JOURNAL HOME CME HOME THIS MONTH PAST ISSUES ETOC COLLECTIONS
AUTHORS REVIEWERS EDITORIAL BOARD FEEDBACK RSS HELP
A&A International Anesthesia Research Society
 QUICK SEARCH:   [advanced]


     


Anesth Analg 2008; 107:791-792
© 2008 International Anesthesia Research Society
doi: 10.1213/ane.0b013e3181806009
This Article
Right arrow Full Text (PDF)
Right arrow Echo Rounds
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mauermann, W. J.
Right arrow Articles by Park, S. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mauermann, W. J.
Right arrow Articles by Park, S. J.
Related Collections
Right arrow Cardiovascular
Right arrow Echo Rounds
Right arrow Video Clip
Right arrow Technology


CARDIOVASCULAR ANESTHESIOLOGY

Transesophageal Echocardiography in a Patient in Hemodynamic Compromise After Jarvik 2000 TM Implantation: The Suckdown Effect

William J. Mauermann, MD*, Kent H. Rehfeldt, MD*, and Soon J. Park, MD{dagger}

From the Departments of *Anesthesia, and {dagger}Cardiovascular Surgery, Mayo Clinic College of Medicine, Rochester, Minnesota.

Address correspondence and reprint requests to William J. Mauermann, MD, Department of Anesthesiology, Division of Cardiovascular Anesthesia, Mayo Clinic College of Medicine, 200 First St. SW, Rochester, MN 55905. Address e-mail to mauermann.william{at}mayo.edu.

A 39-yr-old man presented with decompensated heart failure for placement of a Jarvik 2000 TM (Jarvik Heart Inc., New York, NY) left ventricular (LV) assist device (LVAD) as a bridge to transplant. He was inotrope-dependent with a LV ejection fraction of 14% and severe right ventricular (RV) dysfunction. The device was placed at the LV apex with the outflow cannula anastomosed to the ascending aorta. He was weaned from cardiopulmonary bypass (CPB) to LVAD support without hemodynamic instability. However, his course was complicated by persistent nonsurgical bleeding.

Nine hours after his initial operation he returned to the operating room in hemodynamic distress for emergent exploration. The patient was tachycardic to 140 bpm with a mean arterial blood pressure of 42–55 mm HG, despite support from the LVAD at setting 3 (approximately 10,000 revolutions per minute [RPM]). Considering the persistent postoperative bleeding, his hemodynamic compromise was likely due, in part, to hypovolemia despite aggressive resuscitation in the intensive care unit. Cardiac tamponade also needed to be excluded. Intraoperative transesophageal echocardiography (TEE) demonstrated no evidence of pericardial fluid or clot accumulation. However, his LV end-diastolic volume was markedly reduced with near collapse of the ventricular walls adjacent to the LVAD inflow site (Fig. 1, Video 1; please see video clip available at www.anesthesia-analgesia.org). Based on the TEE images, LVAD flow was transiently reduced to lower settings and additional resuscitative fluids were administered resulting in adequate LV filling (Fig. 2, Video 2), a mean arterial blood pressure of 75 mm Hg and a heart rate of 90 bpm. Of note, neither the pulmonary artery diastolic pressure nor central venous pressure changed significantly before, during, or after this episode. He was discharged from the hospital without any deficits and subsequently underwent successful orthotopic heart transplantation.


Figure 113
View larger version (77K):
[in this window]
[in a new window]

 
Figure 1. In this midesophageal four chamber view the left ventricle is severely under-filled. The left ventricular assist device pump (large arrow) is contacting the myocardium and inflow obstruction is likely present. Note the small diameter of the mitral annulus and the bowing of the ventricular septum into the left ventricle.

 

Figure 213
View larger version (64K):
[in this window]
[in a new window]

 
Figure 2. In this midesophageal four chamber view the left ventricle is now well filled and the left ventricular assist device pump (arrow) is ideally positioned to receive the left atrial inflow through the mitral valve.

 

DISCUSSION

TEE is essential in the perioperative management of patients undergoing placement of LVADs. In the pre-bypass period, the overall structure of the heart is assessed with careful attention to the competency of the aortic valve and the presence or absence of intracardiac shunts.1,2 After weaning from CPB, TEE is used to assess RV function, LVAD cannulae placement, intracardiac shunting, and LV filling.1,3 We report the use of TEE in the diagnosis and management of a patient in hemodynamic distress after placement of a Jarvik 2000 LVAD.

The second generation, axial flow LVADs, (Jarvik 2000 and Heartmate II TM) represent a change in technology from the pulsatile devices. Axial flow devices use an impeller rotating at 6000 to 15,000 RPM with maximum outputs of up to 10 L/min, depending on the settings. Thus, the impeller of the device continues at its set speed regardless of inflow to the ventricular cannulae. Clearly, a LVAD is capable of providing hemodynamic support only if there is adequate inflow to the device. Delivery of blood to the LV cannulae is dependent on adequate intravascular volume, RV function, and a well positioned inflow cannulae.

If hemodynamic compromise occurs, the position of the inflow cannula (Heartmate II) or pump (Jarvik 2000) in the LV, RV function, and LV filling may all be easily assessed by TEE using midesophageal windows.1 The aorta should be interrogated for signs of dissection. A careful exploration should be undertaken looking for thrombus in the left-sided chambers that may embolize and obstruct the LVAD. Obstruction of the device inflow can occur from thrombus or by collapse of the myocardial walls over the device inflow. Inflow cannula obstruction may be detected by turbulent flow with color Doppler imaging or by a peak velocity more than 2.3 m/s using continuous wave Doppler.1 We find this to be difficult in patients with a Jarvik 2000. With this device, the pump resides within the LV and the inflow is essentially directly into the pump. Although it is still possible to track inflow from the mitral orifice to the device with color Doppler, there is always significant aliasing from the mechanical artifact of the device. In our experience, neither continuous nor pulse wave Doppler evaluations of the Jarvik 2000 inflow have been possible due to reverberation artifact from the pump.

This report demonstrates a phenomenon often seen when patients are initially weaned from CPB and onto support from the axial flow LVAD. The transition must be slow with meticulous attention paid to LV filling. LV filling may be assessed by TEE or direct left atrial pressure measurements. If the RPMs of the device are increased too quickly without adequate ventricular filling, a "suck down" effect quickly occurs whereby the ventricular walls begin to collapse and contact the LVAD inflow leading to obstruction. The ventricular septum bows into the LV and the RV assumes an unfavorable geometry leading to RV dysfunction and further worsening of volume delivery to the LV. Position of the LVAD cannula or pump, LV filling, and the ventricular septum are assessed with the midesopheageal four chamber view. If the midesophageal two chamber view is also used, LV filling and position of the LVAD inflow can be assessed in two planes.

When a patient with a LVAD becomes hemodynamically unstable, the natural tendency is to increase the RPMs of the device in an effort to increase support. However, in the situation of inadequate volume delivery to the device, increasing the RPMs will worsen the clinical picture by causing LV wall collapse at the site of inflow and inducing RV dysfunction. The correct maneuver is to decrease the RPMs while restoring intravascular volume. Once the hyopvolemia is corrected, full support from the LVAD may be reinstituted slowly.

It may be possible to diagnose LV "suckdown" and other causes of hemodynamic instability using transthoracic echocardiography. However, the postoperative surgical changes including drains and bandages, and the position of the device in the chest, make transthoracic imaging and interpretation difficult. We maintain a low threshold for using TEE to evaluate these patients in the immediate postoperative period, and we consider doing so in the operating suite where surgical therapy, if needed, can be swiftly instituted.

Footnotes

This article has supplementary material on the Web site: www.anesthesia-analgesia.org.

Accepted for publication May 1, 2008.

Supported by Departmental and Institutional funds.

REFERENCES

  1. Chumnanvej S, Wood MJ, MacGillivray TE, Melo MF. Perioperative echocardiographic examination for ventricular assist device implantation. Anesth Analg 2007;105:583–601[Abstract/Free Full Text]
  2. Mets B. Anesthesia for left ventricular assist device placement. J Cardiothorac Vasc Anesth 2000;14:316–26[Web of Science][Medline]
  3. Baker JE, Stratmann G, Hoopes C, Donateillo R, Tseng E, Russell IA. Profound hypoxemia resulting from shunting across an inadvertent atrial septal tear after left ventricular assist device placement. Anesth Analg 2004;98:937–40, table of contents[Web of Science]




This Article
Right arrow Full Text (PDF)
Right arrow Echo Rounds
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a colleague
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mauermann, W. J.
Right arrow Articles by Park, S. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mauermann, W. J.
Right arrow Articles by Park, S. J.
Related Collections
Right arrow Cardiovascular
Right arrow Echo Rounds
Right arrow Video Clip
Right arrow Technology


Lippincott, Williams & Wilkins Anesthesia & Analgesia® is published for the International Anesthesia Research Society® by Lippincott Williams & Wilkins and Stanford University Libraries' HighWire Press®. Copyright 2008 by the International Anesthesia Research Society. Online ISSN: 1526-7598   Print ISSN: 0003-2999 HighWire Press