*Department of Medicine, Cardiovascular Division, Cardiac Arrhythmia Center, University of Minnesota, Minneapolis, Minnesota; and
Department of Anesthesiology and Critical Care Medicine, Leopold-Franzens-University, Innsbruck, Austria
Address correspondence and reprint requests to Keith G. Lurie, MD, Department of Medicine/Cardiovascular Division, University of Minnesota Medical School, MMC 508, AHC, 420 Delaware St. SE, Minneapolis, MN 55455. Address e-mail to lurie002{at}tc.umn.edu
| Abstract |
|---|
|
|
|---|
IMPLICATIONS: This study describes a new inspiratory impedance valve that is attached to a facemask or breathing tube and used during cardiopulmonary resuscitation. The valve helps to draw more blood back into the heart after each chest compression. In a pig study of cardiac arrest, use of this new valve resulted in a doubling of the blood flow to the heart and a near normalization of blood flow to the brain. Clinical studies are underway to determine the potential benefit of this more efficient way to perform cardiopulmonary resuscitation in patients.
| Introduction |
|---|
|
|
|---|
Investigations of the mechanism of ACD CPR led to the realization that complete airway occlusion during the decompression phase of ACD CPR resulted in greater negative intrathoracic pressure, enhanced venous refill of the heart, and therefore, a significant increase in CPR efficiency (4). Consequently, a new inspiratory impedance valve was developed to prevent inspiratory gas exchange only during the decompression phase of CPR, when intrathoracic pressure is less than atmospheric pressure (4,5). Most recently, the new valve has been evaluated in laboratory (4,5) and clinical studies (6). When the impedance valve was used during ACD CPR, coronary perfusion pressure (CPP) and vital organ blood flow were significantly increased compared with ACD CPR alone (4,6).
Because the impedance valve can be easily inserted in the respiratory circuit and does not require specific training when used during standard CPR, this device may also enhance standard CPR efficacy (7). Furthermore, if the new valve improves CPP above a threshold that usually correlates with return of spontaneous circulation, the administration of vasopressor drugs during CPR, such as epinephrine, may be spared (8). Accordingly, the primary goal of the current investigation was to assess the effects of the new valve during standard CPR on CPP. Other endpoints, including vital organ blood flow and return of spontaneous circulation were also evaluated in an animal model of cardiac arrest.
| Materials and Methods |
|---|
|
|
|---|
The study was performed according to Utstein-style guidelines (9) on 22 healthy, 1216-wk-old female domestic farm pigs weighing 2834 kg. The pigs were anesthetized with a single bolus dose of ketamine (20 mg/kg IM), and pentobarbital (15 mg/kg IV bolus followed by 15 mg/kg per hour IV infusion) given via an ear vein. The pigs were intubated during spontaneous respiration with a 7.5-mm cuffed endotracheal tube (Mallinckrodt Critical Care, Glens Falls, NY), and mechanically ventilated (model 607; Harvard Apparatus, Dover, MA) at a volume-controlled setting of 20 mL/kg. During the preparation time, respiratory frequency was adjusted at 1012 breaths/min according to end-tidal and arterial carbon dioxide partial pressure values to maintain the mean arterial carbon dioxide partial pressure at 35 torr; inspiratory oxygen concentration was 21%. Total fluid management consisted of 5001000 mL of normal saline solution administered IV (Flo-Gard 6201; Baxter Healthcare, Hooksett, NH) throughout the 3-h preparation period.
Left ventricular and ascending aortic arch blood pressures were monitored by using a single high-fidelity micromanometer-tipped catheter (Millar Instruments, Houston, TX). This luminal aorto-left-ventricular micromanometer catheter was positioned under fluoroscopic guidance by femoral cutdown, and used for injection of radiolabeled microspheres. To monitor right atrial pressures, another micromanometer-tipped catheter was inserted through a right jugular vein sheath. Reference blood samples for measurement of organ blood flow were withdrawn from a 5F catheter placed in the descending aorta by femoral cutdown. Intrathoracic pressures were measured by using another micromanometer-tipped catheter positioned 2 cm below the tip of the endotracheal tube. It was inserted into the respiratory circuit via a sealed Y-connector, which was positioned between the endotracheal tube and the test valves, as described previously (4,5). Pressures were recorded continuously and the maximal negative intrathoracic pressure during each compression-decompression cycle was measured. For the measurement of body temperature, a thermometer was placed in the rectum. Body temperature was maintained with a heating blanket (Bair Hugger; Augustine Medical, Eden Prairie, MN) between 38.0° and 39.0°C.
Measurements
Pressure tracings obtained from the high-fidelity micromanometer catheters were continuously monitored with a data acquisition and recording system (Superscope II v1.295, GW Instruments, Somerville, MA). Digitized data were analyzed electronically to provide hemodynamic measurements. Heart rate was determined from a simultaneously recorded electrocardiogram signal. CPP calculated during diastole (relaxation) was defined as the arterio-venous pressure difference (time-coincident difference between aortic and right atrial pressure). Arterial blood gases were analyzed (IL 1301; Instrumentation Laboratory, Lexington, MA) every 30 min to ensure adequate acid base status and oxygenation. Organ blood flow was assessed with radiolabeled microspheres according to the method described by Heymann et al. (10).
Microspheres used in this study were radiolabeled with 141Ce, 51Cr, and 113Sr (DuPont-New England Nuclear, Boston, MA). Each microsphere vial was placed in a water bath, and subjected to ultrasonic vibration for 1 min before injection. Approximately 5 x 106 microspheres were then immediately injected into the left ventricle through the lumen of the Millar catheter. The syringe and catheter were flushed with 10 mL of heparinized saline. With an automatic pump (Masterflex; Cole-Parmer, Vernon Hills, IL), arterial blood was continuously withdrawn from the descending aorta at a rate of 6 mL/min just before the microspheres injection to 4 min thereafter. At the end of the experiment, the entire heart, cerebrum, and kidneys were removed. The left ventricular free wall was sectioned into three layers. Aliquots of each tissue were weighed, and placed into vials. Radioactivities from tissues and blood were measured with a
scintillation spectrometer (COBRA II, Auto-Gamma; Packard Instrument Co., Downers Grove, IL), and blood flow values were subsequently calculated.
Experimental Protocol
This prospective randomized blinded evaluation was performed by using two different types of impedance valves (Figs. 1, 2). The functional impedance threshold valve (ITVTM) (CPRx LLC, Minneapolis, MN) included the silicone diaphragm, which occludes the airway within the valve during the decompression phase. In the sham valve, the diaphragm was intentionally omitted during the manufacturing process. Both valves were pigmented so that it was not possible for the research team to know whether or not a functional or sham valve was used during the study. A computer-generated list was used to randomize the valves during the study.
|
|
|
We defined return of spontaneous circulation as an unassisted pulse with a systolic arterial pressure of >80 mm Hg, and pulse pressure of >40 mm Hg, a slightly more stringent criteria than the >60 mm Hg arterial pressure as recommended by the Utstein guidelines in laboratory CPR research. In animals that were successfully resuscitated, an IV saline infusion (300 mL per hour) was started. No other therapy was delivered. Fifteen minutes after return of spontaneous circulation, a third radiolabeled microsphere was injected into the left ventricle to assess vital organ perfusion after return of spontaneous circulation. After finishing the experimental protocol, the animals were euthanized with an overdose of pentobarbital and potassium chloride. All pigs underwent necropsy to check correct positioning of the catheters, damage to the rib cage, and to harvest organs for blood flow measurement.
Values were expressed as mean ± SEM. The primary endpoint was CPP. The comparability of weight and baseline data was tested with the t-test for continuous variables. A two-factor analysis of variance with repeated measures on one factor was used to determine statistical significance between groups. Because blood flow data were unevenly distributed, the Mann-Whitney U-test (two-tailed) was applied to determine a significant difference between each group. Statistical significance was considered to be at P< 0.05 after Bonferroni correction for multiple comparisons.
| Results |
|---|
|
|
|---|
|
|
|
Return of spontaneous circulation could be achieved in 6 of 11 animals in the ITV group versus 3 of 11 in the standard CPR group (not significant). No vasopressors were used to achieve successful resuscitation. In surviving animals, there were no statistically significant differences in hemodynamic variables, organ blood flow, and blood gases between groups during the postresuscitation phase. However, only 2 of 11 pigs in the Control group survived for 15 min, whereas 6 of 11 animals treated with the functional valve were still alive at this point (Tables 1 and 2). Necropsy confirmed appropriate catheter positions, and revealed no injuries of the rib cage or intrathoracic organs in any animals.
Assessment of all valves after the experiment revealed proper function in all cases.
| Discussion |
|---|
|
|
|---|
A distinctive difference between key hemodynamic variables indicates that use of the new valve is beneficial for "priming the pump" and increasing CPR efficiency. The results from these studies suggest that preventing inspiratory gas exchange during standard CPR, when not actively ventilating the pigs, caused an increased negative intrathoracic pressure during the decompression phase of CPR. The consequences, in the absence of powerful vasopressors typically used to increase systemic vascular resistance, were striking. Both hemodynamic data and blood flow measurements support the fundamental importance of enhancing blood return into the thorax during CPR. Accordingly, we found a significant improvement in myocardial and cerebral perfusion. Although increasing cardiac arrest times will increase the resistance to sufficient reflow in vital organ systems (11), brain blood flow was nearly 75% of normal values in the Functional Valve group. Myocardial perfusion was also increased by nearly twofold when compared with standard CPR with the addition of the valve. However, even with the addition of the valve, total myocardial perfusion fell far short of normal values. This is in contrast to our findings when we compared vital organ blood flow during ACD CPR with and without an impedance valve (4,6). With active decompression and the valve, left ventricular and brain blood flow were approximately 55% and 125% of baseline, respectively (4). In patients undergoing CPR, the mean blood pressure was 110/56 when ACD/CPR was performed with the valve versus 90/35 with ACD/CPR alone (6).
The hemodynamic effects of the impedance valve during standard CPR may be closely related to the arrest time, and other further factors such as fluid status, levels of endogenous hormones, diaphragmatic tone, the compliance of the ventricular wall, and duration of CPR. Both chest wall and respiratory system compliance during CPR decrease significantly over time (1216). The effectiveness of the valve depends on many of these factors during standard CPR. This may explain why we observed a greater effect of the valve when the cardiac arrest time, in the absence of CPR, was increased from four to six minutes. In an earlier study, with a shorter arrest time, the effect of the impedance valve was evident but not as striking until the arrest time was lengthened (5).
As reported previously, insertion of the impedance valve in the respiratory circuit decreased minute ventilation by preventing airflow during chest compressions (4). This resulted in higher-than-normal arterial carbon dioxide values in the Functional Valve group (4,5). Moreover, progressive arterial acidosis occurred to a greater extent when the valve was used. The latter finding is likely to reflect the beneficial effects of airway occlusion on systemic circulation, whereas normal arterial pH values during low flow states such as CPR may not reflect the degree of cardio-circulatory failure (16). Because we did not measure mixed venous blood samples, it is not possible to state definitively whether the lower pH in the Active Valve group is secondary to increased metabolic acidosis or worsening respiratory acidosis. Although the decrease in gas exchange in the Functional Valve group was obvious, all variables remained within adequate levels. Furthermore, the trend of 6 of 11 ITV versus 3 of 11 standard pigs with return of spontaneous circulation suggests that decreased minute ventilation in the Functional Valve group had no negative impact on survival rates. Fifteen minutes after return of spontaneous circulation, 6 of 11 pigs treated with the functional valve were alive versus 2 of 11 in controls.
Some limitations of the present study should be noted. We did not assess mixed venous blood gases; accordingly, we are unable to interpret gradients between mixed venous and arterial blood gas values. Ventilation was synchronized with chest compressions and delivered primarily during the decompression phase. Therefore, we were unable to report on effects of the valve when chest compressions and ventilation were not synchronized. In addition, we determined the sample size based on CPP as the primary endpoint. Consequently, a potential survival benefit from the valve will require another study (continuing) with a larger number of animals.
Additional limitations include that we did not measure chest wall compliance in this study, which varies in pigs and humans from one chest to the next. Inter-individual differences in chest wall may strongly influence the effect of intermittent airway occlusion on hemodynamics during standard CPR. As such, it is important to reemphasize that, when performing standard CPR on a pig or on a patient, the chest must be allowed to recoil to its natural resting position to provide the chest recoil needed to create a slight vacuum within the thorax relative to the rest of the body and atmospheric pressure. We also observed some differences in the baseline CPPs between the groups. Although these were not statistically significant, they could have altered the overall results. Finally, it is important to note that we observed no significant adverse effects of the impedance valve in this model.
In conclusion, the use of an impedance valve during standard CPR resulted in a marked increase in CPP and vital organ blood flow when compared with standard CPR alone. On the basis of these favorable results, examination of the potential benefit of intermittent impedance of inspiratory gas exchange during standard CPR in patients seems warranted.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. A. Rickards, K. L. Ryan, W. H. Cooke, K. G. Lurie, and V. A. Convertino Inspiratory resistance delays the reporting of symptoms with central hypovolemia: association with cerebral blood flow Am J Physiol Regulatory Integrative Comp Physiol, July 1, 2007; 293(1): R243 - R250. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Srinivasan, V. M. Nadkarni, D. Yannopoulos, B. S. Marino, G. Sigurdsson, S. H. McKnite, M. Zook, D. G. Benditt, and K. G. Lurie Rapid Induction of Cerebral Hypothermia Is Enhanced With Active Compression-Decompression Plus Inspiratory Impedance Threshold Device Cardiopulmonary Resusitation in a Porcine Model of Cardiac Arrest J. Am. Coll. Cardiol., February 21, 2006; 47(4): 835 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
Part 4: Advanced Life Support Circulation, November 29, 2005; 112(22_suppl): III-25 - III-54. [Full Text] [PDF] |
||||
![]() |
D. Yannopoulos, V. M. Nadkarni, S. H. McKnite, A. Rao, K. Kruger, A. Metzger, D. G. Benditt, and K. G. Lurie Intrathoracic Pressure Regulator During Continuous-Chest-Compression Advanced Cardiac Resuscitation Improves Vital Organ Perfusion Pressures in a Porcine Model of Cardiac Arrest Circulation, August 9, 2005; 112(6): 803 - 811. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. B. Wolcke, D. K. Mauer, M. F. Schoefmann, H. Teichmann, T. A. Provo, K. H. Lindner, W. F. Dick, D. Aeppli, and K. G. Lurie Comparison of Standard Cardiopulmonary Resuscitation Versus the Combination of Active Compression-Decompression Cardiopulmonary Resuscitation and an Inspiratory Impedance Threshold Device for Out-of-Hospital Cardiac Arrest Circulation, November 4, 2003; 108(18): 2201 - 2205. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. G. Lurie, T. Zielinski, S. McKnite, T. Aufderheide, and W. Voelckel Use of an Inspiratory Impedance Valve Improves Neurologically Intact Survival in a Porcine Model of Ventricular Fibrillation Circulation, January 1, 2002; 105(1): 124 - 129. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|