Anesth Analg 2001;92:246-248
© 2001 International Anesthesia Research Society
REGIONAL ANESTHESIA AND PAIN MEDICINE
Breakage of Epidural Catheters: A Comparison of an Arrow Reinforced Catheter and Other Nonreinforced Catheters
Takashi Asai, MD, PhD,
Kei Yamamoto, MD,
Takuji Hirose, MD,
Hitoshi Taguchi, MD, and
Koh Shingu, MD
Department of Anesthesiology, Kansai Medical University, Osaka, Japan
Address correspondence and reprint request to Takashi Asai, Department of Anesthesiology, Kansai Medical University, 10-15 Fumizono-cho, Moriguchi City, 570-8507, Osaka, Japan. Address e-mail to asait{at}takii.kmu.ac.jp
 |
Abstract
|
|---|
Implications: Although breakage of an epidural catheter is rare, our ex vivo study suggests that compared with three other epidural catheters, the Arrow catheter is more likely to stretch and break. The segment between the 7- and 8-cm marks may be an area especially vulnerable to breakage.
 |
Introduction
|
|---|
The Arrow reinforced epidural catheter ( Flex Tip Plus TM; Arrow International, Reading, PA), which contains a stainless steel coil, has advantages over conventional nonreinforced catheters because it is easy to insert (1) and has an infrequent incidence of associated paresthesias1 and epidural vein cannulation2. However, we report a case of breakage of the Arrow catheter during its removal.
 |
Case Report
|
|---|
A 57-yr-old healthy woman was scheduled for total knee arthroplasty. With the patient in the right decubitus position, the L3-4 epidural space was located via a paramedian approach by using a 17-gauge Arrow Tuohy needle 4.5 cm from the skin. A 19-gauge undamaged Arrow catheter was passed easily through the needle 5.5 cm into the epidural space. The catheter was not pulled back through the needle, and the patient had no paresthesia during the procedure. The catheter was used successfully for anesthesia during surgery and for analgesia during the first 4 days after surgery. No apparent damage to, or migration of, the catheter was found at daily inspection. No anticoagulants were given after surgery.
On the fourth day after surgery, we attempted to remove the catheter with the patient on her side, but we encountered resistance with gentle tugging of the catheter at the skin insertion site. Changing the patients position did not help. Further traction stretched the catheter until it became possible to remove it. However, the distal part of the catheter was missing, even though the inner metal coil from the amputated distal segment remained attached to the withdrawn catheter segment. The metal coil was easily pulled from the patients body. The catheter tip, which was surgically removed under local anesthesia, was stretched; the breakage point of the catheter was found to be between the 7- and 8-cm marks. No metal coil was left inside ( Figs. 1, 2).

View larger version (60K):
[in this window]
[in a new window]
|
Figure 1. A segmented Arrow Flex Tip PlusTM epidural catheter. Note that the catheter is cut off between the 7- and 8-cm marks (above), where the density of the inner stainless steel coil changes and where an intact catheter makes a natural curve (below).
|
|

View larger version (94K):
[in this window]
[in a new window]
|
Figure 2. Microscopic view of the proximal edge of a broken Arrow reinforced catheter tip. Note that the inner stainless steel coil is pulled off from the amputated distal segment catheter so that no coil is visible.
|
|
This incident prompted us to design an ex vivo study to examine whether this type of Arrow catheter has an inherent weakness compared with other epidural catheters and to investigate whether the segment between the 7- and 8-cm markswhere the inner metal coil becomes less dense distally (and where there is a natural curve) (Fig. 1)was weaker than the surrounding segments. The Arrow catheter and three types of 19-gauge nonreinforced catheters ( Perifix catheterTM ; B. Braun, Bethlehem, PA; Perisafe catheterTM ; Becton Dickinson, Franklin Lakes, NJ; Portex catheterTM ; SIMS Portex, Kent, UK) were examined for their degree of stretching, the force required to snap each catheter, and the site of breakage of each catheter.
A pair of forceps was applied vertically to the catheter 5 cm from the distal tip, and the forceps were fixed to a stand, simulating the catheter being caught, for example, by a bony structure in the patients body. The catheter was suspended in air, and a small loop was attached 10 cm from the tip with a silk string, simulating the site where the catheter would be grasped during removal.
A weight (200 and 300 g, in turn) was hung from the string loop (at the 10-cm mark), and the length between the 5- and 10-cm marks was measured. The weight was removed. A hand-held scale (TB-611T; Nihon Kohden, Tokyo, Japan) was attached to the string loop, the catheter was pulled manually, and the force required to snap the catheter was recorded. The transducer response is linear over the range of 0 to 1200 g, and this range covers the reported pulling force required to remove an epidural catheter from a patient (24). The accuracy of the scale was confirmed before each use by hanging 100, 200, 500, 1000, and 1200-g weights on it. The speed of pulling was not quantified. The site of breakage was recorded.
We specifically assessed whether the site between the 7- and 8-cm marks of the Arrow catheter was inherently weaker. We repeated the study but fixed the catheter at 7.5 cm and pulled it manually at 12.5 cm.
The F test showed that there were significant differences between groups for the variability of required weight for snapping the catheter, whereas variables were not significantly different between groups for the length of the catheter (at either 200 or 300 g weight). Therefore, the Kruskal-Wallis test was used to compare the required weight to snap the catheter, followed (if significant) by the Mann-Whitney U-test to compare the Arrow catheter with each of the other catheters. One-way analysis of variance was used to compare the length of the catheter, followed (if significant) by Students t-test between the two groups. Fishers exact test was used to compare the incidence between groups of the catheter snapping close to ( 0.5 cm distance from) the fixed point (i.e., at 5 or 7.5 cm). A P value of <0.05 was considered significant. The 95% confidence intervals were also calculated.
The Arrow catheter stretched significantly more (P < 0.001) and snapped at a significantly lower weight (P < 0.01) than each of the other three catheters ( Tables 1, 2). All catheters snapped at the fixed site or pulling site and never between. The incidence of the catheter breaking at 5 cm was significantly more for the Arrow catheter than any of the other three catheters (P < 0.01) (Table 2). In the Arrow catheter, the inside metal coil was exposed and snapped somewhere between the 5- and 10-cm marks for two catheters ( Fig. 3) and did not break for the remaining four catheters. For the Arrow catheter, a significantly lower weight was required to snap the 7.512.5-cm segment than the 510-cm segment (P < 0.05) (Table 2).

View larger version (85K):
[in this window]
[in a new window]
|
Figure 3. An example of an Arrow catheter broken in an ex vivo study. The catheter snapped at the fixed site, but the inner coil, which was not glued to the catheter, was pulled off from the amputated distal segment, and no metal coil was remaining in the tip. In this figure, the tip of the metal cannula was left in the broken catheter tip but could be easily pulled off.
|
|
 |
Discussion
|
|---|
The mean force (measured with a portable force gauge) required to remove an epidural catheter from the patient ranges from 130 to 390 g (24), and the maximum withdrawal force is 1170 g (2). In our study, the snapping forces for the Arrow catheter were always less than the reported maximum withdrawal force, indicating that this catheter is more likely than the other catheters studied to snap (at the fixed site rather than at the pulling site) if the maximum force is applied in the clinical setting. In addition, the inner metal coil may be pulled out of the catheter (Fig. 3), as occurred in our case. However, caution is required for interpreting clinical implications of these results, because this ex vivo study may not accurately reflect the clinical setting.
Although our ex vivo study has limitations and different results might have been obtained by using different methods (5), we attempted to simulate as much as possible the difficulty in removing an epidural catheter. First, we applied a pair of forceps to fix, and a silk loop to pull, the catheter, and this fixation method might have encouraged snapping at the fixed sites that occurred in our study. To measure the tensile strength more accurately, the catheter segment of interest should be fixed by rolling adjacent segments around cylinders with large diameters. Second, the speed of stretching the catheter may also affect when it snaps: the faster the speed, the more likely it is to snap. We attempted only to use a speed similar to that used by clinicians in removing epidural catheters from patients. Third, the breakage of the catheter in our patient might have been caused mainly by prior damage to the catheter. One study has shown that the tensile strength of epidural catheters can be markedly decreased if the catheter has been damaged by the bevel of the Tuohy needle (5).
A few solutions have been proposed for difficulty in removing an epidural catheter (2,3,68). First, the patient should be placed in the lateral decubitus position, because greater force is required to remove the catheter in the sitting position (2,3). Second, flexion of the spine is effective (6). Third, injection of saline through the catheter during removal is effective (7). There was a report of successful removal of a catheter by inserting a Tuohy needle over the catheter (8), but the needle may have cut the catheter during reinsertion.
Although the incidence of breakage of epidural catheters during removal is not known, it appears to be very infrequent (9,10). We routinely use the Arrow catheter because of the ease of insertion and have had only one case of breakage during several years of use. In addition to the current case, we have experienced three other cases of difficulty in removing the Arrow catheter. In all patients, it was impossible to remove the catheter with the patient in any position. However, in all these cases, it became possible to remove the catheter easily when attempts were made again 3060 minutes later. We postulated that patients body movements fortuitously relieved the trapped catheters. Therefore, we suggest that, if there is difficulty in removing the catheter, one should avoid pulling the catheter forcefully and should reattempt removal 3060 minutes later.
 |
Acknowledgments
|
|---|
We thank Japan Medico and Japan Beckton Dickenson for providing us with Portex and Perisafe catheters, respectively.
 |
Footnotes
|
|---|
1Junega M, Kargas GA, Miller DL, et al. Comparison of epidural induced paresthesia in parturients [abstract]. Reg Anesth 1995;20(Suppl):152. 
2Junega M, Kargas GA, Miller DL, et al. Incidence of epidural vein cannulation in parturients with three different epidural catheters [abstract]. Reg Anesth 1995;20(Suppl):4. 
 |
References
|
|---|
-
Banwell BR, Morley-Forster P, Krause R. Decreased incidence of complications in parturients with the Arrow (FlexTip Plus) epidural catheter. Can J Anaesth 1998; 45: 3702.[Web of Science][Medline]
-
Boey SJ, Carrie LES. Withdrawal forces during removal of lumbar extradural catheters. Br J Anaesth 1994; 73: 8335.[Abstract/Free Full Text]
-
Gravenstein N, Blackshear RH, Wissler RN. An approach to spinal or epidural catheters that are difficult to remove [letter]. Anesthesiology 1991; 75: 544.
-
Morris GN, Warren BB, Hanson EW, et al. Influence of patient position on withdrawal forces during removal of lumbar extradural catheters. Br J Anaesth 1996; 77: 41920.[Abstract/Free Full Text]
-
Ates Y, Yücesoy CA, Ünlü MA, et al. The mechanical properties of intact and traumatized epidural catheters. Anesth Analg 2000; 90: 3939.[Abstract/Free Full Text]
-
Sia-Kho E, Kudlak TT. How to dislodge a severely trapped epidural catheter [letter]. Anesth Analg 1992; 74: 933.[Free Full Text]
-
Gadalla F. Removal of tenacious epidural catheter. Anesth Analg 1992; 75: 10712.
-
Shantha TR, Mani M. A simple method to retrieve irretrievable epidural catheters. Anesth Analg 1991; 73: 5089.
-
Ballance JHW. Difficulty in the removal of an epidural catheter. Anaesthesia 1981; 36: 712.
-
Tio TO, Macmurdo SD, McKenzie R. Mishap with an epidural catheter. Anesthesiology 1979; 50: 2602.[Web of Science][Medline]
Accepted for publication September 19, 2000.
This article has been cited by other articles:

|
 |

|
 |
 
J. E. Spiegel, A. Vasudevan, Y. Li, and P. E. Hess
A randomized prospective study comparing two flexible epidural catheters for labour analgesia
Br. J. Anaesth.,
September 1, 2009;
103(3):
400 - 405.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Motamed, F. Farhat, F. Remerand, J. Stephanazzi, A. Laplanche, and C. Jayr
An Analysis of Postoperative Epidural Analgesia Failure by Computed Tomography Epidurography
Anesth. Analg.,
October 1, 2006;
103(4):
1026 - 1032.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Tanizaki, K. Kabashima, K. Takahashi, T. Sakurai, Y. Tokura, and Y. Miyachi
Artificial foreign body embolism after percutaneous cardiac catheterization.
Arch Dermatol,
August 1, 2006;
142(8):
1077 - 1078.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y. Demiraran, I. Yucel, and B. Erdogmus
Subcutaneous effusion resulting from an epidural catheter fragment
Br. J. Anaesth.,
April 1, 2006;
96(4):
508 - 509.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Deusch, J. Benrath, L. Weigl, K. Neumann, and S. A. Kozek-Langenecker
The Mechanical Properties of Continuous Spinal Small-Bore Catheters
Anesth. Analg.,
December 1, 2004;
99(6):
1844 - 1847.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. C. H. Tsui and B. Finucane
Tensile Strength of 19- and 20-Gauge Arrow Epidural Catheters
Anesth. Analg.,
November 1, 2003;
97(5):
1524 - 1526.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Ugboma, X. Au-Truong, L. I. Kranzler, S. H. Rifai, N. J. Joseph, and M. R. Salem
The Breaking of an Intrathecally-Placed Epidural Catheter During Extraction
Anesth. Analg.,
October 1, 2002;
95(4):
1087 - 1089.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|