*Department of Anesthesiology, Leiden University Medical Center, Leiden, The Netherlands;
Scottish National Sleep Laboratory; and
Department of Anaesthesia, Critical Care and Pain Medicine, Royal Infirmary, Edinburgh, United Kingdom
Address correspondence and reprint requests to Dr. D. Nieuwenhuijs, Department of Anesthesiology, Leiden University Medical Center (P5-Q) PO Box 9600, 2300 RC Leiden, The Netherlands. Address e-mail to djfn{at}worldonline.nl
| Abstract |
|---|
|
|
|---|
IMPLICATIONS: Although computer processing of the electroencephalogram can provide an adequate index of depth of anesthesia, the same processing cannot reliably convey depth of natural sleep. At each sleep stage, the output signal has a wide range of possible values.
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
The polysomnograph variables were recorded with a computerized 16-channel recording system (Compumedics, Melbourne, Australia). The recorded signals from all of the subjects were analyzed manually by the same trained sleep technician according to standard criteria (11). Each recording was divided in epochs of 30 s and each epoch classified as one of the following sleep stages: awake (i.e., roused from sleep), stage 1, 2, 3, or 4 sleep, or rapid eye movement (REM) sleep. BIS in arbitrary values from 0 to 100, SEF (units Hz), and signal quality index (SQI) (percent satisfactory values) were recorded at 5-s intervals from the serial port of the A-2000 monitor to a laptop computer (Toshiba Satellite Pro 4200; Toshiba, Tokyo, Japan), by using the Microsoft HyperTerminal program (Hilgraeve, Monroe, MI). For further analysis, values of the BIS and SEF were obtained at the end of each period of uniform sleep/roused stage and note was taken of the duration of this period. Only stages lasting 2 min or more were included in the analysis (stable sleep/awake state). Signals with an SQI <50% were not analyzed.
The distribution of BIS and SEF values for each sleep stage was not normal. We used descriptive nonparametric statistics to examine the relationship between these signals and the categories of sleep. All values were presented as median (lower [25%] and upper [75%] interquartile values) and SD unless otherwise stated. To simplify the presentation of the data, and because relatively little time was spent in sleep stages 1 and 3, we considered stages 1 and 2 as "light sleep" and stages 3 and 4 as "slow wave sleep" (SWS).
| Results |
|---|
|
|
|---|
From polysomnography, the median recording time (i.e., time in bed) was 470 min (451500 min); the median time subjects needed to fall asleep was 8 min (415 min); median sleeping time was 412 min (396458 min); and median time aroused was 49 min (3560 min). Of the time spent asleep, 11 min (812 min; 2.5% of total sleep time) was scored as stage 1, 253 min (220267 min; 59.1% of total sleep time) as stage 2, 16 min (1018 min; 3.5% of total sleep time) as stage 3, 60 min (4878 min; 15.7% of total sleep time) as stage 4, and 80 min (5398 min; 18.9% of total sleep time) as REM sleep. The sleep efficiency (time asleep/time in bed) was 89% (84%91%).
From the BIS analysis, SQI was >50% during the studies in 9 subjects. In one subject, multiple episodes of SQI <50% were observed (45% of sleep time), and these data were not used. Before the subjects fell asleep, their median BIS values were 97 (9498) ± 12.1 and their median SEF values were 23 Hz (2025 Hz) ± 4.2. The BIS and SEF values both decreased with increasing sleep stages (Fig. 1), but the overlap in values at each stage was considerable (Figs. 2 and 3). After sleep initiation, the median BIS values for light sleep (stages 1 and 2 sleep) and SWS (stages 3 and 4 sleep) were 50 (4465) ± 16.5 and 42 (3549) ± 11.2, respectively, and the median SEF values were 15 Hz (1318 Hz) ± 3.6 and 10 Hz (911 Hz) ± 2.6, respectively. During REM sleep, the respective BIS and SEF values were 48 (4451) ± 7.1 and 19 Hz (1621 Hz) ± 4.1. Arousal caused increases in BIS and SEF values to 67 (5288) ± 20.2 and 20 Hz (1522 Hz) ± 4.7, respectively. These BIS values (after arousal) were significantly different from presleep awake values (see above) (P < 0.01,
2 test). SEF aroused and presleep awake values did not differ (P = 0.06).
|
|
|
| Discussion |
|---|
|
|
|---|
Our subjects were studied in the sleep laboratory and may not have slept as well as at home. However, we hoped to apply the BIS measurements in circumstances in the hospital where sleep quality is disturbed, such as after surgery and in high-dependency units, so this is not necessarily a weakness in this study. We selected sleep periods that had not been interrupted for at least two minutes (stable sleep), and sampled only the BIS and SEF signals from the end of the stable sleep stages. The BIS signal is derived from at least 15 seconds of preceding data [variable calculation, moving average, and artifact rejection (13)], so this precaution should provide values that would allow the BIS and sleep state values to agree. In practice, if BIS was to be used as an automatic index of sleep state, this "selection" of stable signals would not be possible, and any possible concordance with sleep stage could be reduced further.
The BIS is a complex variable that, among other things, reflects the coherence of different frequency components of the EEG. It uses a combination of analysis in time and frequency domains, and generates the index in relation to a database that has been obtained from patients during stable anesthesia of different depths (10). It is thus to be expected that different degrees of drug effect can be predicted from BIS analysis. The BIS signal seems to be a valid measure of anesthetic effect; for example, the dose of anesthetic needed to generate a given BIS value is related to the age of the subject, in the same way as anesthetic potency measured conventionally by minimum alveolar concentration value is related to age (14). Nevertheless, BIS values can vary considerably, for a given dose of anesthetic, or a given drug effect. For example, approximately 50% of the BIS values obtained when end-tidal sevoflurane is 2% overlap with the values present when end-tidal sevoflurane is 1% (14), and BIS values at different sedation scores in patients given sevoflurane can overlap completely (15). Although others have suggested that values obtained by automatic processing of the EEG may indicate the depth of natural sleep, natural sleep is clearly not the condition that was used to "calibrate" the BIS device, and the capacity to predict sleep stage from BIS measurements has not been fully explored in previous reports (9). We explored this relationship in greater detail, because an automatic indicator of sleep state would be of great use in studies of postoperative breathing and respiratory control.
We observed a decrease in BIS values and SEF frequencies with increasing sleep depth from light sleep to SWS (Fig. 1). Although this suggests that both BIS and SEF are useful to discriminate sleep stages, the spread of values measured at each stage of sleep clearly precludes the use of these signals as measures of sleep stage. Fell et al. (16) used nonlinear methods to analyze the EEG during the different sleep stages. Their results indicate that nonlinear EEG measures (such as correlation dimension) yield additional information to the classical spectral measures in discriminating between sleep stages. However, the standard deviations of the measures at each of the sleep stages indicate considerable overlap for both spectral and nonlinear measures of the EEG, in agreement with our observations.
Our findings do not support those of Sleigh et al. (9). They measured BIS and SEF values in five volunteers who were sleeping at home. They reported greater BIS values for light sleep, SWS, and REM sleep than we did (light sleep 81 versus 50; SWS 59 versus 42; REM sleep 83 versus 48). However, they only analyzed the first sleep cycle of the night and restricted their analysis to testing for significant differences in the values measured at the start of each sleep stage, which may explain why the BIS values they report are considerably greater than those in the present study.
The BIS values we found in our aroused subjects were frequently lower than the values observed in the awake, presleep state. We suspect that this may be, in part, because the BIS value takes a long time to recover after arousal. Two examples of such a delay are shown in Figure 4. Because the BIS value output by the A-2000 represents the mean of at least 15 but sometimes 60 s of data (Olofsen and Dahan, unpublished observation), a delay of perhaps 1 minute might be expected, but in these examples, the delays are 2 minutes or longer. An alternative possibility is that the EEG and electromyogram features of arousal become apparent quickly, whereas the temporal synchrony measured by the BIS takes longer to dissipate after arousal. During anesthetic measurements relating BIS to sevoflurane and isoflurane concentrations, the BIS seems to lag behind the calculated time for drug effect, and one potential explanation for this delay may be the slow changes in interactions between separate cortical regions (13).
|
| Acknowledgments |
|---|
The EEG monitor and electrodes were provided by Aspect Medical Systems International BV, Leiden, The Netherlands.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
K. E. Kotliar, B. Mucke, W. Vilser, R. Schilling, and I. M. Lanzl Effect of Aging on Retinal Artery Blood Column Diameter Measured along the Vessel Axis Invest. Ophthalmol. Vis. Sci., May 1, 2008; 49(5): 2094 - 2102. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Lamas, J. Lopez-Herce, L. Sancho, S. Mencia, A. Carrillo, M. J. Santiago, and V. Martinez Bispectral Index and Middle Latency Auditory Evoked Potentials in Children Younger Than Two-Years-Old Anesth. Analg., February 1, 2008; 106(2): 426 - 432. [Abstract] [Full Text] [PDF] |
||||
![]() |
A A Dahaba, H C Worm, S M Zhu, F P Bao, A Salah, S Zakaria, H Bornemann, V Stadlbauer, P H Rehak, H Metzler, et al. Sensitivity and specificity of bispectral index for classification of overt hepatic encephalopathy: a multicentre, observer blinded, validation study Gut, January 1, 2008; 57(1): 77 - 83. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Howard, J. J. Finneran, and S. H. Ridgway Bispectral index monitoring of unihemispheric effects in dolphins. Anesth. Analg., September 1, 2006; 103(3): 626 - 632. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. J. De Jong, S. M. Burns, M. L. Campbell, M. Chulay, M. J. Grap, L. N.B. Pierce, and T. Simpson Development of the American Association of Critical-Care Nurses' Sedation Assessment Scale for Critically Ill Patients Am. J. Crit. Care., November 1, 2005; 14(6): 531 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Benini, M. Trapanotto, S. Sartori, A. Capretta, D. Gobber, C. Boniver, and F. Zacchello Analysis of the Bispectral Index During Natural Sleep in Children Anesth. Analg., September 1, 2005; 101(3): 641 - 644. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Soto, T. C. Nguyen, and R. A. Smith A Comparison of Bispectral Index and Entropy, or How to Misinterpret Both Anesth. Analg., April 1, 2005; 100(4): 1059 - 1061. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nishiyama and K. Hanaoka The A-line ARX index may be a more sensitive detector of arousal than the bispectral index during propofol-fentanyl-nitrous oxide anesthesia: a preliminary investigation: [L'index < Can J Anesth, June 1, 2004; 51(6): 539 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Rodriguez, L. E. Hall, S. Duggan, and W. M. Splinter The bispectral index does not correlate with clinical signs of inhalational anesthesia during sevoflurane induction and arousal in children: [L'index bispectral ne correspond pas aux signes cliniques de l'anesthesie par inhalation pendant l'induction au sevoflurane et le reveil chez les enfants] Can J Anesth, May 1, 2004; 51(5): 472 - 480. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Nishiyama, T. Matsukawa, and K. Hanaoka A Comparison of the Clinical Usefulness of Three Different Electroencephalogram Monitors: Bispectral Index, Processed Electroencephalogram, and Alaris Auditory Evoked Potentials Anesth. Analg., May 1, 2004; 98(5): 1341 - 1345. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Barker Too Much Technology? Anesth. Analg., October 1, 2003; 97(4): 938 - 939. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|