Anesth Analg 2009;0:ane.0b013e31818f87e9
© 2009 International Anesthesia Research Society
doi: 10.1213/ane.0b013e31818f87e9
Eliminating Arterial Injury During Central Venous Catheterization Using Manometry
Catalin S. Ezaru, MD,
Michael P. Mangione, MD,
Todd M. Oravitz, MD,
James W. Ibinson, MD, and
Richard J. Bjerke, MD
From the Department of Anesthesiology, Veterans Affairs Medical Center, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania.
Address correspondence and reprint requests to Catalin S. Ezaru, MD, Department of Anesthesiology, 3A101, Veterans Affairs Medical Center, University Dr. C, Pittsburgh, PA 15240. Address e-mail to ezaruc{at}anes.upmc.edu.
Abstract
Background: Unintended arterial puncture occurs in 2%–4.5% of central venous catheterizations, resulting in arterial injury in 0.1%–0.5% of patients. Routine performance of manometry during catheterization may successfully identify unintended arterial puncture and avoid arterial cannulation and injury.
Methods: We conducted a retrospective review of all cases of central venous catheter placement during a 15-yr period after implementation of a safety program requiring mandatory use of manometry to verify venous access. Arterial injuries were defined as unintended arterial cannulations with a 7-French or larger catheter or dilator. Arterial punctures were defined as the unintended placement of an 18-gauge catheter or needle into the artery. Data were reviewed for all arterial injuries during the entire 15-yr period. In addition, data on both arterial puncture and subsequent arterial injury were evaluated during the final year of analysis.
Results: A total of 9348 central venous catheters were placed during the observation period. During the full 15 yr of observation, there were no cases of arterial injury. During the final year of assessment, 511 central venous catheters were placed, with arterial punctures in 28 patients (5%). Arterial puncture was recognized without manometry in 24 cases. Arterial puncture was identified only with manometry in 4 cases, with no incidents of arterial injury.
Conclusions: Consistent use of manometry, to verify venous placement, during central venous catheterization effectively eliminated arterial injury from unintended arterial cannulation during the 15-yr assessment.
This article has been cited by other articles:

|
 |

|
 |
 
J. W. Ibinson, T. M. Oravitz, C. S. Ezaru, and M. P. Mangione
Clarification of Issues Regarding Manometry for Central Venous Catheterization
Anesth. Analg.,
December 1, 2009;
109(6):
2028 - 2028.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. J. C. Mittnacht
Ultrasound-Guided Central Venous Cannulation: False Sense of Security
Anesth. Analg.,
December 1, 2009;
109(6):
2029 - 2029.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Bowdle, E. Kharasch, and H. Schwid
Pressure Waveform Monitoring During Central Venous Catheterization
Anesth. Analg.,
December 1, 2009;
109(6):
2030 - 2031.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Ellison, D. R. Jobes, and A. J. Schwartz
Primum Non Nocere
Anesth. Analg.,
December 1, 2009;
109(6):
2028 - 2029.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. Sandberg
"Just Do It"
Anesth. Analg.,
December 1, 2009;
109(6):
2029 - 2030.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. B. Leibowitz and M. A. Rozner
PRO: Manometry Should Routinely Be Used During Central Venous Catheterization
Anesth. Analg.,
July 1, 2009;
109(1):
3 - 5.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. J. Leone
CON: Manometry During Internal Jugular Cannulation: Case Not Proven
Anesth. Analg.,
July 1, 2009;
109(1):
6 - 7.
[Full Text]
[PDF]
|
 |
|
|