| Abstract |
|---|
|
|
|---|
IMPLICATIONS: In 96.6% of patients intubated with the Laryngeal Tube®, it was possible to maintain oxygenation, ventilation, and respiratory mechanics during mechanical ventilation.
| Introduction |
|---|
|
|
|---|
|
| Methods |
|---|
|
|
|---|
Three attending anesthesiologists used the Laryngeal Tube during the study. Each had performed at least 10 Laryngeal Tube insertions before the implementation of the study.
All patients received premedication with oral diazepam 1020 mg. After 3 min of preoxygenation, anesthesia was induced with up to 3 µg/kg of fentanyl and 23 mg/kg of propofol and was maintained with 70% N2O in 30% (remainder) oxygen and isoflurane. Neuromuscular blockade was obtained with vecuronium 0.1 mg/kg and maintained throughout the surgery to train-of-four count of 1/4, as assessed by using a peripheral nerve stimulator with electrodes placed over the ulnar nerve. After confirmation of complete neuromuscular blockade, a Size 4 Laryngeal Tube was inserted in accordance with the manufacturers recommendations. The patients were positioned in sniffing position.
Both balloons were sequentially inflated with air by using a manometer (Cuff Pressure Gauge; VBM Medizintechnik GmbH) until intraballoon pressure reached 80 cm H2O, and then the deflate valve was pressed to adjust the pressure to 60 cm H2O. Proper positioning of the Laryngeal Tube was confirmed by bilateral chest movement and auscultation, absence of gastric insufflation, and partial pressure of end-tidal CO2 (ETCO2).
The lungs were ventilated with volume-controlled mechanical ventilation by using a ventilator of an AS/3TM anesthesia delivery unit (Datex-Ohmeda, Helsinki, Finland) with a semiclosed circuit incorporating a CO2 absorber. Ventilatory settings included an inspiratory/expiratory ratio of 1:2 and an average tidal volume of 12 mL/kg. The initial respiratory rate was 12 breaths/min and was adjusted to obtain an ETCO2 of 40 mm Hg. The fresh gas flow was 3 L/min. Blood pressure, heart rate, oxygen saturation (SpO2), ETCO2, and end-tidal isoflurane concentration were measured with the AS/3 monitor. Breath-by-breath spirometry data were obtained by using a sidestream spirometry device (D-liteTM flow sensor; Datex-Ohmeda) attached between the proximal end of the Laryngeal Tube and the Y-piece. Data measured included airway pressures (peak, plateau, and positive end-expiratory pressure, lung volumes (minute and tidal volume), graphically displayed loops (pressure volume and flow volume) and curves (pressure and flow), airway resistance, and dynamic compliance. Data were recorded at 5-min intervals after the introduction of the Laryngeal Tube.
Gas leak was determined by auscultation with a stethoscope placed on the neck region and/or as a nonclosing flow-volume loop. Upper airway trauma was assessed by checking for the presence of blood on the Laryngeal Tube after its removal (0 = no blood and 1 = blood stains present) and by examining the patients for sore throat and hoarseness 24 h after surgery.
The number of attempts taken to insert the Laryngeal Tube was recorded. The insertion time was noted from removal of the face mask to attachment of the breathing system to the Laryngeal Tube after inflation of the cuffs.
Intraballoon pressures (proximal and distal) were measured by using the manometer connected to the pilot tube of the pharyngeal cuff of the Laryngeal Tube and were recorded at 10-min intervals, beginning after the introduction of the Laryngeal Tube, for 30 min of anesthesia. The intraballoon pressure was measured during the expiratory phase of the ventilatory cycle to prevent changes in airway pressure from affecting intraballoon pressure. After 30 min of anesthesia, the intraballoon pressure was released to the initial value of 60 cm H2O.
The study was terminated immediately if ventilation of the patients lungs was clinically unacceptable, peak airway pressure exceeded 40 cm H2O, or SpO2 decreased to <90%. At the end of the procedure, anesthesia was discontinued and the device was removed as the patients reflexes returned, in accordance with the manufacturers recommendations.
Statistical process control (x control charts) was used to determine the stability of the spirometry data. The data were considered stable if they were within the limits of the mean value ± 3 SD and in the absence of a "run" or a "trend." Warning limits were fixed within the mean value ± 2 SD. A "run" was defined as at least seven successive points on the same side of the mean value. A "trend" was defined as a succession of seven values that were increasing or decreasing (4).
Data were also analyzed with the Friedman analysis of variance test, followed by Wilcoxons matched pairs signed rank tests with an appropriate correction analysis of variance for repeated measurements. A P value of <0.05 was considered significant.
| Results |
|---|
|
|
|---|
|
|
A gas leak was detected at a ventilation pressure of 25 cm H2O in 10 patients; the leak disappeared after overinflation of the balloon with an additional 10 mL of air. Gastric insufflation was not detected by epigastric auscultation in any patient. The pharyngeal balloon pressure was noted to increase progressively during the procedure, with a mean increase of 14.1 ± 4 cm H2O at 30 min.
Statistical process control charts showed stable systems for all data. Examples of x-charts are presented in Figures 24. No significant statistical differences were observed by the Friedman analysis of variance test for different time points of recording for all data.
|
|
|
| Discussion |
|---|
|
|
|---|
One of the advantages of the Laryngeal Tube is its small intraballoon pressure (60 cm H2O). Inflation with the maximum recommended cuff volume produces intraballoon pressures more than 100 cm H2O in other supraglottic ventilatory devices (49). The incidence and degree of mucosal trauma caused by the pressure exerted by the pharyngeal balloon of the Laryngeal Tube are unknown. Although there is no clear correlation between the intracuff pressure of the supraglottic ventilatory devices and the pressure exerted on the pharyngeal mucosa (5,6,10), it is obvious that decreased intraballoon pressure is a desirable feature in these devices.
With volume-controlled mechanical ventilation, the mean peak respiratory pressure (22 cm H2O) was close to the values obtained with the laryngeal mask airway (LMA) in similar conditions (11,12). We believe that monitoring the cuff pressure of the Laryngeal Tube with the aid of a manometer and readjusting the pressure to its initial values may help avoid excessive pressure on the surrounding tissues. The cuff pressure increase can be explained by the fact that nitrous oxide diffuses more rapidly into the cuff than nitrogen diffuses out of it (13).
The leak fraction calculated by subtracting expired from inspired tidal volume was 10%; however, by auscultation with a stethoscope placed on the neck region, no significant air leak was detected at a peak respiratory pressure of 22 cm H2O with an intra-balloon pressure of 6070 cm H2O. This finding is in agreement with a report of Doerges et al. (14), who found that airway pressures even up to 40 cm H2O would have been possible without gastric insufflation, although an intra-balloon pressure of 90 cm H2O was required.
In this study, the Laryngeal Tube proved to be remarkably atraumatic. The small percentage of cases with upper airway trauma (2.3%), evaluated by checking for the presence of blood on the Laryngeal Tube after removal and by the presence of sore throat (6.8%), is probably due to its design and its low-pressure, high-volume oropharyngeal balloons. Comparative sore throat percentages for the LMA and Combitube® are 5.8%18% (15,16) and 25%48% (10,17), respectively.
Previous studies used the first version of the Laryngeal Tube with two separate pilot tubes for inflation of the two balloons (1,14,18). The improved version of the Laryngeal Tube used in this study has a unique connector to inflate both balloons, minimizing the possibility of overinflation of the distal balloon and, consequently, the risk of esophageal trauma. One of the disadvantages of the Laryngeal Tube is that it is impossible to empty the gastric contents, because the esophagus is blocked.
Recently, Miller et al. (19) reported a frequent failure rate of the Laryngeal Tube during spontaneous ventilation, possibly because of an obstruction of the single ventilation hole by the epiglottis, which occupies a considerable part of its area. This form of obstruction could be more easily overcome by the ventilator during positive pressure mechanical ventilation than by the spontaneously breathing patients efforts.
The improved version of the Laryngeal Tube used in this study has two additional small lateral holes in addition to the principal ventilation hole (2 mm long and 7 mm wide). These additional holes are intended to improve the ventilation by offering an additional route for air passage into the larynx.
This is an observational study, and further comparative studies are needed to compare the Laryngeal Tube with other supraglottic airway devices. However, the frequent rate of successful use (96.6%) of the Laryngeal Tube in mechanical ventilation makes it comparable to the LMA and the Combitube® (12,20).
The Laryngeal Tube is a promising device in the evolution of supraglottic devices and might be an important alternative for airway management. We believe that it is suitable for airway management with mechanical ventilation during general anesthesia of at least moderate duration.