 in which PEEP is increased from 0 to 20 cm H2O whereas IPs of around 40 cm H2O are used to actively recruit the lungs. In our study, this procedure was once again slightly modified by choosing an IP of 45 cm H2O for the recruitment maneuver to reach a complete recruitment in all patients.7 The effects of a recruitment maneuver on Pao2,7,9 on compliance,9,22,23,34 on the relation of ventilation to perfusion,22 on dead space fraction10,20,25 and on FRC7,22 have been investigated in various studies. However, in the context of recruitment and PEEP, FRC has only been assessed by indirect CT methods, not direct measurements.2,35 There are limitations to our study. When searching for the optimal PEEP for an individual patient we suggest performing a titration. In our study, however, this titration was done in PEEP steps of 5 cm H2O because of overall operating time restrictions in relation to the time needed for performing each FRC measurement. Therefore, our results suffer from a poor resolution for detecting the truly optimal PEEP. In future studies, PEEP and IP should be decreased in smaller steps (i.e., 2 cm H2O) for a more exact identification of optimal PEEP. In this investigation, study periods were long because we determined both FRC in duplicate and blood gas values at the end of each protocol step. To merely recruit collapsed lung tissue, it would have been sufficient to perform a short recruitment maneuver lasting only for a few breaths and a quicker PEEP titration, thereby minimizing the side effects of the recruitment maneuver and PEEP in the clinical setting. In our study, we did not provide outcome data but physiological surrogates for a better patient condition. There is some evidence that adverse effects of atelectasis persist in the postoperative period and can affect patient recovery.29 Since use of recruitment maneuvers in conjunction with sufficient PEEP can overcome atelectasis, it seems to be consequential that lung recruitment might also improve outcome. However, to date it has only been demonstrated that lung recruitment improves physiologic variables such as oxygenation and work of breathing,29 and it remains to be shown the extent to which lung recruitment may affect hospital stay and other variables of outcome.36 Further limitations of this study are the lack of data on oxygen transport, the simplified method for determining dead space fraction, and the small number of patients studied. In conclusion, significant increases in FRC, compliance, and Pao2 and a significant reduction of the dead space fraction could be induced by recruitment in conjunction with sufficient PEEP in patients with healthy lungs. PEEP applied without recruitment did not show the same effects. Compliance and dead space fraction were indicators for efficient ventilation at an optimal PEEP. Although both FRC and Pao2 changed with recruitment and PEEP, these variables proved insensitive to alveolar distension.
Accepted for publication August 23, 2007. Supported by the Department of Anesthesiology of the University Hospital Hamburg. Dieter Weismann is employed by Drägerwerk AG, Lübeck, Germany; for the other authors no conflicts of interest can be stated. Reprints will not be available from the author.
This article has been cited by other articles:
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|