Perfluorocarbon emulsions (PFEs) appear as platelets in automated cell counters, which may affect samples from thrombocytopenic patients (less than 100,000/µL). Therefore, we mixed clinically relevant concentrations of perfluorodichlorooctane (Oxyfluor®; Hemagen, Inc., St. Louis, MO) in vitro with whole blood samples ranging from 0 to 150,000 platelets/µL and compared a new counter that uses optical platelet recognition (Abbott CellDyn 3200; Santa Clara, CA) with conventional electroimpedance-based counters (Abbott CellDyn 3500 and CellDyn 1700). We found that emulsion particles appear as small-sized platelets either in diluent or in blood. The emulsion results in a reproducible overestimate of the platelet counts, of greater importance as PFE concentration increases, and as the actual platelet count of the blood samples decreases. The new optical technology yields smaller overestimates but, even at low PFE concentrations, gives an unacceptable relative error at platelet counts near the transfusion thresholds recommended by the American Society of Anesthesiologists guidelines for blood component therapy. Unexpected interference in the leukocyte and erythrocyte channels is also reported. Experimental limitations preclude extrapolation of these findings to other automated cell counters, because differences in technology or software may affect their capacity to separate PFE particles from platelets. Implications: Perfluorocarbons are being investigated under conditions in which thrombocytopenia is likely to occur. In this in vitro study, we demonstrate significant overestimates in platelet counts from automated cell counters at clinically relevant perfluorocarbon concentrations in thrombocytopenic blood samples.
Perfluorocarbon emulsions (PFEs) are oxygen-carrying solutions now in widespread clinical trials (1). PFEs may reduce requirements for allogeneic blood transfusion and find an application in cardiac surgery or in situations wherein tissue ischemia may occur. At our institution, a second generation PFE, Oxyfluor® (Hemagen, Inc., St. Louis, MO) has been in a Phase I trial for coronary artery bypass graft patients. During this trial, we observed a significant overestimate of the platelet count with impedance-type automated counters, particularly in thrombocytopenic patients. This suggested that PFEs might interfere with automated hematology analyzers. Plasma samples of patients receiving PFEs appear lipemic and, because of light-scattering effects, are known to cause interference in many colorimetric and spectrophotometric methods used in clinical laboratories. These compounds may have impact not only on hematology tests, but also on other laboratory-automated analyzers, including those used in chemistry study (2,3). Hematology tests are important in the care of patients likely to receive PFEs, but few studies have addressed the potential modification of these tests by oxygen-carrying solutions, and have only investigated normal blood samples (36). In this study, we characterize the interference when Oxyfluor® is present at relevant concentrations in human blood. We also compare conventional electroimpedence-type analyzers with a new analyzer that uses optical platelet recognition, and we evaluate the consequences of this interference on thrombocytopenic human blood.
Perfluorochemical Oxyfluor® is a PFE that contains egg yolk phospholipids and safflower oil as the emulsifying agents and perfluorodichlorooctane 40% by volume (PFDCO; Hemagen, Inc., St. Louis, MO).
Hematology Analyzers
Patient Material
Methodology
Mechanism of the Interference.
Impact of the Interference on Thrombocytopenic Blood.
Characterization of the Interference The presence of PFDCO resulted in dose-dependent elevations in the PLT count with the impedance methodology (Table 1). The changes were significant only at 8% PFE (P < 0.01), but demonstrated a trend toward increase even at 1% PFE. There was no change in PLT count with the optical methodology. The successive measurements presented coefficients of variation smaller than 4.5% for the worst performance, not significantly different from control with 1% or 8% PFE, within the normal ranges of each instrument and accounting for reproducibility. The measurements over 4-h presented coefficients of variation smaller than 5.1% and were considered relatively stable, except for the CD1700 where 8% PFE seemed to cause significant variations in PLT counts (P < 0.01 for coefficient of variation 8% PFE vs control or vs 1%; Table 1). RBC counts were not affected with both methodologies, except in the CD1700, where high PFDCO concentrations (8%) caused unexpected decreases (Table 2). WBC counts were not affected with the optical methodology, but were overestimated by impedance methodologies, with significant increases at 8% PFDCO only (P < 0.01; Table 2).
Mechanism of the Interference When PFDCO was alone in diluent, there was a strong linear correlation between measured PLT counts (r2 > 0.99) or WBC counts (r2 > 0.98), and PFDCO concentration with the same differences between the analyzers as described (Fig. 1). In all instruments, PFDCO alone produces the same signal as small PLTs (<10 pL, normal PLT range 225 pL). As PFDCO concentration increases in blood, this signal overwhelms the PLT signal from normal blood faster in impedance instruments than in the optical counter, which performs better (Fig. 2). Similarly, impedence instruments included PFDCO as lymphocytes, which caused a decrease in the granulocytes part of the LD (data not shown).
Impact of the Interference The emulsion always produced an overestimate of the PLT count, with increasing error as PFE concentration increased and as the actual platelet count of the blood samples decreased. Overestimates of PLT counts were smaller, although significant, using optical methodology (Table 3).
We demonstrated a significant overestimate of the PLT count with automated hematology analyzers when PFDCO is present at relevant concentrations in human blood (Table 1). This overestimate is directly related to PFDCO concentration, as well as to the severity of the thrombocytopenia, but it is smaller using optical rather than impedance technology. Unexpectedly, we also found less significant dose-dependent overestimates of the WBCs with the impedance technology, and slight overestimates of the RBCs at large PFDCO concentrations in only one of the two impedence-based analyzers (Table 2). These effects of the PFEs rely on the size of their particles, which range from 0.1 to 0.350 µM (10). All types of analyzers consider these particles as small PLTs (Fig. 2A), although with errors of different magnitude (Fig. 1A). Since the relationship between measured PLT counts versus PFEs alone demonstrated a good linear correlation (r2 = 0.99), this could be potentially used to generate a correction factor for PLT count measurements in the presence of this agent using the PFE concentration in blood, which is easy to measure by centrifugation (11) or to estimate by calculation (12). However, when mixed with blood, this dose-dependent relationship may not be linear, because PFE particles and PLT may interact. Furthermore, this correction factor would be instrument-specific, because each analyzer appears to have a different ability to differentiate PFDCO particles from PLT (Fig. 2, C and D). This ability varies mainly between optical and impedance instruments, although differences in the software algorithms used may also be a factor since two impedance analyzers from the same manufacturer separate PLTs differently from non-PLT signals (Fig. 1A). Both methodologies use either the change in an electrical current (impedance), either the light-scattering angles of a laser beam (optical) produced by each individual cell or other particles to measure their characteristics (8). Combining the information from up to four angles of light scatter, optical sensors can recognize intracellular organelles and offer a better cell discrimination than the single measurement of impedance variation. These findings are apparently contradictory to the decrease in PLT count described by previous investigators using impedance analyzers and attributed to particle agglutination of an unspecified PFE with PLTs (5). This agglutination was possibly related to the large 80% PFE concentration used, which may explain differences from our study, as we used lower, but more clinically relevant, concentrations. Agglutination increased over time while the low coefficients of variation from our measurements over four hours (<5.1%) describe a rather stable phenomenon, occurring immediately after mixing and not disappearing at least for four hoursthe estimated maximum latency before processing. We could not observe any particle size effects because all particles were probably trapped in the agglutinates. This size effect has already been described as reporting the tendency of impedance analyzers to falsely include acellular aggregates in their count (4). The direct relationship between the error in PLT count versus the severity of thrombocytopenia may be problematic. Blood substitutes are currently being investigated in situations in which thrombocytopenia is likely to develop (hemorrhage, extracorporeal bypass) and in which PLT transfusion can reduce morbidity or even be life-saving. In settings such as cardiac surgery, identification of nonsurgical bleeding or PLT transfusion triggers should rely on testing PLT function as well as PLT count, because those procedures are known to affect both PLT quality and quantity (13). In addition, PFEs have been reported to dose-dependently inhibit PLT activation and aggregation (14). Unfortunately, none of the current point-of-care tests of PLT function (bleeding time, thromboelastography, and laboratory testing) achieves the necessary specificity, reliability, simplicity, or rapidity (15). Hence, the identification of PLT transfusion triggers relies on the clinicians experience and on PLT counts. The American Society of Anesthesiologists guidelines for blood component therapy recommend transfusion when the PLT count is below 50,000 or even 100,000/µL for interventions at risk for blood loss and/or PLT consumption (16). Relative errors for these two values are, respectively, 296% or 152% under the best conditions (i.e., optical methodology at 1% PFDCO) (Table 3). Therefore, significant thrombocytopenia in a patient receiving PFEs may go uncorrected if automated PLT counts are not questioned. This study has limitations that may moderate these conclusions. We have studied only one compound, and each of the substances that are currently under investigation may have different impacts on a variety of automated analyzersnot only hematological, but also others including chemistry studies. Another PFE, Perflubron® (Alliance Pharmaceutical Corp., San Diego, CA), may also present the same problem since it has a similar, although somewhat smaller, particle distribution size (0.10.250 µM) as PFDCO (17). However, the published results of Phase II clinical studies with intravascular Perflubron® do not report such interference at equivalent doses (1.5 g/kg), perhaps due to a more uniform particle size distribution and a better stability of the emulsion (18). Because all three studied analyzers from the same manufacturer reacted differently, we did not evaluate whether this effect would differ with other manufacturers instruments. The current study was also limited to mixing PFE with blood in vitro, and it is unknown whether these elevations in blood cell counts may similarly occur with PFEs circulating in vivo for a few hours. The latter issues need to be addressed in patients receiving this and other similar agents by measuring the concentration of the respective agents, the PLT counts via several instruments and via manual counting to assess actual counts. Manual counting would be mandatory to confirm that the actual PLT count is substantially lower than that measured with the analyzers. We processed blood samples in the institutional reference automated analyzer (CD3500) before mixing them with PFEs and corrected these counts for dilution. We chose this approach because manual counting is time-consuming, and some reports suggest that it might not be as accurate as automated analysis when actual counts are low (19). The small number of analyzed human blood samples limits the degree of statistical significance of our conclusions, although the interference in the group of normal blood samples shows a small variability when compared with the importance of the errors (Table 1). As for the reported interference of PFDCO with thrombocytopenia, we chose to obtain thrombocytopenic samples by random selection as needed, rather than by dilution, which made it difficult to constitute a large pool of blood samples and prevented reliable statistical analysis. This should be done in further studies to present the variability inherent in the data and the statistical significance. Despite the limitations of this study, we have demonstrated that PFDCO may interfere significantly with automated hematological analyzers. This error must be considered when evaluating PLT counts in thrombocytopenic patients receiving PFEs, such as PFDCO.