*Department of Anaesthesiology and Intensive Care Medicine, University of Bonn, Bonn; and
Division of Pulmonary Pharmacology, Research Center, Borstel, Germany
Address correspondence and reprint requests to Hermann Wrigge, MD, Department of Anesthesiology and Intensive Care Medicine, University of Bonn, Sigmund-Freud-Strasse 25, D-53105 Bonn, Germany. Address e-mail to hermann.wrigge{at}ukb.uni-bonn.de
| Abstract |
|---|
|
|
|---|
IMPLICATIONS: In 62 patients undergoing elective major thoracic or abdominal surgery, mechanical ventilation with low tidal volumes and positive end-expiratory pressure or high tidal volumes and zero end-expiratory pressure did not result in different pulmonary or systemic levels of measured inflammatory markers.
| Introduction |
|---|
|
|
|---|
Clinical studies suggest that mechanical ventilation can modify inflammatory responses in patients with acute lung injury (5). In such patients, with existing pulmonary and systemic inflammation, ventilation with tidal volumes (VT) of 1015 mL/kg ideal body weight and low-to-moderate levels of positive end-expiratory pressure (PEEP) was associated with increased intraalveolar and systemic levels of inflammatory mediators (57). In contrast, mechanical ventilation with moderate-to-high levels of PEEP and low VT of approximately 6 mL/kg ideal body weight assured adequate gas exchange, decreased intraalveolar and systemic mediator levels, and improved outcome (58). Experimental data suggest that mechanical ventilation with higher VT and zero end-expiratory pressure (ZEEP) induces not only cytokine release (911) but also translocation of cytokines from the lungs to the systemic circulation and even vice versa (12). In addition, inflammatory responses to mechanical stress caused by mechanical ventilation were aggravated by inflammatory co-stimulation (9).
In contrast to patients with acute lung injury having a continuing systemic inflammatory reaction, short-term mechanical ventilation alone of adult patients with healthy lungs did not induce a systemic inflammatory response (13). However, it is not known whether mechanical ventilation will aggravate production and/or translocation of cytokines released by an inflammatory co-stimulus such as major surgery. Therefore, we studied pulmonary and systemic mediator levels depending on ventilatory settings in patients undergoing major thoracic or abdominal surgery as a clinical model of ventilator-associated lung injury.
| Methods |
|---|
|
|
|---|
Sixty-four adult patients scheduled for major elective thoracic (n = 34) or abdominal (n = 30) surgery under general anesthesia were eligible to participate in the study. Patients with immunosuppression by drugs or patients preoperative underlying condition and those with an increased white blood cell count or clinical signs of a systemic or pulmonary infection before surgery were not included in the study.
All patients received a standard premedication with midazolam. Anesthesia was performed using propofol, remifentanil, and cisatracurium. Patients who underwent thoracotomy were tracheally intubated with a double-lumen tube (Mallinckrodt, Athlone, Ireland) to allow single-lung ventilation. Mechanical ventilation was provided with an anesthesia ventilator connected to a circle system (Julian®; Dräger, Lübeck, Germany) using a fraction of inspired oxygen of 0.30 and 1.0 during single-lung ventilation. Routine perioperative monitoring included measurement of invasive arterial and central-venous blood pressures (AS/3®; Datex-Ohmeda, Helsinki, Finland). End-tidal fractions of carbon dioxide were monitored using infrared absorption capnography (Julian®; Dräger). Arterial blood gases, oxygen saturation, hemoglobin, and pH were determined immediately after sampling with standard blood gas electrodes (ABL; Radiometer, Copenhagen, Denmark) or spectrophotometry (OSM3; Radiometer).
Ventilatory Measurements
Gas flow and airway pressure were measured at the proximal end of the tracheal tube with a standard monitor for ventilatory measurements (CP 100; Bicore, Irvine, CA). All signals were sampled with an analog/digital converter board (PCM-DAS16S/12, Computer Boards Inc., Mansfield, MA) installed in a personal computer. Digitized signals were stored on magnetic media for off-line analysis. Inspiratory time and ventilatory cycle time were identified by analyzing the gas flow curve. VT and minute ventilation were derived from the integrated gas flow signal.
Cytokine and Chemokine Measurements
Arterial ethylenediaminetetraacetic acid blood samples of 5 mL were centrifuged at 1500g for 5 min; the plasma was aspirated and stored at -70°C. Tracheal fluid was aspirated after instillation of 10 mL of sterile isotonic saline. In thoracic surgery patients, fluid was always instilled and aspirated in the ventilated lumen of the double-lumen tube. The aspirated tracheal fluid was centrifuged at 500g for 15 min and the supernatants were stored at -70°C. Commercially available Cytometric Bead Array (BD Biosciences, Heidelberg, Germany), which allows simultaneous detection of multiple soluble analytes in a particle-based immunoassay, was used to quantitatively measure interleukin (IL)-8, IL-1ß, IL-6, IL-10, tumor necrosis factor (TNF)-
, and IL-12p70 protein levels in tracheal aspirate and serum samples. The assays were performed according to the manufacturers instructions and were analyzed using the BD Cytometric Bead Array software.
Protocol
Baseline blood samples were obtained immediately after induction of anesthesia. Patients were then randomly (sealed envelopes) assigned to receive either mechanical ventilation with high VT (12 or 15 mL/kg ideal body weight during thoracic or abdominal surgery, respectively) and ZEEP (high VT and ZEEP group) or VT of 6 mL/kg ideal body weight and PEEP of 10 cm H2O (low VT and PEEP group). The ventilator rate was adjusted to maintain PaCO2 between 35 and 45 mm Hg and ventilatory measurements were performed during stable conditions. In the thoracic surgery group, the ventilatory setting remained unchanged during one-lung ventilation but inspiratory pressure was limited to 35 cm H2O. In case this pressure limit was reached, the setting was changed to pressure-limited ventilation with decelerating inspiratory flow usually allowing decreasing of pressure limit below the upper limit to meet the desired VT. Additional blood was sampled 1, 2, and 3 h after initiation of mechanical ventilation finally followed by tracheal aspiration after 3 h only.
To detect differences in mediator levels between the two ventilatory settings with respect to the subgroups thoracic/abdominal surgery with the given two-sided parallel design at a significance level of 5% (
= 0.05) with a probability of 80% (ß = 0.20) based on an estimated difference of 0.84 of the parameters mean SD, the number of patients to be studied in each group is at least 30.
Results are expressed as mean ± SD unless otherwise stated. All statistical analyses were performed using a statistical software package (STATISTICA for Windows 6.0; StatSoft, Inc., Tulsa, OK). Data were tested for normal distribution with the Shapiro-Wilks W test and analyzed by using one-way or repeated-measures analysis of variance. Data of plasma mediators were analyzed separately after log10 transfor-mation to achieve homoscedasticity (homogenous variances of data sets) and normal distribution. Groups were analyzed by Tukeys post hoc test only when a significant F ratio was obtained, with differences either within the same type of surgery or within the same ventilatory setting between surgery. Because mediator concentrations in tracheal aspirates still differed significantly from normal distribution even after log10 transformation, these data were analyzed using the Mann-Whitney U-test. Although this study was not designed to control for the influence of possible cofactors, such as age, gender, and smoking status, the potential influence of these factors was studied by backward regression together with the factors mode of ventilation and surgery. Differences were considered to be statistically significant if P < 0.05.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Our findings seem to be in contrast to basic research demonstrating initiation or perpetuation of local and systemic inflammatory responses to injurious ventilatory strategies using high VT and/or low or zero PEEP (5,9,11,14,15). High VT mechanical ventilation resulted in increased production of proinflammatory cytokines within a few hours whereas low VT and high PEEP ventilatory strategies were associated with reduced mediator levels.
Clinically, it was demonstrated that low VT ventilation of 68 mL/kg ideal body weight with moderate (7,16) or high levels of PEEP (8) can decrease mortality in acute lung injury or acute respiratory distress syndrome when compared with mechanical ventilation with high VT of at least 12 mL/kg ideal body weight, and that this correlates with lower pulmonary and/or systemic mediator concentrations (5,6). Thus, although ample evidence indicates that mechanical ventilation may alter an existing inflammatory response in patients with acute lung injury, it is not entirely clear whether mechanical ventilation alone can stimulate cytokine production from healthy lungs. In patients with normal lungs ventilated without surgical co-stimuli for one hour, we did not observe changes in the plasma levels of inflammatory mediators (13). Plötz et al. (17) reported an increase in pulmonary TNF and IL-6 levels in anesthetized infants during diagnostic heart catheterization after two hours of conventional mechanical ventilation (17). Unfortunately, the authors did not control for the factors mechanical ventilation, anesthesia, radiation, and the use of a contrast agent.
Whereas in small animal models mechanical ventilation with sufficiently high VT alone is able to trigger an inflammatory response, the somewhat conflicting observations in adults with and without previous pulmonary inflammation may be explained by a two-hit model. According to this model, pulmonary inflammation must already be present (first hit) for injurious mechanical ventilation (second hit) to aggravate the inflammatory response. The physiological basis for such a two-hit model may be given by the fact that lungs from acute respiratory distress syndrome patients are inhomogeneously injured (18), so that the healthy alveoli may overinflate because the atelectatic alveoli produce no counterpressure (19). This hypothesis is supported by several experimental studies showing increased inflammatory responses to high VT mechanical ventilation after an inflammatory first hit (9,10,20).
Previous studies have demonstrated a certain amount of alveolar collapse and atelectasis soon after induction of anesthesia and mechanical ventilation in previously healthy patients (21) which can be prevented with PEEP of 10 cm H2O (22). Because the amount of atelectasis might even increase during surgery and lung volume is reduced during one-lung ventilation, high VT could lead to inspiratory overdistension of lung units which is a contributing factor for ventilator-associated lung injury (23). Thus, the low VT and PEEP ventilatory setting chosen in this study should have attenuated both the alveolar collapse and the overdistension in comparison to the high VT and ZEEP setting (22).
Both of these settings may not reflect the average clinical setting mainly used in the operating room; however, this study was designed to investigate inflammatory effects of different ventilatory settings in a clinical model of ventilator-associated lung injury. In addition, ventilation with high VT and ZEEP may still be used during one-lung ventilation (24), although recent experimental data suggest that ventilator-induced lung injury occurs with this setting (25).
If the two-hit theory were true, the lack of difference between the two ventilatory settings would suggest that even during thoracic surgery, only minimal lung injury was inflicted and that the inflammatory responses to major surgery were not a sufficient "first hit." This is in contrast to patients with acute lung injury already having pulmonary and systemic inflammation and requiring higher inspiratory airway pressures as did our patients with healthy lungs (5,6).
Surgical stimuli during major abdominal and thoracic surgery are associated with increased plasma levels of proinflammatory markers (14). In the present study, we extended those previous findings by showing that systemic, but not pulmonary cytokine levels are slightly higher during abdominal surgery. These differences, which likely are biologically insignificant, may be related to minor exposure to intestinal bacteria. Levels of all mediators, except IL-8 were rather low. The higher IL-8 levels are in line with other observations showing increased pulmonary and systemic IL-8 levels in patients undergoing esophagectomy (26,27).
It is of importance to note that we tested each ventilatory setting for only three hours. However, in previous experimental and clinical studies, this period of time was sufficient to cause ventilation-induced mediator release. Experimentally, intraalveolar TNF expression (28) and increased TNF levels in the systemic circulation (10) occurred after one hour of high VT mechanical ventilation in lung injury models. Clinically, in patients with acute lung injury, alveolar and systemic cytokine concentrations increased within one hour after initiating mechanical ventilation with low PEEP and high VT (6). Therefore, and because we observed an inflammatory response to surgery within the study period of three hours, it is not likely that three hours was too short for ventilation-induced cytokine release to occur.
Our data suggest that, in uninjured normal lungs of patients during major abdominal or thoracic surgery, mechanical ventilation with high VT and ZEEP does not increase the pulmonary or systemic inflammatory responses to surgery within three hours. This observation is indirect evidence for the two-hit hypothesis suggesting that high VT and low PEEP mechanical ventilation may induce lung inflammation to clinically important levels in preinjured or infected lungs in the short term, as previously shown, but not in normal lungs even during major surgery.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E R Fernandez-Perez, J Sprung, B Afessa, D O Warner, C M Vachon, D R Schroeder, D R Brown, R D Hubmayr, and O Gajic Intraoperative ventilator settings and acute lung injury after elective surgery: a nested case control study Thorax, February 1, 2009; 64(2): 121 - 127. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. T. Lytle and D. R. Brown Appropriate Ventilatory Settings for Thoracic Surgery: Intraoperative and Postoperative Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2008; 12(2): 97 - 108. [Abstract] [PDF] |
||||
![]() |
R. M. Determann, E. K. Wolthuis, G. Choi, P. Bresser, A. Bernard, R. Lutter, and M. J. Schultz Lung epithelial injury markers are not influenced by use of lower tidal volumes during elective surgery in patients without preexisting lung injury Am J Physiol Lung Cell Mol Physiol, February 1, 2008; 294(2): L344 - L350. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Pruszkowski, N. Dalibon, M. Moutafis, E. Jugan, J. D. Law-Koune, P. A. Laloe, and M. Fischler Effects of propofol vs sevoflurane on arterial oxygenation during one-lung ventilation Br. J. Anaesth., April 1, 2007; 98(4): 539 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Wurfel Microarray-based Analysis of Ventilator-induced Lung Injury Proceedings of the ATS, January 1, 2007; 4(1): 77 - 84. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Frank, P. E. Parsons, and M. A. Matthay Pathogenetic Significance of Biological Markers of Ventilator-Associated Lung Injury in Experimental and Clinical Studies Chest, December 1, 2006; 130(6): 1906 - 1914. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Slinger Pro: low tidal volume is indicated during one-lung ventilation. Anesth. Analg., August 1, 2006; 103(2): 268 - 270. [Full Text] [PDF] |
||||
![]() |
E. Crimi, H. Zhang, R. N. N. Han, L. D. Sorbo, V.M. Ranieri, and A. S. Slutsky Ischemia and Reperfusion Increases Susceptibility to Ventilator-induced Lung Injury in Rats Am. J. Respir. Crit. Care Med., July 15, 2006; 174(2): 178 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schilling, A. Kozian, C. Huth, F. Buhling, M. Kretzschmar, T. Welte, and T. Hachenberg The Pulmonary Immune Effects of Mechanical Ventilation in Patients Undergoing Thoracic Surgery Anesth. Analg., October 1, 2005; 101(4): 957 - 965. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Zupancich, D. Paparella, F. Turani, C. Munch, A. Rossi, S. Massaccesi, and V. M. Ranieri Mechanical ventilation affects inflammatory mediators in patients undergoing cardiopulmonary bypass for cardiac surgery: A randomized clinical trial J. Thorac. Cardiovasc. Surg., August 1, 2005; 130(2): 378 - 383. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. P. Grichnik and T. A. D'Amico Acute Lung Injury and Acute Respiratory Distress Syndrome After Pulmonary Resection Seminars in Cardiothoracic and Vascular Anesthesia, December 1, 2004; 8(4): 317 - 334. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|