| Abstract |
|---|
|
|
|---|
IMPLICATIONS: The laryngeal mask airway ProSealTM is an effective temporary ventilatory device in grossly and morbidly obese patients before laryngoscope-guided tracheal intubation.
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
3550 kg/m2) or morbidly (BMI
50 kg/m2) obese patients scheduled for elective surgery, who preferred airway management under general anesthesia. Patients were excluded if they had more than 1 symptom of gastroesophageal reflux per week, mouth opening <2.5 cm, or had a known difficult airway. In our institutes, anesthesia management for these patients would normally comprise preoxygenation, induction of anesthesia, face mask ventilation, and intubation. In this feasibility study, the PLMA was used as an alternative to the face mask before tracheal intubation. Patients were premedicated with ranitidine 150 mg per os 1.5 h preoperatively. Monitoring was applied before induction and included an electrocardiograph, pulse oximeter, gas analyzer, arterial line, tidal volume monitor, airway pressure monitor, and peripheral nerve stimulator. Anesthesia was given with the patient in the supine position with the patients head on a standard pillow 8 cm in height. Midazolam 0.02 mg/kg and fentanyl 1 µg/kg were administered. Patients were preoxygenated for 5 min. Anesthesia was induced with propofol 23 mg/kg given over 30 s, and the PLMA inserted when there was no response to jaw thrust (5). Additional boluses of propofol 0.5 mg/kg were given as required until an adequate level of anesthesia was achieved for placement. Anesthesia was maintained with sevoflurane 1%3% in oxygen 100%. Face mask ventilation was performed if conditions for insertion were not suitable within 30 s of completion of the induction dose.
The size 5 PLMA was inserted/fixed according to the manufacturers instructions (6). The introducer tool was used. The PLMA was connected to a circle anesthesia breathing system and the cuff inflated with air until an effective airway was established or the maximal recommended inflation volume reached (40 mL). An effective airway was judged by a square wave capnograph trace and no audible leak with peak airway pressures
12 cm H2O during manual ventilation. The PLMA was removed and reinserted if air leaked up the drainage tube, or there was airway obstruction, or the bite block protruded from the teeth/gums. Once an effective airway was obtained, rocuronium 0.5 mg/kg IV was administered for paralysis. The intracuff pressure was set at 60 cm H2O, and the oropharyngeal leak pressure was determined (7). Any gastric insufflation was noted during oropharyngeal leak pressure testing by listening over the epigastrium with a stethoscope (8). A well-lubricated 14F gastric tube was inserted through the drainage tube if there was no air leak up the drainage tube. Correct gastric tube placement was assessed by suction of fluid or detection of injected air by epigastric stethoscopy. The gastric tube was then aspirated, removed, and the volume of aspirate noted. A fiberoptic scope was passed down to the distal aperture of the airway tube and the visibility of the vocal cords and/or esophagus noted. The fiberoptic position of the drainage tube was determined by passing a fiberoptic scope down the drainage tube to a position just proximal to the end of the tube. The view was cataloged as hypopharynx (mucosa), open upper esophageal sphincter (a clear view down the esophagus), and others (glottis, epiglottis, arytenoids). Patients were ventilated for 3 min with the cuff fully inflated at an inspired tidal volume of 8 mL/kg (true weight), a respiratory rate of 12/min, and an inspiratory/expiratory ratio of 1:1.5. The presence/absence of oropharyngeal air leaks (detected by listening over the mouth (7)), gastric air leaks (detected by listening with a stethoscope over the epigastrium (8)), drainage tube air leaks (detected by placing lubricant over the proximal end of the drainage tube), or an end-tidal CO2 >45 mm Hg was noted. Once these data were collected and the train-of-four count was zero, the PLMA was removed and laryngoscope-guided intubation attempted by using a MacIntosh laryngoscope blade (Parker Healthcare, Mansfield, Australia) and a size 8.0-mm tracheal tube. Conditions for intubation were optimized, as described by Benumof (9). If the vocal cords were not visible, a gum elastic bougie was used. The Cormack and Lehane (10) score was noted at laryngoscopy: grade 1, glottis visible including anterior and posterior commissures; grade 2, glottis visible, but anterior commissure not seen; grade 3, glottis not visible, but corniculate cartilages visible; grade 4, glottis and corniculate cartilages not visible. The number of insertion and intubation attempts was recorded. A maximum of two attempts was allowed before insertion or intubation was considered a failure. A failed attempt was defined as removal of the PLMA or laryngoscope from the mouth. If either technique failed, the alternative technique was used. The time between picking up the PLMA or laryngoscope and obtaining an effective airway was recorded. Any episodes of hypoxia (SpO2 <90%) or other adverse events were documented.
All cases were conducted by anesthesiologists with more than 10 yr clinical experience of laryngoscope-guided tracheal intubation and more than 100 uses of the PLMA. Data were collected by unblinded observers. Statistical analysis was performed with paired t-test,
2 test, and Kendells rank correlation. Data are mean ± SD (range) or numbers unless otherwise stated. Significance was taken as P < 0.05
| Results |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
We found that positive pressure ventilation with a tidal volume of 8 mL/kg was possible in 95%98% of patients. This is comparable to the efficacy of the PLMA (12) and LMA1(13) as a ventilatory device for nonobese patients. We found that there were no air leaks up the drainage tube during oropharyngeal leak pressure testing, that gastric insufflation did not occur, and that gastric tube insertion was easy. This suggests that the respiratory tract was effectively isolated from the gastrointestinal tract. There is evidence from cadavers (14) and an anesthetized patient (15) that the PLMA will protect the lungs from regurgitated gastric contents if it is correctly placed. The frequency of the upper esophageal sphincter being open was similar to the previous study of nonobese patients (2). Assuming a binomial distribution for the incidence of adverse events, the upper limit for the probability of an adverse event occurring when no such events have been observed in 60 patients is 0.07 (16). Therefore, we can state with 95% confidence that the technique seems to be safe in at least 93% of grossly and morbidly obese patients.
We found that there was no relationship between Mallampati or Cormack and Lehane scoring and any measured variable for the PLMA. This suggests that the PLMA insertion is feasible in obese patients with predicted and unpredicted difficult airways. Similar findings have been obtained for the LMA for both Mallampati (1719) and Cormack and Lehane2 (20) scoring. Interestingly, in all six patients in the current study who required more than one intubation attempt, the PLMA formed an effective airway at the first insertion attempt. Insertion of the PLMA is more difficult than the LMA in nonobese patients with normal airways (2), but the implications of this for the difficult airway scenario in obese and nonobese patients is unknown.
Our study has a number of limitations. First, we only used the PLMA for a period of 510 minutes before the patient was intubated and our data must be interpreted cautiously regarding its use for longer periods. However, we consider it unlikely that airway quality would have deteriorated with time. There is evidence that airway quality does not change with time for the LMA (21,22) and the stability of the LMA and PLMA are similar in different head and neck positions (23). Also, there was no change in airway quality in the two patients managed with the PLMA during the procedure after failed intubation. Second, airway management was performed by experienced users and our findings might not apply to less experienced users. Third, any comparisons with tracheal intubation must be interpreted cautiously because PLMA insertion always preceded intubation. We elected to intubate patients after PLMA insertion because this is the current "gold standard" for airway management of grossly and morbidly obese patients. We considered that using the PLMA for the entire procedure was not justified given the current lack of knowledge about the device in this high-risk population. However, based on our findings, we consider that the PLMA could be used as an alternative ventilatory device in grossly and morbidly obese patients. Fourth, data collection was unblinded. Fifth, we did not use a control group managed with the LMA.
We conclude that the PLMA is an effective temporary ventilatory device in grossly and morbidly obese patients before laryngoscope-guided tracheal intubation. Further investigations into the prolonged use of the PLMA in this population are justified.
| Footnotes |
|---|
2 Mahiou P, Narchi P, Veyrac P, et al. Is laryngeal mask easy to use in case of difficult intubation? [abstract] Anesthesiology 1992;77:A1228. ![]()
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
U. Freo, M. Carron, M. Micaglio, and C. Ori ProSeal laryngeal mask airway for laparoscopic gastric banding in a myasthenic, morbidly obese patient Br. J. Anaesth., December 1, 2007; 99(6): 921 - 922. [Full Text] [PDF] |
||||
![]() |
A. von Goedecke, H. Herff, P. Paal, V. Dorges, and V. Wenzel Field Airway Management Disasters Anesth. Analg., March 1, 2007; 104(3): 481 - 483. [Full Text] [PDF] |
||||
![]() |
R. Garcia-Aguado, J. Vinoles, J. Brimacombe, M. Vivo, R. Lopez-Estudillo, and G. Ayala Suction catheter guided insertion of the ProSealTM laryngeal mask airway is superior to the digital technique: [L'insertion du masque larynge ProSealTM, guidee par une sonde d'aspiration, est superieure a la technique digitale]. Can J Anesth, April 1, 2006; 53(4): 398 - 403. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Cook, G. Lee, and J. P. Nolan The ProSealTM laryngeal mask airway: a review of the literature: [Le masque larynge ProSealTM : un examen des publications] Can J Anesth, August 1, 2005; 52(7): 739 - 760. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Brimacombe and C. Keller A Proposed Algorithm for the Management of Airway Obstruction with the ProsealTM Laryngeal Mask Airway Anesth. Analg., January 1, 2005; 100(1): 298 - 299. [Full Text] [PDF] |
||||
![]() |
A. A. Matioc and J. Olson Use of the Laryngeal TubeTM in two unexpected difficult airway situations: lingual tonsillar hyperplasia and morbid obesity: [L'usage du Laryngeal TubeTM dans deux situations d'intubation difficile inattendue : hyperplasie des amygdales linguales et obesite morbide] Can J Anesth, December 1, 2004; 51(10): 1018 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Gaszynski and P. Juvin Standard Clinical Tests for Predicting Difficult Intubation Are Not Useful Among Morbidly Obese Patients * Response Anesth. Analg., September 1, 2004; 99(3): 956 - 956. [Full Text] [PDF] |
||||
![]() |
C. Keller, J. Brimacombe, P. Lirk, and F. Puhringer Failed Obstetric Tracheal Intubation and Postoperative Respiratory Support with the ProSealTM Laryngeal Mask Airway Anesth. Analg., May 1, 2004; 98(5): 1467 - 1470. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ovassapian, A. Norris, and D. Hinchliffe Management of failed intubation in a septic parturient Br. J. Anaesth., July 1, 2003; 91(1): 154 - 155. [Full Text] [PDF] |
||||
![]() |
J. B. Brodsky, H. J. M. Lemmens, J. G. Brock-Utne, L. J. Saidman, R. Levitan, B. O. Ogunnaike, S. B. Jones, C. W. Whitten, D. B. Jones, and D. Provost Anesthetic Considerations for Bariatric Surgery: Proper Positioning is Important for Laryngoscopy * Response Anesth. Analg., June 1, 2003; 96(6): 1841 - 1842. [Full Text] [PDF] |
||||
![]() |
R. M. Cooper The LMA, laparoscopic surgery and the obese patient - can vs should: Le ML, la chirurgie laparoscopique et le patient obese - pouvoir vs devoir Can J Anesth, January 1, 2003; 50(1): 5 - 10. [Full Text] [PDF] |
||||
![]() |
J. R. Maltby, M. T. Beriault, N. C. Watson, D. J. Liepert, and G. H. Fick LMA-ClassicTM and LMA-ProSealTM are effective alternatives to endotracheal intubation for gynecologic laparoscopy: [Le ML ClassiqueTM et le ML ProSealTM peuvent remplacer efficacement l'intubation endotracheale pour la laparoscopie gynecologique] Can J Anesth, January 1, 2003; 50(1): 71 - 77. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|