| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
We have introduced a novel, parallel design into a new clinical bymixer (patent pending), named for the bypass of a constant fraction of total flow through a mixing chamber. Over a wide range of tidal volumes (3001200 mL), frequency (620 breaths/min), and PCO2 (650 mm Hg), the bymixer provided accurate measurement of mixed expired gas fractions in the ventilation circuit compared with an expired gas collection in a metabolic lung bench setup (average slope, 1.00; average y intercept, -0.01; average coefficient of determination, R2 = 0.9988). Simple changes in mixing chamber volume provided adjustable bymixer response times. The fast bymixer response (time constant, 6.4 s) should allow measurements to be updated every 20 s (where 95% response occurs by three time constants). The new clinical bymixer is constructed from standard anesthesia circuit components, attaches easily to the anesthesia machine inspired outlet and expired inlet ports, is simple to clean and sterilize, and has no reservoir to trap condensed water vapor from expired gas. The new clinical bymixer may facilitate indirect calorimetry (CO2 elimination, CO2, and oxygen uptake, CO2) during anesthesia and the noninvasive detection of metabolic upset (e.g., onset of anaerobic metabolism) and critical events (e.g., pulmonary embolism). IMPLICATIONS: A new clinical bymixer (inline mixing chamber) provides a fast response and accurate measurements of mixed expired gas fractions in the anesthesia circle circuit. A novel parallel design facilitates adjustable response, easy cleaning, and construction from standard airway circuit components. The new clinical bymixer may facilitate widespread introduction of indirect calorimetry during anesthesia.
Measurement of mixed expired gas concentrations is an essential component of the methodology to measure CO2 elimination ( CO2) and pulmonary oxygen uptake ( O2) at the airway opening (1). In the normal condition in which CO2 is absent from inspired gas, CO2 is given by
where
Conversely,
where I denotes inspiration. Because expired volume is increased by increased temperature and added water vapor (increased humidity), volumes must be corrected to standard temperature and pressure, dry (STPD) conditions, or the error in
Regardless of whether temperature and humidity differences between inspiration and expiration are managed by the Haldane transformation (Equation 3) or by separate measurements of airway temperature and relative humidity (RH) (Equation 2) (3), the determination of
To solve these problems, we have developed a new clinical bymixer from common anesthesia circuit components (Fig. 2). Instead of diverting gas flow into a separate, large mixing chamber, the new clinical bymixer incorporates a novel parallel tubing design (patent pending). A constant fraction of total is diverted through the mixing chamber (tubing), whose volume can be adjusted by varying the length of corrugated tubing (collapsible/expandable pediatric anesthesia circuit tubing). The resistor controls the fraction of bypass to total flow. As gas passes through the mixing chamber, it mixes longitudinally in the tubing. Flow-averaged mixed gas is sampled at the port for analysis by a sidestream sampling monitor.
This study describes the development and construction of the new clinical bymixer. The following questions were tested and answered: Is a constant fraction of main gas flow diverted through the mixing chamber (mandatory for mixed bypass gas samples to accurately represent total gas flow)? Does the longitudinal design of the tubular mixing chamber provide adequate mixing (no significant breath-to-breath variation of mixed gas fractions)? What is the fastest accurate response of the new clinical bymixer when the mixing chamber volume is decreased (shortest length of mixing tubing)? Does the continuous sidestream sampling flow rate affect the measurement of the mixed gas fraction? To test the performance of the new clinical bymixer during cyclical changes in gas fractions under actual expiratory flow conditions, the bymixer was interposed in the exhalation limb of a ventilator attached to a CO2-producing mechanical lung (Fig. 3). Measurement of bymixer mixed expired PCO2 was compared with the value in a gas collection from the exhaust port of the open circuit ventilator. The use of the metabolic lung simulator was mandatory for the execution of this study to provide a wide and controlled range of tidal volume (VT), respiratory frequency (f), and mixed expired PCO2.
Design and Construction of the New Clinical Bymixer (Patent Pending) The bymixer divided the incoming total gas flow into two parallel channels: main flow and bypass flow (Fig. 2). The main flow channel was a 24-cm length of standard 3/4-in. polyvinyl chloride (PVC) pipe (22-mm inner diameter [ID]). The bypass flow/mixing chamber channel was a length of expandable/collapsible pediatric anesthesia circuit tubing (15-mm ID, Expandoflex; Cleveland Tubing Inc., Cleveland, TN). The adjustable tubing was connected, in series, to a sampling port adapter (Datex-Engstrom Division, Instrumentarium Corp., Helsinki, Finland), a flow resistor, and a 12-cm length of standard 3/4-in. PVC tubing. The flow resistor was constructed by drilling a 4-mm-diameter hole in a plastic cap (NAS-820-10; Niagra Plastics, Erie, PA) placed inside a connector (Multi Adapter; Hudson RCI, Temecula, CA; 15-mm ID; 22-mm outer diameter). In this study, the adjustable tubing lengths were 50, 65.5, and 121 cm, which generated mixing chamber volumes (measured up to the sampling port) of 100, 150, and 200 mL, respectively. The volume of the bypass channel from the sampling port to the downstream Y-connector was 53 mL. The main flow and bypass flow channels were con-nected at each end by identical Y-connectors (supplied with standard anesthesia circle circuits). Volumes of channel components were determined by water displacement.
Determination of Time Constant (Bymixer Response) The bymixer response was also tested in the exhalation limb during mechanical ventilation of the CO2-producing metabolic lung simulator (Fig. 3; see below). Steady-state ventilation was established at minute ventilation of 4, 8, or 12 L/min (the respiratory f was 10 breaths/min, and the inspiration/expiration time ratio was 1:2). The bymixer (100-mL mixing chamber) was separately flushed with air. During the inspiratory phase, the bymixer was abruptly interposed in the exhalation limb of the ventilation circuit (Time 0). Because bymixer data during ventilation were periodic and available only during expiration, we measured the time for bymixer PCO2 (infrared analysis) to reach 95% of its maximum value. This time for 95% response was corrected for the PCO2 transport delay (1.76 s) down the sidestream sampling system (9).
Validation of the Accuracy of the New Clinical Bymixer
For each length of mixing chamber expandable tubing, the bymixer was tested during different ventilatory patterns, encompassing combinations of VT (3001200 mL) and respiratory f (620 breaths/min). The inspiratory/expiratory ratio was 1:2. Gas was continuously sampled from the bymixer by the sidestream capnometer (bymixer P
Effect of Tidal Volume and Respiratory Frequency on Oscillations of Bymixer F
Effect of Intermittent (Instead of Continuous) Sampling from the Bymixer Port
Data Analysis
where VBYPASS was the volume of the mixing chamber (measured up to the sampling port) and
where
In the validation of average bymixer PCO2 versus the value measured in the expired gas collection (metabolic lung simulator),
where dt is the digital sampling interval (1/100 Hz) and t0 and tend were the beginning and end sampling times (s) of bymixer PCO2.
Bymixer P
Figure 4 displays the excellent linear regression correlation between bymixer P CO2 and the value measured in the expired gas collection over a wide range of P CO2 (650 mm Hg) for the bymixer with mixing chamber volumes set to 100, 150, and 200 mL. There was no significant difference in bymixer P CO2 accuracy among the mixing chamber volumes (analysis of variance of the P CO2 differences between the bymixer and expired gas collection measurements). For the bymixer set to a mixing chamber volume of 150 mL, there was no significant difference in P CO2 accuracy between continuous and intermittent sampling from the bymixer port.
The Bland-Altman analysis (12) derived the LOA as 0.07 ± 0.93 mm Hg (Fig. 5, top). Measurements for mixing chamber volumes of 100, 150, and 200 mL were combined. Inspection of the graph revealed that the PCO2 difference between the bymixer measurement and the simultaneous value measured in the expired gas collection increased as the measurement increased along the x axis. To correct for this effect, the lower panel of Figure 5 plotted the PCO2 ratio (bymixer/bag) versus the average of the two values (13). The calculation of LOA (1.00 ± 0.03) demonstrated that 95% of the bymixer PCO2 measurements were within 3% of the expired gas collection value.
Figure 6 displays the breath-by-breath oscillations in PCO2 measured during continuous aspiration from the bymixer into the sidestream sampling gas analyzer. Oscillations in PCO2 were larger with the smaller bymixer mixing chamber volume. Table 1 shows that the average PCO2 oscillations increased from 0.1 to 0.7 mm Hg as the bymixer mixing chamber volume decreased from 200 to 100 mL. The plot of PCO2 oscillation (mm Hg) versus VT (mL) (constant f) generated a significant direct relationship (slope = 0.0016; y intercept = -0.69; R2 = 0.92). The plot of PCO2 oscillation (mm Hg) versus f (breaths/min) (constant VT) resulted in a significant inverse relationship (slope = -0.062; y intercept = 1.22; R2 = 0.91). Thus, PCO2 oscillations increased in magnitude as VT increased and f decreased.
The ratio of bypass flow to total flow was similar (1:9) for the three mixing chamber volumes (Table 1). The of the bymixer response to a change in input gas concentration (at 8 L/min) ranged from 6.4 to 14.1 s for the smallest (100 mL) to largest (200 mL) mixing chamber volumes. Tripling of the predicted 95% response. During minute ventilation of the metabolic lung simulator at 4, 8, and 12 L/min, the times for 95% response of the bymixer (100-mL volume) were 19.0, 12.6, and 6.6 s, respectivelysignificantly less than the values of 3 (Table 1).
The new clinical bymixer incorporates a radically new design, compared with the classic bymixer (4,8). Instead of diverting a portion of main flow into a surrounding reservoir, as in the classic bymixer (Fig. 1), the new clinical bymixer diverts a fraction of main flow through a parallel, longitudinal, and adjustable mixing chamber (Fig. 2). The flow resistor (variable orifice) provided an easy control of fraction of bypass flow. To provide an accurate mixed average gas fraction of total flow, the ratio of bypass flow to total flow must remain constant, and gas must adequately mix by the time it reaches the sampling port. Figure 4 demonstrates the excellent correlation of bymixer P CO2 compared with the simultaneous value measured in the expired gas collection over a wide range of VT, f, and PCO2. The Bland-Altman LOA analysis (12,13) (Fig. 5) demonstrates excellent bymixer measurement accuracy, where 95% of the bymixer mea-surements were within 3% of the simultaneous value measured in the mixed expired gas collection. If present, the small bymixer PCO2 oscillations (Fig. 6, Table 1) were time-averaged and did not degrade bymixer performance for mixing chamber volumes of 100, 150, and 200 mL.
For these mixing chamber volumes, the ratio of bypass flow to total flow was similar (1:9) because the major impedance to gas flow was the flow resistor. The increased length of the large-bore mixing chamber tubing did not materially add to bypass flow resistance. Thus, the dynamic response of the bymixer can be improved by decreasing the volume of the mixing chamber tubing (decreased length). Table 1 suggests that the bymixer dynamic response (at 8 L/min) could be improved, beyond (less than) the 9.3-second There was no difference in bymixer accuracy between continuous and intermittent aspiration at the sampling port. Accordingly, the downstream volume (measured from the sampling port) of the bypass channel was sufficiently large that sidestream sampling (200 mL/min) did not spuriously sample gas from the main flow outlet during inspiration (when gas flow through the bymixer was zero). The small bymixer FCO2 oscillations, when present, represented slight incomplete mixing in bypass flow. FCO2 oscillations decreased with smaller VT because the ratio of VT to mixing chamber volume decreased. FCO2 oscillations decreased with higher f (at constant VT) because increased overall gas flow (and velocity) improved gas mixing. The corrugations of the mixing chamber tubing presumably added to gas mixing. The presence of bymixer PCO2 oscillations was not significant, because simple time averaging of the oscillations resulted in excellent bymixer accuracy (Figs. 4 and 5).
In summary, the novel, parallel design of the new clinical bymixer should provide accurate measurement of mixed expired gas fractions in the anesthesia circle circuit. Simple changes in mixing chamber volume allow adjustable bymixer response time. The fast bymixer response (
Supported by National Heart Lung and Blood Grant R01 HL-42637 (PHB, principal investigator). Supported in part by the Department of Anesthesiology, University of California-Irvine, and by National Center for Research Resources Grant M01 RR00827. The authors acknowledge Duoyou Wang, MD, PhD, for preliminary studies of the clinical bymixer. The authors thank David Chien, BSc, computer support specialist, for assistance in bymixer design/testing and digital data acquisition program development and Chris Kirby, BSc, research associate, for assistance in validation studies. The authors also thank Jeffrey C. Milliken, MD, W. Lane Parker, CCP, and Berend J. Ages, CCP, Division of Cardio-Thoracic Surgery, for providing and helping with the precision occlusion roller pump.
This article has been cited by other articles:
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|